Evaluation of Automated Platelet Aggregation Test Using a Sysmex CS-5100 Analyzer

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

초록

Background: The platelet aggregation test is widely used to measure antiplatelet therapy response and to detect platelet function disorders. CS-5100 (Sysmex Co., Japan) is a recently introduced coagulation analyzer that can also measure platelet aggregation. We evaluated the performance of CS-5100 in the platelet aggregation test for use in clinical laboratories. Methods: We investigated the precision, stability, dilution test, and correlation of CS-5100 with a traditional aggregometer. Precision was tested using normal and patient samples. Stability was assessed over 5 different time points for 8 hours. The dilution test was performed with normal samples using ADP agonists. We tested correlations between the results of Chrono-log aggregometer (Chrono-Log Co., U.S.A) and CS-5100 using 10 samples with 5 agonists each. We also calculated the reference range of 5 agonists using 22-30 normal samples. Results: The coefficients of variation using normal samples were 7.45% and 3.27% for ADP and arachidonic acid, respectively. Stability was maintained for up to 2 hours in most samples. Dilution tests showed similar results until reaching a dilution factor of 2. Strong correlations of the results between Chrono-log and CS-5100 were found, except for ristocetin. The reference ranges of 5 reagents in CS-5100 were similar to those obtained with the Chrono-log aggregometer. Conclusions: The performance of CS-5100 in platelet aggregation tests showed reliable results compared to a traditional aggregometer. CS5100 can perform coagulation test and platelet aggregation test, simultaneously. Thus, CS-5100 can enable cost saving and reduce turn-aroundtime by reducing the inspection time.

Abstract

Background: The platelet aggregation test is widely used to measure antiplatelet therapy response and to detect platelet function disorders. CS-5100 (Sysmex Co., Japan) is a recently introduced coagulation analyzer that can also measure platelet aggregation. We evaluated the performance of CS-5100 in the platelet aggregation test for use in clinical laboratories. Methods: We investigated the precision, stability, dilution test, and correlation of CS-5100 with a traditional aggregometer. Precision was tested using normal and patient samples. Stability was assessed over 5 different time points for 8 hours. The dilution test was performed with normal samples using ADP agonists. We tested correlations between the results of Chrono-log aggregometer (Chrono-Log Co., U.S.A) and CS-5100 using 10 samples with 5 agonists each. We also calculated the reference range of 5 agonists using 22-30 normal samples. Results: The coefficients of variation using normal samples were 7.45% and 3.27% for ADP and arachidonic acid, respectively. Stability was maintained for up to 2 hours in most samples. Dilution tests showed similar results until reaching a dilution factor of 2. Strong correlations of the results between Chrono-log and CS-5100 were found, except for ristocetin. The reference ranges of 5 reagents in CS-5100 were similar to those obtained with the Chrono-log aggregometer. Conclusions: The performance of CS-5100 in platelet aggregation tests showed reliable results compared to a traditional aggregometer. CS5100 can perform coagulation test and platelet aggregation test, simultaneously. Thus, CS-5100 can enable cost saving and reduce turn-aroundtime by reducing the inspection time.

참고문헌 (17)

  1. Gale AJ, "Continuing education course #2 : Current understanding of hemostasis", Toxicol Pathol, 39, pp.273~280 (2011)
  2. Nurden AT, "Congenital platelet disorders and understanding of platelet function", Br J Haematol, 165, pp.165~178 (2014)
  3. Konkle BA, "Acquired disorders of platelet function", Hematology Am Soc Hematol Educ Program, 2011, pp.391~396 (2011)
  4. Born GV, "The aggregation of blood platelets", J Physiol, 168, pp.178~195 (1963)
  5. Harrison P, "Platelet function analysis", Blood Rev, 19, pp.111~123 (2005)
  6. Cattaneo M, "Results of a worldwide survey on the assessment of platelet function by light transmission aggregometry : a report from the platelet physiology subcommittee of the SSC of the ISTH", J Thromb Haemost, 7, pp.1029~ (2009)
  7. Sakayori T, "Introduction and evaluation of light transmission platelet aggregation method on the Sysmex CS-series automated coagulation analyzer", Sysmex Journal International, 26, pp.1~10 (2016)
  8. Cattaneo M, "Recommendations for the standardization of light transmission aggregometry : a consensus of the working party from the platelet physiology subcommittee of SSC", ISTH, 2013, pp.11, 1183~1189
  9. Clinical and Laboratory Standard Institute, "Evaluation of precision of quantitative measurement procedures; Approved guideline" (2014), Clinical and Laboratory Standard Institute
  10. Clinical and Laboratory Standard Institute, "Measurement procedure comparison and bias estimation using patient samples; Approved guideline" (2013), Clinical and Laboratory Standard Institute
  11. Clinical and Laboratory Standard Institute, "Defining, establishing, and verifying reference intervals in the clinical laboratory; Approved guideline" (2010), Clinical and Laboratory Standard Institute
  12. Taylor R, "Interpretation of the correlation coefficient : a basic review", J Diagn Med Sonogr, 6, pp.35~39 (1990)
  13. Mukaka MM, "Statistics corner : A guide to appropriate use of correlation coefficient in medical research", Malawi Med J, 24, pp.69~71 (2012)
  14. Hayward CP, "An evaluation of methods for determining reference intervals for light transmission platelet aggregation tests on samples with normal or reduced platelet counts", Thromb Haemost, 100, pp.134~145 (2008)
  15. Miller CH, "Gender, race and diet affect platelet function tests in normal subjects, contributing to a high rate of abnormal results", Br J Haematol, 165, pp.842~853 (2014)
  16. Ling LQ, "Evaluation of an automated light transmission aggregometry", Platelets, 28, pp.712~719 (2017)
  17. Lawrie AS, "The automation of routine light transmission platelet aggregation", Int J Lab Hematol, 36, pp.431~438 (2014)