본문 바로가기
KERD
논문KCI 등재

Evaluation of Test Standardization for 25-OH Vitamin D Immunoassays

2

조회

0

다운로드

1

인용

0

참고문헌

키워드

초록

Background: Accurately measuring 25-OH vitamin D (25-OH-D) is important for monitoring vitamin D status. Despite the efforts of all involved parties, achieving standardization remains challenging. The present study assessed the standardization status of commercial immunoassays for measuring 25-OH-D. Methods: The diagnostic accuracy of commercial immunoassays Roche Elecsys Vitamin D Total, Siemens Atellica IM Vitamin D Total, Abbott Alinity 25-OH Vitamin D, and Beckman Coulter Access 25-OH Vitamin D Total was evaluated by comparing the quantitative values and qualitative decisions for 150 residual serums with various concentrations obtained from each test to the results determined using mass spectrometry (MS). Then, correlation and concordance were estimated. Results: When the measured 25-OH-D values from each assay were compared to those determined using MS ranging from 5.86 to 67.75 ng/mL, no evaluated immunoassay was equivalent to MS by showing slope and intercept not containing 1 and 0 in their 95% confidence interval, respectively. Regression estimates at the medical decision limits of 10, 20, and 30 ng/mL showed that only 30 ng/mL of the Beckman Coulter test was within the acceptable bias limit of 5.0% recognized by the VDSCP. The concordance of decisions with MS ranged from 90.0 to 94.0% in immunoassays. Conclusions: Despite achieving the certification by the VDSCP, most immunoassays differed significantly regarding acceptance criteria in measuring 25-OH-D, which may affect clinical decision-making. Therefore, attention should be paid to interpreting the results, and further efforts for test standardization are required.

Abstract

Background: Accurately measuring 25-OH vitamin D (25-OH-D) is important for monitoring vitamin D status. Despite the efforts of all involved parties, achieving standardization remains challenging. The present study assessed the standardization status of commercial immunoassays for measuring 25-OH-D. Methods: The diagnostic accuracy of commercial immunoassays Roche Elecsys Vitamin D Total, Siemens Atellica IM Vitamin D Total, Abbott Alinity 25-OH Vitamin D, and Beckman Coulter Access 25-OH Vitamin D Total was evaluated by comparing the quantitative values and qualitative decisions for 150 residual serums with various concentrations obtained from each test to the results determined using mass spectrometry (MS). Then, correlation and concordance were estimated. Results: When the measured 25-OH-D values from each assay were compared to those determined using MS ranging from 5.86 to 67.75 ng/mL, no evaluated immunoassay was equivalent to MS by showing slope and intercept not containing 1 and 0 in their 95% confidence interval, respectively. Regression estimates at the medical decision limits of 10, 20, and 30 ng/mL showed that only 30 ng/mL of the Beckman Coulter test was within the acceptable bias limit of 5.0% recognized by the VDSCP. The concordance of decisions with MS ranged from 90.0 to 94.0% in immunoassays. Conclusions: Despite achieving the certification by the VDSCP, most immunoassays differed significantly regarding acceptance criteria in measuring 25-OH-D, which may affect clinical decision-making. Therefore, attention should be paid to interpreting the results, and further efforts for test standardization are required.

참고문헌 (19)

