Development of Fully-Implantable Middle Ear Hearing Device with Differential Floating Mass Transducer : Current Status
2
조회
0
다운로드
5
인용
0
참고문헌
키워드
초록
It is expected that fully-implantable middle-ear hearing devices (FIMEHDs) will soon be available with the advantages of complete concealment, easy surgical implantation, and low power operation to resolve the problems of semi-implantable middle-ear hearing devices (SIMEHDs) such as discomfort of wearing an external device and replacement of battery. Over the last 3 years, a Korean research team at Kyungpook National University has developed an FIMEHD called ACRHS-1 based on a differential floating mass transducer (DFMT). The main research focus was functional improvement, the establishment of easy surgical procedures for implantation, miniaturization, and a low-power operation. Accordingly, this paper reviews the overall system architecture, functions, and experimental results for ACRHS-1 and its related accessories, including a wireless battery charger and remote controller.
Abstract
It is expected that fully-implantable middle-ear hearing devices (FIMEHDs) will soon be available with the advantages of complete concealment, easy surgical implantation, and low power operation to resolve the problems of semi-implantable middle-ear hearing devices (SIMEHDs) such as discomfort of wearing an external device and replacement of battery. Over the last 3 years, a Korean research team at Kyungpook National University has developed an FIMEHD called ACRHS-1 based on a differential floating mass transducer (DFMT). The main research focus was functional improvement, the establishment of easy surgical procedures for implantation, miniaturization, and a low-power operation. Accordingly, this paper reviews the overall system architecture, functions, and experimental results for ACRHS-1 and its related accessories, including a wireless battery charger and remote controller.
참고문헌 (28)
- "Construction and Indication of the Middle Ear Implant Advanced in Audiology", 4, pp.15~21 (1988), Principle
- "The History and Development of the Implantable Hearing Aid", 28, pp.1~16 (1995)
- "Acoustic and Electromagnetic Evaluation of Semi-implantable Middle Ear Hearing Device", 76(5), pp.321~327 (1997)
- "Development of a totally implantable hearing aid", 1, pp.235~238 (2002)
- "An implantable piezoelectric hearing aid transducer for sensorineural hearing loss", 45, pp.792~800 (1997)
- "Optimizing the position and force of a piezoelectric malleus sensor National Institutes of Health", pp.22~24 (1997)
- "Surgical rehabilitation of deafness with partially implantable hearing aid using piezoelectric ceramic bimorpli ossicular vibrator", 24, pp.91~98 (1997)
- "Belgian experience with the Vibrant Soundbridge prosthesis", 56, pp.375~378 (2002)
- "A middle ear implant, the Symphonix Vibrant Soundbridge: Retrospective study of the first 125 patients implanted in France", 24, pp.427~436 (2003)
- "Middle ear implants: A new technology", 54(8), pp.33~48 (2001)
- "A Multicentric experience and clinical results with the Vibrant Soundbridge middle ear implant", 22, pp.952~961 (2001)
- "The effect on the floating mass transducer in the middle ear on hearing sensitivity", 21, pp.42~48 (2000)
- "Long-term Results Using A Piezoelectric Semi-implantable Middle Ear Hearing Device", 34(2), pp.389~400 (2001)
- "Phase 1 clinical trial results of the envoy system:A totally implantable middle ear device for sensorinural hearing loss", pp.904~916 (2004), The American Academy of Otolaryngology, Head and Neck Surgery Foundation Inc
- "Proposal and evaluation of vibration transducer with minimal magnetic field interference for use in IME system by in-vitro experiment IEICE Transactions on electronics", 85-C(6), pp.1374~1377 (2002)
- "Electromagnetic vibration transducer using polyimide elastic body for implantable middle ear hearing aid", pp.201~207 (2002)
- "Design of a Differential Electromagnetic Transducer for Use in IME System", e87-d(5), pp.1231~1237 (2004)
- "Proposal and evaluation of vibration transducer with minimal magnetic field interference for use in IME system by in-vitro experiment IEICE Transactions on electronics", E.85-C(6), pp.1374~1377 (2002)
- "Analysis and Design of Differential Electromagnetic Transducer for Implantable Middle Ear using Finite Element Analysis", 1, pp.796~799 (2005)
- "The effects of skin thickness on microphone function American Journal of Otolaryngology", 11, pp.1~4
- "Design of Implantable Microphone for Artificial Middle Ear System", 26(3), pp.139~144 (2005)
- "Amplification for the hearing-impaired", pp.281~283 (1988)
- "Improved transcutaneous signal coils for detecting two-dimensional dislocation", 35(5), pp.4127~4129 (1999)
- "Transcutaneous DC-DC converter for totally implantable artificial heart using synchronous rectifier IEEE Transactions on Magnetics", 32(5), pp.5118~5120 (1996)
- "Low profile contactless battery charger using planner printed circuit board windings as energy transfer device", 2, pp.23~27 (2002)
- "The Design of an Infrared Transcutaneous Control Unit for Totally Implantable Middle Ear System", 41, pp.71~78 (2004)
- "Inductively-Coupled Power and Data Link for Neural Prostheses using a Class-E Oscillator and FSK Modulation", pp.3376~3379 (2003)
- "Inductive links and drivers for remotely-powered telemetry systems", pp.60~62 (2000)