초록

In this study, we investigated the rotational characteristics which were comprised of directionality and linearity of target registration error (TRE) as a study in advance to enhance the accuracy of contour-based registration in neuronavigation. For the experiment, two rigid head phantoms that have different faces with specially designed target frame fixed inside of the phantoms were used. Three-dimensional coordinates of facial surface point cloud and target point of the phantoms were acquired using computed tomography (CT) and 3D scanner. Iterative closest point (ICP) method was used for registration of two different point cloud and the directionality and linearity of TRE in overall head were calculated by using 3D position of targets after registration. As a result, it was represented that TRE had consistent direction in overall head region and was increased in linear fashion as distance from facial surface, but did not show high linearity. These results indicated that it is possible for decrease TRE by controlling orientation of facial surface point cloud acquired from scanner, and the prediction of TRE from surface registration error can decrease the registration accuracy in lesion. In the further studies, we have to develop the contour-based registration method for improvement of accuracy by considering rotational characteristics of TRE.

Abstract

In this study, we investigated the rotational characteristics which were comprised of directionality and linearity of target registration error (TRE) as a study in advance to enhance the accuracy of contour-based registration in neuronavigation. For the experiment, two rigid head phantoms that have different faces with specially designed target frame fixed inside of the phantoms were used. Three-dimensional coordinates of facial surface point cloud and target point of the phantoms were acquired using computed tomography (CT) and 3D scanner. Iterative closest point (ICP) method was used for registration of two different point cloud and the directionality and linearity of TRE in overall head were calculated by using 3D position of targets after registration. As a result, it was represented that TRE had consistent direction in overall head region and was increased in linear fashion as distance from facial surface, but did not show high linearity. These results indicated that it is possible for decrease TRE by controlling orientation of facial surface point cloud acquired from scanner, and the prediction of TRE from surface registration error can decrease the registration accuracy in lesion. In the further studies, we have to develop the contour-based registration method for improvement of accuracy by considering rotational characteristics of TRE.

참고문헌 (20)

  1. A. Mert, "Advanced cranial navigation", Neurosurgery, 72(Suppl 1), pp.43~53 (2013)
  2. R. Marmulla, "Automated laser registration in image-guided surgery : evaluation of the correlation between laser scan resolution and navigation accuracy", Int. J. Oral Maxillofac. Surg, 33(7), pp.642~648 (2004)
  3. W. K. Pfisterer, "Fiducial versus nonfiducial neuronavigation registration assessment and considerations of accuracy", Neurosurgery, 62(3 Suppl 1), pp.201~207 (2008)
  4. J. Schlaier, "Registration accuracy and practicability of laser-directed surface matching", Comput. Aided Surg, 7(5), pp.284~290 (2002)
  5. N. McLaughlin, "Neuronavigation in endonasal pituitary and skull base surgery using an autoregistration mask without head fixation : an assessment of accuracy and practicality", J. Neurol. Surg. A. Cent. Eur. Neurosurg, 73(6), pp.351~357 (2012)
  6. A. I. Omara, "Anatomical landmarks for point-matching registration in image-guided neurosurgery", The Int. J. Med. Robot. Comp, 10(1), pp.55~64 (2014)
  7. J. M. Fitzpatrick, "The role of registration in accurate surgical guidance", Proc. Inst. Mech. Eng. H, 224(5), pp.607~622 (2010)
  8. G. Eggers, "Image-to-patient registration techniques in head surgery", Int. J. Oral Maxillofac. Surg, 35(12), pp.1081~1095 (2006)
  9. Y. Fan, "A new markerless patient-to-image registration method using a portable 3D scanner", Med. Phys, 41(10), pp.101910~ (2014)
  10. R. R. Shamir, "Surface-based facial scan registration in neuronavigation procedures : a clinical study : Clinical article", J. Neurosurg, 111(6), pp.1201~1206 (2009)
  11. A. Raabe, "Laser surface scanning for patient registration in intracranial image-guided surgery", Neurosurgery, 50(4), pp.797~803 (2002)
  12. G. J. Ledderose, "Surface laser registration in ENT-surgery : accuracy in the paranasal sinuses-a cadaveric study", Rhinology, 45(4), pp.281~ (2007)
  13. C. R. Mascott, "Quantification of true in vivo(application)accuracy in cranial image-guided surgery : influence of mode of patient registration", Neurosurgery, 59(1 Suppl 1), pp.ONS146~ONS156 (2006)
  14. P. A. Woerdeman, "Application accuracy in frameless image-guided neurosurgery : a comparison study of three patient-to-image registration methods", J. Neurosurg, 106(6), pp.1012~1016 (2007)
  15. S. M. Hardy, "A comparison of computer-aided surgery registration methods for endoscopic sinus surgery", Am. J. Rhinol, 20(1), pp.48~52 (2006)
  16. M. N. Wang, "Properties of the target registration error for surface matching in neuronavigation", Comput. Aided Surg, 16(4), pp.161~169 (2011)
  17. K. Schicho, "Comparison of laser surface scanning and fiducial marker-based registration in frameless stereotaxy : technical note", J. Neurosurg, 106(4), pp.704~709 (2007)
  18. W. E. Lorensen, "Marching cubes: A high resolution 3D surface construction algorithm", ACM siggraph computer graphics, pp.163~169 (1987)
  19. P. BESL, "A method for registration of 3-D shapes", IEEE Trans. Pattern Anal. Mach. Intell, 14(2), pp.239~256 (1992)

관련 논문AI추천