초고주파 가열치료를 위한 MR 호환 동축 슬롯 안테나의 개발

Development of MR Compatible Coaxial-slot Antenna for Microwave Hyperthermia

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

초록

MR compatible coaxial-slot antenna for microwave hyperthermia was developed while its structure and size of each part were determined by computer simulation using finite element method(FEM). Its local heating performance was evaluated using tissue-mimic phantom and swine muscles. 2% agarose gel mixed with 6mM/ℓ MnCl2 as a biological tissue-mimic phantom was heated by the proposed antenna driven by a 2.45GHz microwave generator. The temperature changes of the phantom were monitored using multi-channel digital thermometer at the distance of 0mm, 5mm, 10mm and 20mm from the tip center of the antenna. Also muscle tissue of swine was heated for 2 and 5minutes with 50W and 30W of microwave generator powers, respectively, to evaluate the local heating performance of the antenna. MRI compatibility was also verified by acquiring MR images and MR temperature map. MR signals were acquired from the agarose gel phantom using T2* GRE sequence with 1.5T clinical MRI scanner(Signa Echospeed, GE, Milwaukee, WI, U.S.A.) at Pusan Paik Hospital and were transferred to PC in order to reconstruct MR images and temperature map using proton resonance frequency(PRF) method and laboratory-developed phase unwrapping algorithm. Authors found that it has no severe distortion due to the antenna inserted into the phantom. Finally, we can conclude that the suggested coaxial-slot antenna has an excellent local heating performance for both of tissue-mimic phantom and swine muscle, and it is compatible to 1.5T MRI scanner

Abstract

MR compatible coaxial-slot antenna for microwave hyperthermia was developed while its structure and size of each part were determined by computer simulation using finite element method(FEM). Its local heating performance was evaluated using tissue-mimic phantom and swine muscles. 2% agarose gel mixed with 6mM/ℓ MnCl2 as a biological tissue-mimic phantom was heated by the proposed antenna driven by a 2.45GHz microwave generator. The temperature changes of the phantom were monitored using multi-channel digital thermometer at the distance of 0mm, 5mm, 10mm and 20mm from the tip center of the antenna. Also muscle tissue of swine was heated for 2 and 5minutes with 50W and 30W of microwave generator powers, respectively, to evaluate the local heating performance of the antenna. MRI compatibility was also verified by acquiring MR images and MR temperature map. MR signals were acquired from the agarose gel phantom using T2* GRE sequence with 1.5T clinical MRI scanner(Signa Echospeed, GE, Milwaukee, WI, U.S.A.) at Pusan Paik Hospital and were transferred to PC in order to reconstruct MR images and temperature map using proton resonance frequency(PRF) method and laboratory-developed phase unwrapping algorithm. Authors found that it has no severe distortion due to the antenna inserted into the phantom. Finally, we can conclude that the suggested coaxial-slot antenna has an excellent local heating performance for both of tissue-mimic phantom and swine muscle, and it is compatible to 1.5T MRI scanner

참고문헌 (19)

  1. 신국선, "Hyperthermia용 Coaxial-Slot 안테나의 SAR 분석", 한국전자파학회 논문지, 13(7), pp.732~739 (2002), 한국전자파학회
  2. S.S. Han, "Extracorporeal High intensity focused ultrasound therapy", J. of Korean Bone & Joint Tumor Soc, 11(1), pp.17~24 (2005)
  3. O. Rouviere, "Transrectal high-intensity focused ultrasound ablation of prostate cancer: Effective treatment requiring accurate imaging", European J. of Radiology, 63, pp.317~327 (2007)
  4. K.H. Haraldsdottir, "Interstitial laser thermotherapy(ILT) of breast cancer", EJSO, 34, pp.739~745 (2008)
  5. S.A. Curley, "Radiofrequency ablation of malignant liver tumors", The Oncologist, 6, pp.14~23 (2001)
  6. K. Sato, "Alternate biplanar MR navigation for microwave ablation of liver tumors", Magnetic Resonance in Medical Sciences, 4(2), pp.89~94 (2005)
  7. M.S. Breen, "MRI-guided laser thermal ablation: Model to predict cell death from MR thermometry images for real-time therapy monitoring", Proc. 26th IEEE EMBS, pp.1028~1031 (2004)
  8. C.M.C. Tempany, "MR imaging-guided focused ultrasound surgery of uterine leiomyomas: A feasibility study", Radiology, 226(3), pp.897~905 (2003)
  9. 참고문헌 목록을 붙여넣기 하십시오. 이전 다음 자료유형 학술지(정기간행물) 학술대회논문 단행본 보고서 학위논문 인터넷자원 기타자료 E.A. stewart, "Clinical outcomes of focused ultrasound surgery for the treatment of uterine fibroids", ASRM, 85(1), pp.22~29 (2006)
  10. H. Furusawa, "Magnetin resonance- guided focused ultrasound surgery of breast cancer: Reliability and Effectiveness", The American College of Surgeons, 203(1), pp.54~63 (2006)
  11. T. Uchida, "Five years experience of transrectal high-intensity focused ultrasound using the sonablate device in the treatment of localized prostate cancer", J. of Urology, 13, pp.228~233 (2006)
  12. D.B. Zippel, "The use of MR imaging guided focused ultrasound in breast cancer patients; a preliminary phase one study and review", Breast Cancer, 12(1), pp.32~38 (2005)
  13. M.L. Coiffe, "Real-time monitoring of radiofre- quency ablation of rabbit liver by respiratoty-gated quantitative temperature MRI", J. of MRM, 24, pp.152~159 (2006)
  14. D. Germain, "MR monitoring of laser-induced lesion of the liver in vivo in a low-field open magnet: Temperature mapping and lesion size prediction", J. of MRM, 13, pp.42~49 (2001)
  15. C. Mougenot, "Automatic spatial and temporal temperature control for MR-guided focused ultrasound using fast 3D MR thermometry and multispiral tragectory of the focal point", Magnetic Resonance in Medicine, 52, pp.1005~1015 (2004)
  16. B.D. Senneville, "MR thermometry for monitoring tumor ablation", Eur Radiol, 17, pp.2401~2410 (2007)
  17. F. Sterzer, "Microwave Medical Devices", pp.65~70 (2002), IEEE microwave magazine
  18. Compilation of the dielectric properties of body tissues at RF and microwave frequencies, "C. Gabriel", ~, pp.Brooks Air Force Technical Report AL (1996), OE-TR-1996-0037
  19. Kee Chin Tan, "Center Array-Sequencing 위상펼침 기법의 MR 온도영상 적용에 관한 기초연구", 대한자기공명의과학회지, 12(2), pp.131~141 (2008), 대한자기공명의과학회

관련 논문AI추천

이 간행물의 다른 논문 (10건)