표준전자식 혈압계와 수은혈압계, 전자혈압계의 비교 선행연구
Comparison of Sphygmomanometer, Fully Automatic Electronic Blood Pressure Meters with Standard Digital Blood Pressure Monitor : Pilot Study
2
조회
0
다운로드
1
인용
0
참고문헌
키워드
초록
Devices to measure the blood pressure of patients are being used without any calibration in a hospital. It is an important to show consistent values when any medical devices measure the same patients regardless they are sphygmomanometer or fully automatic electronic blood pressure meter. We compared sphygmomanometer and fully automatic electronic blood pressure meters with standard digital blood pressure monitor (SDBPM) to evaluate the consistency of the small healthy subjects. We measured the blood pressure from six healthy subjects (three of 20~40 years and three of 40~60 years old). Two sphygmomanometer and two fully automatic electronic blood pressure meters were used and compared with the SDBPM. Blood pressures measured from right and left arms each and were compared. All six healthy subjects showed normal blood pressure values. In general, left blood pressure values showed higher values than right side. Comparing SDBPM, with the other monitors, the systolic pressure showed ± 34.8% difference and ± 33.3% for the diastolic pressure. Correlation between SDBPM and Sphygmomanometer was 0.59~0.71, and 0.50~0.70 for fully automated digital BP monitors. It fell in grade-D when we apply the BHS(British hypertension society). AAMI(American association for the advancement of medical instrumentation)also showed unsatisfactory results for the mean value (≤ 5 mmHg) and standard deviation (≤ 8 mmHg). We tested sphygmomanometer and fully automatic electronic blood pressure meters and compared with a standard digital blood pressure monitor. All devices showed inconsistent blood pressures. A reliable calibration system is highly needed for all devices in all hospitals.
Abstract
Devices to measure the blood pressure of patients are being used without any calibration in a hospital. It is an important to show consistent values when any medical devices measure the same patients regardless they are sphygmomanometer or fully automatic electronic blood pressure meter. We compared sphygmomanometer and fully automatic electronic blood pressure meters with standard digital blood pressure monitor (SDBPM) to evaluate the consistency of the small healthy subjects. We measured the blood pressure from six healthy subjects (three of 20~40 years and three of 40~60 years old). Two sphygmomanometer and two fully automatic electronic blood pressure meters were used and compared with the SDBPM. Blood pressures measured from right and left arms each and were compared. All six healthy subjects showed normal blood pressure values. In general, left blood pressure values showed higher values than right side. Comparing SDBPM, with the other monitors, the systolic pressure showed ± 34.8% difference and ± 33.3% for the diastolic pressure. Correlation between SDBPM and Sphygmomanometer was 0.59~0.71, and 0.50~0.70 for fully automated digital BP monitors. It fell in grade-D when we apply the BHS(British hypertension society). AAMI(American association for the advancement of medical instrumentation)also showed unsatisfactory results for the mean value (≤ 5 mmHg) and standard deviation (≤ 8 mmHg). We tested sphygmomanometer and fully automatic electronic blood pressure meters and compared with a standard digital blood pressure monitor. All devices showed inconsistent blood pressures. A reliable calibration system is highly needed for all devices in all hospitals.
참고문헌 (10)
- 2009 patient survey In-depth Analysis, "C. G. Son", Ministry of Health and Welfare, ~ (2011)
- K. Kang, "Medical Instrumentation Engineering", pp.395~449 (2012), Yeomungak
- L. A. Taylor, "The application of signal averaging to the NIBP oscillometricwaveform", Proceedings of Annual International Conference of the IEEE EMBS, 1, pp.691~692 (1995)
- C. T. Lin, "Reduction of interference in oscillometric arteiral blood pressure measurement using fuzzy logic", IEEE transactions on Biomedical Engineering, 50(4), pp.432~441 (2003)
- C. Huo, "Blood pressure measurement based on wavelet analysis and fuzzy inference method", Proceedings of the 6th World Congress on Intelligent Control and Automation, 1, pp.4382~4386 (2006)
- T. G. Pickering, "What will replace the mercury sphygmomanometer?", Blood Press Mount, 8, pp.23~25 (2003)
- 이종국, "오므론 HEM 907 XL 자동 진동식 혈압계와 수동 수은혈압계의 오차에 미치는 연령의 영향", 대한고혈압학회지, 17(3), pp.114~124 (2011), 대한고혈압학회
- J. N. Choo, "Clinical Evaluation of the Accuracy of Electronic Home Blood Pressure Measuring Devices", The Korean journal of fundamentals of nursing, 9(1), pp.101~112 (2002)
- E. O'Brien, "Replacing the mercury sphygmomanometer : Requres clinicians to demend better automated devices", Br Med. J, 320, pp.815~816 (2000)
- American national standard for electronic or automated sphygmomanometers, "American association for the advancement of medical instrumentation", ~ (1987)