본문 바로가기
KERD
논문KCI 등재

방사선촬영 보조기구를 이용한 어린이 흉부 엑스선 검사에 관한 연구

A Study on Chest X-ray Using Ancillary Device for Child Radiography

0

조회

0

다운로드

4

인용

0

참고문헌

키워드

초록

In this study, We developed a Ancillary device for child radiography for X-ray of children under 5 years old and verified its effectiveness. Chest X-rays of children younger than 5 years of age were performed by Supine method at the position of Table detector, Short - Source to Image Receptor Distance(SID). Existing Supine and Short -SID imaging methods cause many problems, such as errors in image reading and excessive radiation exposure dose to patients, but the use of an Ancillary device for child radiography(ADCR) solves these problems. A total of 160 children were divided into the Upright group using ADCR and Supine group without ADCR. The chest X-ray image was visually evaluated by two radiologists with reference to the European Commission's List of Quality Criteria for Diagnostic Radiographic Images in Pediatrics. The total score of the qualitative evaluation was 5.15% higher in the chest upright method using ADCR than in the chest supine method without ADCR, and the chest upright method score was higher than that of the chest supine method in items 1 to 7. whether infants have deep inspiration or not, 4.87% higher for item 1, whether infants rotate or not and the degree of tilting, 0% higher for the item 2, the reproduction of image from just above apices of lungs to T12/L1, 0% for the item 3, reproduction of the vascular pattern in central 2/3 of the lungs, 6.92% higher for the item 4, reproduction of the trachea and the proximal bronchi, 12.9% higher for the item 5, visually sharp reproduction of the diaphragm and costo-phrenic angles, 10% higher for the item 6, reproduction of the spine and paraspinal structures and visualisation of the retrocardiac lung and the mediastinum, and 3.65% higher for the item 7. Items 2 and 3 showed no statistically significant differences(P > 0.05), and items 1, 4, 5, 6, and 7 showed statistically significant differences(P < 0.05). In conclusion, Upright method using ADCR in pediatric chest X-ray is considered as a good alternative to existing Supine method.

Abstract

In this study, We developed a Ancillary device for child radiography for X-ray of children under 5 years old and verified its effectiveness. Chest X-rays of children younger than 5 years of age were performed by Supine method at the position of Table detector, Short - Source to Image Receptor Distance(SID). Existing Supine and Short -SID imaging methods cause many problems, such as errors in image reading and excessive radiation exposure dose to patients, but the use of an Ancillary device for child radiography(ADCR) solves these problems. A total of 160 children were divided into the Upright group using ADCR and Supine group without ADCR. The chest X-ray image was visually evaluated by two radiologists with reference to the European Commission's List of Quality Criteria for Diagnostic Radiographic Images in Pediatrics. The total score of the qualitative evaluation was 5.15% higher in the chest upright method using ADCR than in the chest supine method without ADCR, and the chest upright method score was higher than that of the chest supine method in items 1 to 7. whether infants have deep inspiration or not, 4.87% higher for item 1, whether infants rotate or not and the degree of tilting, 0% higher for the item 2, the reproduction of image from just above apices of lungs to T12/L1, 0% for the item 3, reproduction of the vascular pattern in central 2/3 of the lungs, 6.92% higher for the item 4, reproduction of the trachea and the proximal bronchi, 12.9% higher for the item 5, visually sharp reproduction of the diaphragm and costo-phrenic angles, 10% higher for the item 6, reproduction of the spine and paraspinal structures and visualisation of the retrocardiac lung and the mediastinum, and 3.65% higher for the item 7. Items 2 and 3 showed no statistically significant differences(P > 0.05), and items 1, 4, 5, 6, and 7 showed statistically significant differences(P < 0.05). In conclusion, Upright method using ADCR in pediatric chest X-ray is considered as a good alternative to existing Supine method.

참고문헌 (16)

  1. "Press Releases, Conducted national radiological records management program for CT patients"[URL]
  2. K.S Jung, "Image Analysis of Pediatric Chest Radiography", Proc. 2004 Spring Meeting of the Seoul Radiological Technologists Association, 10(1), pp.83~88 (2004)
  3. C.W Song, "Child and Infant Chest AP examination's usefulness which use assistance tool", Proc. 2005 Spring Meeting of the Seoul Radiological Technologists Association, 11(1), pp.39~46 (2005)
  4. E.D Frank, "Merrill's atlas of radiographic positioning & procedures", pp.496~498 (2012)
  5. N. Raby, "Accident and Emergency Radiology: A Survival Guide E-Book", pp.307~328 (2014), Elsevier Health Sciences
  6. 이희정, "소아 단순 흉부 X-선 사진의 판독 -정상 및 폐렴 소견을 중심으로-", Clinical and Experimental Pediatrics, 45(11), pp.1311~1316 (2002), 대한소아청소년과학회
  7. European guidelines on quality criteria for diagnostic radiographic images in paediatrics, EUR. 16261, "European Commission", ~ (1996)
  8. Protection of the patient in x-ray diagnosis, Publication 16, "International Commission on Radiological Protection", ICRP, 18~20 (1970)
  9. Wagner, Louis K., "Potential biological effects following high X-ray dose interventional procedures", Journal of Vascular and Interventional Radiology, 5(1), pp.71~84 (1994)[DOI]
  10. Kebart, R. C., "Benefits of increasing focal film distance", Radiologic technology, 62(6), pp.434~442 (1991)
  11. Brennan, P. C., "Increasing FFD: an effective dose-reducing tool for lateral lumbar spine investigations", Radiography, 4(4), pp.251~259 (1998)[DOI]
  12. S.J Kim, "The great encyclopedia of nursing science" (1997), Korea Dictionary Research
  13. Jones, A. Kyle, "Calculating the peak skin dose resulting from fluoroscopically guided interventions. Part I: Methods", Journal of Applied Clinical Medical Physics, 12(4), pp.231~244 (2011)
  14. Valentin, Jack, "Radiation and your patient: A guide for medical practitioners: ICRP Supporting Guidance 2Approved by ICRP Committee 3 in September 2001", Annals of the ICRP, 31(4), pp.1~52 (2001)
  15. Robinson, J, "Extended focal-film distance technique: an analysis of the factors in dose reduction for the AP knee radiograph", Radiography, 7(3), pp.165~170 (2001)[DOI]
  16. J.M Lee, "Radiation Imaging Auxiliary Device and Radiation Imaging Apparatus, Korean. Patent 10-1625057", ~ (2016)

관련 논문AI추천