A Study on the Evaluation of Exposure Dose and Image Quality as a Function of Exposure Conditions in Head Radiography for Children
1
조회
0
다운로드
0
인용
0
참고문헌
키워드
초록
In this study, we analyzed exposure dose and image quality as a function of exposureconditions in head radiography in radiation-sensitive children less than five years old. To compareexposure doses, we used measures of the absorbed dose, such as DAP and ESD. For image quality,the PSNR was used to quantitatively assess noise characteristics. The SRS-78 program was usedto simulate changes in X-ray beam quality in order to determine optimal exposure conditions. The DICOM header information of Infinitt Piview was used to examine skull AP and lateralradiography conditions for head radiographies performed in children less than five years oldbetween January 2018 and December 2019. Average exposure conditions and EI and DAP valueswere calculated. An acryl phantom was used as a simulated patient, and a VICTOREEN NEROmAx 8000 and ionization chamber were used to calculate the ESD value and entrance surfacedose as a function of kVp and mAs. Sample images for various exposure conditions were obtainedto measure and analyze SNR, PSNR, RMSE, and MAE. Then, simulations were conducted tostudy the characteristics of photon energy and spectrum changes relating to changes in kV. Forthe first head radiograph, the ESD value was 0.611 mGy using average exposure conditions of66 kVp and 12.5 mAs. As kVp and mAs were increased, the DAP and ESD values also increased,while the EI value decreased. When kVp and mAs were decreased, the DAP and ESD valuesalso deceased, while the EI value increased. For the second radiograph, in which there was nodegradation of image quality, the ESD value was 1.596 mGy at 32 mAs and 0.308 mGy at 6.3 mAs(for a fixed kVp), a 5.2-fold difference from the first radiograph. When the value of mAs wasfixed, the ESD value was 1.029 mGy at 81 kVp and 0.314 mGy at 52 kVp, a 3.3-fold difference fromthe first radiograph. The average photon energy was 43.6 keV at 81 kV, 38.1 keV at 66 kV, and 33keV at 52 kV. The air kerma value was 169.8 at 81 kV, 119.8 at 66 kV, and 75.27 at 52 kV, a 2.25-fold difference. The first HVL was 2.868 mmAl at 81 kV, 2.311 mmAl at 66 kV, and 1.832 mmAlat 52 kV, a difference in beam quality of about 1.56-fold. This study confirmed that the exposuredose reached a maximum difference of 5.2-fold without the degradation of image quality andwith changes in kVp and mAs. The first step toward reductions in exposure doses is to educatehospitals that radiation exposure among children is a problem. We also hypothesize that interestin and study of radiation exposure among children will lead to the acquisition of good-qualityimages with reductions in radiation dose.
Abstract
In this study, we analyzed exposure dose and image quality as a function of exposureconditions in head radiography in radiation-sensitive children less than five years old. To compareexposure doses, we used measures of the absorbed dose, such as DAP and ESD. For image quality,the PSNR was used to quantitatively assess noise characteristics. The SRS-78 program was usedto simulate changes in X-ray beam quality in order to determine optimal exposure conditions. The DICOM header information of Infinitt Piview was used to examine skull AP and lateralradiography conditions for head radiographies performed in children less than five years oldbetween January 2018 and December 2019. Average exposure conditions and EI and DAP valueswere calculated. An acryl phantom was used as a simulated patient, and a VICTOREEN NEROmAx 8000 and ionization chamber were used to calculate the ESD value and entrance surfacedose as a function of kVp and mAs. Sample images for various exposure conditions were obtainedto measure and analyze SNR, PSNR, RMSE, and MAE. Then, simulations were conducted tostudy the characteristics of photon energy and spectrum changes relating to changes in kV. Forthe first head radiograph, the ESD value was 0.611 mGy using average exposure conditions of66 kVp and 12.5 mAs. As kVp and mAs were increased, the DAP and ESD values also increased,while the EI value decreased. When kVp and mAs were decreased, the DAP and ESD valuesalso deceased, while the EI value increased. For the second radiograph, in which there was nodegradation of image quality, the ESD value was 1.596 mGy at 32 mAs and 0.308 mGy at 6.3 mAs(for a fixed kVp), a 5.2-fold difference from the first radiograph. When the value of mAs wasfixed, the ESD value was 1.029 mGy at 81 kVp and 0.314 mGy at 52 kVp, a 3.3-fold difference fromthe first radiograph. The average photon energy was 43.6 keV at 81 kV, 38.1 keV at 66 kV, and 33keV at 52 kV. The air kerma value was 169.8 at 81 kV, 119.8 at 66 kV, and 75.27 at 52 kV, a 2.25-fold difference. The first HVL was 2.868 mmAl at 81 kV, 2.311 mmAl at 66 kV, and 1.832 mmAlat 52 kV, a difference in beam quality of about 1.56-fold. This study confirmed that the exposuredose reached a maximum difference of 5.2-fold without the degradation of image quality andwith changes in kVp and mAs. The first step toward reductions in exposure doses is to educatehospitals that radiation exposure among children is a problem. We also hypothesize that interestin and study of radiation exposure among children will lead to the acquisition of good-qualityimages with reductions in radiation dose.
