선형적 위상배열 코일구조의 시뮬레이션을 통한 민감도지도의 공간 해상도 및 필터링 변화에 따른 MR-SENSE 영상재구성 평가
Evaluation of MR-SENSE Reconstruction by Filtering Effect and Spatial Resolution of the Sensitivity Map for the Simulation-Based Linear Coil Array
1
조회
0
다운로드
1
인용
0
참고문헌
키워드
초록
Parallel imaging technique can provide several advantages for a multitude of MRI applications. Especially,in SENSE technique, sensitivity maps were always required in order to determine the reconstruction matrix, therefore,a number of difference approaches using sensitivity information from coils have been demonstrated to improve of image quality. Moreover, many filtering methods were proposed such as adaptive matched filter and nonlinear diffusion technique to optimize the suppression of background noise and to improve of image quality. In this study, we performed SENSE reconstruction using computer simulations to confirm the most suitable method for the feasibility of filtering effect and according to changing order of polynomial fit that were applied on variation of spatial resolution of sensitivity map. The image was obtained at 0.32T(Magfinder II, Genpia, Korea) MRI system using spin-echo pulse sequence(TR/TE = 500/20 ms, FOV = 300 mm, matrix = 128 × 128, thickness = 8 mm). For the simulation, obtained image was multiplied with four linear-array coil sensitivities which were formed of 2D-gaussian distribution and the image was complex white gaussian noise was added. Image processing was separated to apply two methods which were polynomial fitting and filtering according to spatial resolution of sensitivity map and each coil image was subsampled corresponding to reduction factor(r-factor) of 2 and 4. The results were compared to mean value of geomety factor(g-factor) and artifact power(AP) according to r-factor 2 and 4. Our results were represented while changing of spatial resolution of sensitivity map and r-factor, polynomial fit methods were represented the better results compared with general filtering methods. Although our result had limitation of computer simulation study instead of applying to experiment and coil geometric array such as linear, our method may be useful for determination of optimal sensitivity map in a linear coil array.
Abstract
Parallel imaging technique can provide several advantages for a multitude of MRI applications. Especially,in SENSE technique, sensitivity maps were always required in order to determine the reconstruction matrix, therefore,a number of difference approaches using sensitivity information from coils have been demonstrated to improve of image quality. Moreover, many filtering methods were proposed such as adaptive matched filter and nonlinear diffusion technique to optimize the suppression of background noise and to improve of image quality. In this study, we performed SENSE reconstruction using computer simulations to confirm the most suitable method for the feasibility of filtering effect and according to changing order of polynomial fit that were applied on variation of spatial resolution of sensitivity map. The image was obtained at 0.32T(Magfinder II, Genpia, Korea) MRI system using spin-echo pulse sequence(TR/TE = 500/20 ms, FOV = 300 mm, matrix = 128 × 128, thickness = 8 mm). For the simulation, obtained image was multiplied with four linear-array coil sensitivities which were formed of 2D-gaussian distribution and the image was complex white gaussian noise was added. Image processing was separated to apply two methods which were polynomial fitting and filtering according to spatial resolution of sensitivity map and each coil image was subsampled corresponding to reduction factor(r-factor) of 2 and 4. The results were compared to mean value of geomety factor(g-factor) and artifact power(AP) according to r-factor 2 and 4. Our results were represented while changing of spatial resolution of sensitivity map and r-factor, polynomial fit methods were represented the better results compared with general filtering methods. Although our result had limitation of computer simulation study instead of applying to experiment and coil geometric array such as linear, our method may be useful for determination of optimal sensitivity map in a linear coil array.
참고문헌 (12)
- P. Thunberg, "Accuracy and Reproducibility in Phase Contrast Imaging Using SENSE", Mag. Reson. Med, 50(5), pp.1061~1068 (2003)
- K.P. Pruessmann, "SENSE: Sensitivity Encoding for Fast MRI", Mag. Reson. Med, 42(5), pp.952~962 (1999)
- J.Y. Guo, "Evaluation of motion effects on parallel MR imaging with precalibration", Magn. Reson. Imaging, 25(8), pp.1130~1137 (2007)
- D.W. McRobbie, "MRI From Picture to Proton, 2nd ed", pp.137~ (2004), Cambridge Univ. Press
- H. Omer, "A Graphical Generalized Implementation of SENSE Reconstruction Using Matlab", Concepts Magn. Reson. Part A, 36(3), pp.178~186 (2010)
- H.J. Tian, "A Robust Method for Parallel MRI SENSE Reconstruction", Proc. 1st IEEE ICBBE, pp.679~681 (2007)
- J.F. Glockner, "Parallel MR Imaging: A User’s Guide", Radiographics, 25, pp.1279~1297 (2005)
- J.B. Son, "Adaptive SENSE Reconstruction for Parallel Imaging with Massive Array Coils", Proc. 26th IEEE EMBS, pp.1064~1067 (2004)
- S.O. Schoenberg, "Parallel imaging in Clinical MR Applications", pp.49~62 (2007), Springer
- Jim X. Ji, "PULSAR: A MATLAB Toolbox for Parallel Magnetic Resonance Imaging Using Array Coils and Multiple Channel Receivers", Concepts Magn. Reson. Part B, 31(1), pp.24~36 (2007)
- D.J. Larkman, "Beyond the g-Factor Limit in Sensitivity Encoding Using Joint Histogram Entropy", Mag. Reson. Med, 55(1), pp.153~160 (2006)
- James A. Bankson, "Simulation-Based Investigation of Partially Parallel Imaging With a Linear Array at High Accelerations", Mag. Reson. Med, 47(4), pp.777~786 (2002)