무선 캡슐내시경 조종을 위한 머리부착형 디스플레이 인터페이스 시스템의 개발
Development of Head Mounted Display Interface System for Controlling Wireless Capsule Endoscope
0
조회
0
다운로드
0
인용
0
참고문헌
키워드
초록
The present study proposed a new interface system for capsule endoscopy by using head mounted display (HMD) device, which can control the orientation of the capsule endoscope with electromagnetic actuator (EMA) sys- tem. The orientation information of the HMD user was detected by the gyroscope sensor built into the device and then calculated to as an angle increment using Unity Engine compiler. The measured angle changes from the HMD were converted to the current values of the corresponding coils to be changed in the EMA system. Two experiments were designed to measure the accuracy and the intuitiveness of the HMD interface system. In the angle accuracy measurement, the capsule endoscope driven by HMD interface system showed the averaged errors of 0.68 degrees horizontally and 1.001 degrees vertically for given test angles. In the intuitiveness measurement, HMD interface sys- tem showed 1.33 times faster manipulation speed rather than the joystick interface system. In this respect, the HMD interface system for capsule endoscopy was expected to improve the overall diagnostic environment while main- taining comfort of patients and clinicians.
Abstract
The present study proposed a new interface system for capsule endoscopy by using head mounted display (HMD) device, which can control the orientation of the capsule endoscope with electromagnetic actuator (EMA) sys- tem. The orientation information of the HMD user was detected by the gyroscope sensor built into the device and then calculated to as an angle increment using Unity Engine compiler. The measured angle changes from the HMD were converted to the current values of the corresponding coils to be changed in the EMA system. Two experiments were designed to measure the accuracy and the intuitiveness of the HMD interface system. In the angle accuracy measurement, the capsule endoscope driven by HMD interface system showed the averaged errors of 0.68 degrees horizontally and 1.001 degrees vertically for given test angles. In the intuitiveness measurement, HMD interface sys- tem showed 1.33 times faster manipulation speed rather than the joystick interface system. In this respect, the HMD interface system for capsule endoscopy was expected to improve the overall diagnostic environment while main- taining comfort of patients and clinicians.
참고문헌 (16)
- Yea Som Lee, "Immersive Gesture Interface Design for HMD Based Vir-tual World Navigation", 한국HCI학회 학술대회, pp.9~14 (2016)
- Jinah Park, "Medical Simulation : Virtual Surger", Com-munications of the Korean Institute of Information Scientists and Engineers, 23(10), pp.49~52 (2005)
- Rafa Rahman, "Head-Mounted Display Use in Surgery: A Systematic Review", Surgical Innovation, 27(1), pp.88~100 (2019), SAGE Publications[DOI]
- Viet Ha Le, "Miniaturized biopsy module using gripper tool for active locomotive capsule endoscope", J Mechatronics, 44, pp.52~59 (2017)
- Lee C, "3-D Locomotion of Mag-netic Capsule Endoscope using Electromagnetic Field Con-trol", 한국정밀공학회 학술발표대회 논문집, pp.379~380 (2015)
- Maithel SK, "Simulated laparoscopy using a head-mounted display vs traditional video monitor: an assessment of performance and muscle fatigue", Surgical Endoscopy And Other Inter-ventional Techniques, 19(3), pp.406~411 (2005)
- Gastone Ciuti, "A comparative evaluation of control interfaces for a robotic-aided endoscopic capsule platform", IEEE Trans Robot, 28, pp.534~538 (2012)
- Cheong Lee, "Active Locomotive Intestinal Capsule Endoscope (ALICE) System: A Prospective Feasibility Study", IEEE, 2015, pp.20(5), 2067~2074[DOI]
- Jeon SM, "Magnetic navigation system for the precise helical and translational motions of a microrobot in human blood vessels", J Appl Phys, 111(7), pp.07E702.~ (2012)
- Iddan G, "Wireless Cap-sule Endoscopy", Nature, 405, pp.417~ (2000)
- Kazushi Ishiyama, "Magnetic Actuator for a Capsule Endoscope Navigation System", Journal of Magnetics, 12(2), pp.89~92 (2007), 한국자기학회
- James Brooks, "Comparison of a Head-Mounted Display and Flat Screen Display during a Micro-UAV Target Detection Task", Proceedings of the Human Factors and Ergo-nomics Society, pp.1514~1518 (2017)
- Philipp Tiefenbacher, "Comparison of Mobile Touch Interfaces for Object Identifi-cation and Troubleshooting Tasks in Augmented Reality", IEEE Virtual Reality (2016)
- Henrik Keller, "Method for navigation and control of a magnetically guided capsule endoscope in the human stomach", IEEE RAS & EMBS International Conference on Biomedical Robotics and Bio-mechatronics, pp.859~865 (2012)
- Cheong Lee, "3-D Helical Motion of Electromagnetic Actuated Capsule Endoscope for Diagnosis of Tubular Digestive Organs", 대한기계학회 춘추학술대회, pp.2464~2424 (2014)