고주파 소작기의 전극 디자인에 따른 출력 특성 연구 Ⅱ

A Study on the Output Characteristics of High-Frequency Ablation Devices based on Electrode Design II

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류기계공학

초록

Five novel miniature bipolar radiofrequency (RF) electrode tips with distinct tip geometries (spherical, flat, square, and 45° angled) were developed to enable high-precision tissue ablation. Performance was evaluated on saline-soaked tissue, ex vivo bovine liver, and porcine muscle under consistent RF power settings. All designs produced highly localized lesions only a few millimeters across, confirming precise ablation with minimal damage to surrounding tissue. Tip geometry influenced ablation efficiency: a 45° angled tip created ~5 mm lesions at lower power (highest efficiency), whereas an ultra-fine 1.0 mm tip produced ~1 mm lesions but required higher power. These results indicate that the new bipolar RF electrodes achieve precise, localized tissue ablation with minimal surrounding tissue damage and show promise for precise lesion removal in minimally invasive surgery.

Abstract

Five novel miniature bipolar radiofrequency (RF) electrode tips with distinct tip geometries (spherical, flat, square, and 45° angled) were developed to enable high-precision tissue ablation. Performance was evaluated on saline-soaked tissue, ex vivo bovine liver, and porcine muscle under consistent RF power settings. All designs produced highly localized lesions only a few millimeters across, confirming precise ablation with minimal damage to surrounding tissue. Tip geometry influenced ablation efficiency: a 45° angled tip created ~5 mm lesions at lower power (highest efficiency), whereas an ultra-fine 1.0 mm tip produced ~1 mm lesions but required higher power. These results indicate that the new bipolar RF electrodes achieve precise, localized tissue ablation with minimal surrounding tissue damage and show promise for precise lesion removal in minimally invasive surgery.

관련 논문AI추천

이 간행물의 다른 논문 (38건)