  1. Audran M, "The physiology and pathophysiology of vitamin D", Mayo Clin Proc, 60, pp.851~866 (1985)
  2. Holick MF, "Vitamin D deficiency", N Engl J Med, 357, pp.266~281 (2007)
  3. Hossein-nezhad A, "Vitamin D for health : a global perspective", Mayo Clin Proc, 88, pp.720~755 (2013)
  4. Smit E, "The effect of vitamin D and frailty on mortality among non-institutionalized US older adults", Eur J Clin Nutr, 66, pp.1024~1028 (2012)
  5. "CDC Vitamin D Standardization-Certification Program (VDSCP)-Total 25-hydroxyvitamin D Certified Procedures" (2020)[URL]
  6. Binkley N, "Standardizing vitamin D assays : the way forward", J Bone Miner Res, 29, pp.1709~1714 (2014)
  7. Benton SC, "Evaluation of the 25-hydroxy vitamin D assay on a fully automated liquid chromatography mass spectrometry system, the Thermo Scientific Cascadion SM Clinical Analyzer with the Cascadion 25-hydroxy vitamin D assay in a routine clinical laboratory", Clin Chem Lab Med, 58, pp.1010~1017 (2020)
  8. Lippi G, "Multicenter comparison of seven 25OH vitamin D Automated Immunoassays", J Med Biochem, 34, pp.344~350 (2015)
  9. Moreau E, "Performance characteristics of the VIDAS® 25-OH Vitamin D Total assay-comparison with four immunoassays and two liquid chromatography-tandem mass spectrometry methods in a multicentric study", Clin Chem Lab Med, 54, pp.45~53 (2016)
  10. Sempos CT, "25-Hydroxyvitamin D assay standardisation and vitamin D guidelines paralysis", Public Health Nutr, 23, pp.1153~1164 (2020)
  11. Clinical and Laboratory Standards Institute, "Measurement procedure comparison and bias estimation using patient samples; Approved guideline-" (2013), Clinical and Laboratory Standards Institute
  12. Phinney KW, "Value assignment of vitamin D metabolites in vitamin D standardization program serum samples", J AOAC Int, 100, pp.1253~1259 (2017)
  13. Phinney KW, "Development and certification of a standard reference material for vitamin D metabolites in human serum", Anal Chem, 84, pp.956~962 (2012)
  14. Kim H, "Measurement of serum levels of 25-hydroxyvitamin D3 and 25-hydroxyvitamin D2 using Diels-Alder derivatization and ultra-performance liquid chromatography-tandem mass spectrometry", Lab Med Online, 2, pp.188~196 (2012)
  15. Sempos CT, "Vitamin D assays and the definition of hypovitaminosis D : results from the First International Conference on Controversies in Vitamin D", Br J Clin Pharmacol, 84, pp.2194~2207 (2018)
  16. Nah EH, "Vitamin D levels and prevalence of vitamin D deficiency associated with sex, age, region, and season in Koreans", Lab Med Online, 5, pp.84~91 (2015)
  17. Carter GD, "25-Hydroxyvitamin D assays : potential interference from other circulating vitamin D metabolites", J Steroid Biochem Mol Biol, 164, pp.134~138 (2016)
  18. "EFLM Biological Variation Database" (2023)[URL]
  19. Bjerg LN, "An evaluation of total 25-hydroxyvitamin D assay standardization : where are we today?", J Steroid Biochem Mol Biol, 190, pp.224~233 (2019)

관련 논문AI추천

Discrepancy between Vitamin D Total Immunoassays due to Various Cross-reactivities권오주, 박애자, 이준형, 최지혜대한골대사학회 · 대한골대사학회지2015 · 제22권 제3호Comparison of Three Commercially Available Assays for Measurement of Vitamin D양다해, 황현용대한진단검사의학회 · Laboratory Medicine Online2017 · 제7권 제3호Efficacy of the Measurement of 25-Hydroxyvitamin D2 and D3 Levels by Using PerkinElmer Liquid Chromatography-Tandem Mass Spectrometry Vitamin D Kit Compared With DiaSorin Radioimmunoassay Kit and Elecsys Vitamin D Total Assay곽호석, 구은숙, 박미현, 박은정, 오한진, 정희정, 조동희대한진단검사의학회 · Annals of Laboratory Medicine2015 · 제35권 제2호Six Sigma Analysis of Vitamin D Measurement Using External Quality Assessment ProgramMyungsuk Ji대한임상검사과학회 · 대한임상검사과학회지2020 · 제52권 제2호Clinical Utility of Measurement of Vitamin D-Binding Protein and Calculation of Bioavailable Vitamin D in Assessment of Vitamin D Status김현정, 문희원, 송정한, 윤여민, 지미숙, 허미나대한진단검사의학회 · Annals of Laboratory Medicine2017 · 제37권 제1호Comparison of Six Automated Immunoassays With Isotope-Diluted Liquid Chromatography–Tandem Mass Spectrometry for Total Thyroxine MeasurementChuanbao Zhang, Honglei Li, Jun Lu, Li’an Hou, Ling Qiu, Qian Cheng, Qinghui Meng, Shaowei Xie, Songlin Yu, Weiyan Zhou, Xinqi Cheng대한진단검사의학회 · Annals of Laboratory Medicine2019 · 제39권 제4호

이 간행물의 다른 논문 (11건)