본문 바로가기
KERD
논문KCI 등재

컬러 카메라를 이용한 측면유동 면역 어세이 정량분석 방법

A Method for Quantitative Measurement of Lateral Flow Immunoassay Using Color Camera

0

조회

0

다운로드

0

인용

0

참고문헌

키워드

초록

Among semi-quantitative or fully quantitative lateral flow assay readers, an image sensor-based instrumenthas been widely used because of its simple setup, cheap sensor price, and compact equipment size. For all previousapproaches, monochrome CCD or CMOS cameras were used for lateral flow assay imaging in which the overallintensities of all colors were taken into consideration to estimate the analyte content, although the analyte relatedcolor information is only limited to a narrow wavelength range. In the present work, we introduced a color CCD cameraas a sensor and a color decomposition method to improve the sensitivity of the quantitative biosensor systemwhich utilizes the lateral flow assay successfully. The proposed setup and image processing method were applied toachieve the quantification of imitatively dispensed particles on the surface of a porous membrane first, and the measurementresult was then compared with that using a monochrome CCD. The compensation method was proposedin different illumination conditions. Eventually, the color decomposition method was introduced to the commerciallyavailable lateral flow immunochromatographic assay for the diagnosis of myocardial infarction. The measurement sensitivityutilizing the color image sensor is significantly improved since the slopes of the linear curve fit are enhancedfrom 0.0026 to 0.0040 and from 0.0802 to 0.1141 for myoglobin and creatine kinase (CK)-MB detection, respectively.

Abstract

Among semi-quantitative or fully quantitative lateral flow assay readers, an image sensor-based instrumenthas been widely used because of its simple setup, cheap sensor price, and compact equipment size. For all previousapproaches, monochrome CCD or CMOS cameras were used for lateral flow assay imaging in which the overallintensities of all colors were taken into consideration to estimate the analyte content, although the analyte relatedcolor information is only limited to a narrow wavelength range. In the present work, we introduced a color CCD cameraas a sensor and a color decomposition method to improve the sensitivity of the quantitative biosensor systemwhich utilizes the lateral flow assay successfully. The proposed setup and image processing method were applied toachieve the quantification of imitatively dispensed particles on the surface of a porous membrane first, and the measurementresult was then compared with that using a monochrome CCD. The compensation method was proposedin different illumination conditions. Eventually, the color decomposition method was introduced to the commerciallyavailable lateral flow immunochromatographic assay for the diagnosis of myocardial infarction. The measurement sensitivityutilizing the color image sensor is significantly improved since the slopes of the linear curve fit are enhancedfrom 0.0026 to 0.0040 and from 0.0802 to 0.1141 for myoglobin and creatine kinase (CK)-MB detection, respectively.

참고문헌 (19)

  1. Plotz CM., "The latex fixation test. Application to the serologic diagnosis of rheumatoid arthritis", Am J Med, 21(6), pp.888~892 (1956)
  2. Posthuma-Trumpie GA., "Lateral flow(immuno)assay : its strengths, weaknesses, opportunities and threats. A literature survey", Anal Bioanal Chem, 393(2), pp.569~582 (2009)
  3. Wong R, "Lateral Flow Immunoassay", pp.3~5 (2010), Humana Press
  4. K. Taton., "Lateral flow immunoassay using magnetoresistive sensor", J Magn Mater, 321(10), pp.1679~1682 (2009)
  5. Li Z., "Rapid and sensitive detection of protein biomarker using a portable fluorescence biosensor based on quantum dots and a lateral flow test strip", Anal Chem, 82(16), pp.7008~7014 (2010)
  6. Tang D., "Magnetic nanogold microspheres-based lateral-flow immunodipstick for rapid detection of aflatoxin B2 in food", Biosens Bioelectron, 25(2), pp.514~518 (2009)
  7. Xu Q., "Development of lateral flow immunoassay system based on superparamagnetic nanobeads as labels for rapid quantitative detection of cardiac troponin I", Mat Sci Eng C-Bio S, 29(3), pp.702~707 (2009)
  8. Wong R, "Lateral Flow Immunoassay", pp.23~25 (2010), Humana Press
  9. Park J, "Diagnosis apparatus and method used in-vitro diagnostics", Rep. of Korea Patent 1011579990000, ~ (2010)
  10. Bayer BE, "Color imaging array", U.S. Patent 3,971,065, 065~ (1976)
  11. 서동균, "상이한 스펙트럼을 가지는 객체간의 RGB 색상 차이를 최대화하기 위한 최적조명", 의공학회지, 30(3), pp.263~269 (2009), 대한의용생체공학회
  12. 한병관, "피부병변의 정량적 평가를 위한 디지털 컬러 영상 시스템", 의공학회지, 28(2), pp.195~198 (2007), 대한의용생체공학회
  13. Plebani M., "Cardiac markers : present and future", Int J Clin Lab Res, 29(2), pp.56~63 (1999)
  14. Montague C., "Myoglobin in the early evaluation of acute chest pain", Am J Clin Pathol, 104(4), pp.472~476 (1995)
  15. Stone MJ., "Serum myoglobin level as diagnostic test in patients with acute myocardial infarction", Br Heart J, 39(4), pp.375~380 (1977)
  16. Bessman SP., "The creatine-creatine phosphate energy shuttle", Ann. Rev Biochem, 54, pp.831~862 (1985)

관련 논문AI추천

이 간행물의 다른 논문 (1건)