Evaluating the Diagnostic Performance of Two Rapid Influenza Diagnostic Tests

1

조회

0

다운로드

0

인용

0

참고문헌

키워드

초록

Rapid influenza diagnostic test (RIDT) is widely used for the diagnosis of influenza owing to its simplicity and convenience of use. This study aimed to evaluate the performance of a new RIDT, SD Standard F influenza A/B FIA (SD Biosensor, Inc., Korea) (Standard F) and compare its performance with BD Veritor Flu A+B (Veritor), using the results of real-time reverse transcription PCR (rRT-PCR) analysis as the standard for reference. On comparing the results obtained from both the RIDTs and rRT-PCR qualitatively, it was found that the Veritor and Standard F assays have the sensitivity of 65.6% (21/32) and 71.9% (23/32), respectively, for the detection of influenza A with a specificity of 100.0% (68/68). Additionally, both the assays demonstrated a sensitivity of 66.7% (12/18) and specificity of 100.0% (68/68) for the detection of influenza B. The cutoff index (COI) value of the fluorescence color intensity from the Standard F assay, displayed on the device along with the qualitative results, indicated a negative correlation with the Ct value from rRT-PCR for both influenza A and B (P<0.001). The sensitivity of the new RIDT for the detection of influenza was comparable with that of the Veritor assay and the new RIDT could be used as a substitute for existing RIDTs by providing additional information to predict the approximate viral burden of influenza.

Abstract

Rapid influenza diagnostic test (RIDT) is widely used for the diagnosis of influenza owing to its simplicity and convenience of use. This study aimed to evaluate the performance of a new RIDT, SD Standard F influenza A/B FIA (SD Biosensor, Inc., Korea) (Standard F) and compare its performance with BD Veritor Flu A+B (Veritor), using the results of real-time reverse transcription PCR (rRT-PCR) analysis as the standard for reference. On comparing the results obtained from both the RIDTs and rRT-PCR qualitatively, it was found that the Veritor and Standard F assays have the sensitivity of 65.6% (21/32) and 71.9% (23/32), respectively, for the detection of influenza A with a specificity of 100.0% (68/68). Additionally, both the assays demonstrated a sensitivity of 66.7% (12/18) and specificity of 100.0% (68/68) for the detection of influenza B. The cutoff index (COI) value of the fluorescence color intensity from the Standard F assay, displayed on the device along with the qualitative results, indicated a negative correlation with the Ct value from rRT-PCR for both influenza A and B (P<0.001). The sensitivity of the new RIDT for the detection of influenza was comparable with that of the Veritor assay and the new RIDT could be used as a substitute for existing RIDTs by providing additional information to predict the approximate viral burden of influenza.

참고문헌 (20)

  1. Nicholson KG, "Influenza", Lancet, 362, pp.1733~1745 (2003)
  2. Merckx J, "Diagnostic accuracy of novel and traditional rapid tests for influenza infection compared with reverse transcriptase polymerase chain reaction : A systematic review and meta-analysis", Ann Intern Med, 167, pp.394~409 (2017)
  3. "Information for clinicians on influenza virus testing"[URL]
  4. Tsushima Y, "Quantitative RT-PCR evaluation of a rapid influenza antigen test for efficient diagnosis of influenza virus infection", J Virol Methods, 212, pp.76~79 (2015)
  5. Chartrand C, "Accuracy of rapid influenza diagnostic tests : a meta-analysis", Ann Intern Med, 156, pp.500~511 (2012)
  6. "WHO recommendations on the use of rapid testing for influenza diagnosis"[URL]
  7. 박설희, "새로운 인플루엔자 신속항원검사 BD Veritor System의 평가", Annals of Clinical Microbiology, 18(1), pp.27~32 (2015), 대한임상미생물학회
  8. Yeo SJ, "Improvement of a rapid diagnostic application of monoclonal antibodies against avian influenza H7 subtype virus using Europium nanoparticles", Sci Rep, 7, pp.7933~ (2017)
  9. Peaper DR, "Rapid diagnosis of influenza : state of the art", Clin Lab Med, 34, pp.365~385 (2014)
  10. Webster RG, "Evolution and ecology of influenza A viruses", Microbiol Mol Biol Rev, 56, pp.152~179 (1992)
  11. 김재석, "Performance Evaluation of the Automated Fluorescent Immunoassay System Rotavirus Assay in Clinical Samples", Annals of Laboratory Medicine, 39(1), pp.50~57 (2019), 대한진단검사의학회[DOI]
  12. 류지형, "Development of a Rapid Automated Fluorescent Lateral Flow Immunoassay to Detect Hepatitis B Surface Antigen (HBsAg), Antibody to HBsAg, and Antibody to Hepatitis C", Annals of Laboratory Medicine, 38(6), pp.578~584 (2018), 대한진단검사의학회
  13. Yoo Y, "Clinical evaluation of the SD Bioline influenza virus antigen test for rapid detection of influenza viruses A and B in children and adults during the influenza season", Clin Vaccine Immunol, 14, pp.1050~1052 (2007)
  14. Ryu SW, "Comparison of three rapid influenza diagnostic tests with digital readout systems and one conventional rapid influenza diagnostic test", J Clin Lab Anal, 32, pp.e22234~ (2018)
  15. Self WH, "Diagnostic performance of the BinaxNow Influenza A&B rapid antigen test in ED patients", Am J Emerg Med, 30, pp.1955~1961 (2012)
  16. Mese S, "Analytical performance of the BD Veritor™ System for rapid detection of influenza virus A and B in a primary healthcare setting", BMC Infect Dis, 16, pp.481~ (2016)
  17. Hassan F, "Comparison of the BD Veritor™ System for Flu A+B with the Alere BinaxNOW® Influenza A&B card for detection of influenza A and B viruses in respiratory specimens from pediatric patients", J Clin Microbiol, 52, pp.906~910 (2014)
  18. Ryu SW, "Comparison of two new generation influenza rapid diagnostic tests with instrument-based digital readout systems for influenza virus detection", Br J Biomed Sci, 73, pp.115~120 (2016)
  19. Dunn J, "Prompt detection of influenza A and B viruses using the BD Veritor™ System Flu A+B, Quidel® Sofia® Influenza A+B FIA, and Alere BinaxNOW® Influenza A&B compared to real-time reverse transcription-polymerase chain reaction(RT-PCR)", Diagn Microbiol Infect Dis, 79, pp.10~13 (2014)
  20. Selove W, "Performance of rapid SOFIA Influenza A+B test compared to Luminex x-TAG respiratory viral panel assay in the diagnosis of influenza A, B, and subtype H3", J Investig Med, 64, pp.905~907 (2016)

관련 논문AI추천

이 간행물의 다른 논문 (11건)