참고문헌 (18)
- Sung-Hyun Choi, "A Study on Evaluating the Image Quality in DR a System with and without a Grid", Journal of the Korean Physical Society, 60(11), pp.1977~1982 (2012), 한국물리학회[DOI]
- Don S, "Radiosensitivity of children : potential for overexposure in CR and DR and magnitude of doses in ordinary radiographic examinations", Pediatr. Radiol, 34, pp.167~172 (2004)
- Patient dose measurement guideline, "FDA", Food and Drug Administration, ~, pp.Radiation Safety Management Series No 14 (2007)
- Food and Drug Administration, Development of the diagnostic reference level of pediatric chest radiography, "FD", Food and Drug Administration, ~ (2009)
- Hashimoto Y, "Channel monitor-based unequal error protection with dynamic OFDM subcarrier assignment for video transmission", IEEE 56th, Vehicular Technology Conference (2002)
- Hennigs SP, "Digital chest radiography with a large-area flat-panel silicon X-ray detector:clinical comparison with conventional radiography", Eur. Radiol., 11(9), pp.688~1696 (2001)
- Huda W, "Relative speeds of Kodak computed radiography phosphors and screen-film systems", Med. Phys, 24(10), pp.1621~1628 (1997)
- Recommendations of the International Commission on Radiological Protection, "ICRP", ICRP, ~, pp.ICRP Publication 60 (1990)
- International Commission on Radiological Protection, Managing Patient Dose in Digital Radiology, "ICRP", ICRP, ~, pp.ICRP publication 93 (2003)
- 정아영, "소아촬영의 방사선 피폭과 저감화 방법", 대한의사협회지, 54(12), pp.1277~1283 (2011), 대한의사협회
- 김광현, "무선 네트워크 환경에서 Wi-Fi CCTV 영상서비스 품질 측정", 정보과학회 컴퓨팅의 실제 논문지, 18(8), pp.607~611 (2012), 한국정보과학회[DOI]
- 김원준, "IPTV 서비스 영상에 대한 객관적 품질측정 방안 연구", 방송공학회 논문지, 13(4), pp.440~451 (2008), 한국방송∙미디어공학회
- Kim YH, "Patient exposure doses from medical x-ray examinations in Korea", J. Radiol. Sci. Technol, 28(4), pp.241~248 (2005)
- Rong XJ, "Comparison of an amorphous silicon", cesium iodide flat-panel digital chest radiography system with screen, Med. Phys, pp.2001, 28(11)[DOI]
- Shu XO, "Diagnostic X-rays and ultrasound exposure and risk of childhood acute lymphoblastic leukemia by immunophenotype", Cancer Epidemiol. Biomarkers Prev, 11(2), pp.177~185 (2002)
- Tanimura M, "Increased risk of hepatoblastoma among immature children with a lower birth weight", Cancer Res, 58(14), pp.3032~3035 (1998)
- United Nations Scientific Committee on the Effects of Atomic Radiation, Source and Effcets of Ionization Radiation, "UNSCEAR", United Nations Scientific Committee on the Effects of Atomic Radiation, ~ (2000)
- 윤오복, "사고로 소아응급실을 내원한 아동에 대한 실태 고찰", 임상간호연구, 15(1), pp.79~91 (2009), 병원간호사회