소형 GC 모듈의 개발 및 특성
Development and Characteristic Study of a Portable Gas Chromatography
1
조회
0
다운로드
1
인용
0
참고문헌
키워드
초록
In the present study, we developed a portable GC module for real-time, quantitative determinations of gas mixtures in air sample. Capillary or packed column was coiled together with a heater wire and thermocouple in a small case. Together with the small and light weight sensors and valves as well as the rechargeable carrier gas canister, which permits collection and separation of samples, this system can determine the components of complex mixtures of air contaminants at low concentrations with a duty cycle of 10 min. When measured the various samples with a FID and TCD, the system showed,for a capillary column, a good resolution (R=8.3), high sensitivity, reproducibility, and linear dynamic range greater than three orders of magnitude. These results indicate that the portable GC module is expected to be used for a wide range of applications,particularly for in situ environmental monitoring, chemical processes, and regulation of contaminant emission.
Abstract
In the present study, we developed a portable GC module for real-time, quantitative determinations of gas mixtures in air sample. Capillary or packed column was coiled together with a heater wire and thermocouple in a small case. Together with the small and light weight sensors and valves as well as the rechargeable carrier gas canister, which permits collection and separation of samples, this system can determine the components of complex mixtures of air contaminants at low concentrations with a duty cycle of 10 min. When measured the various samples with a FID and TCD, the system showed,for a capillary column, a good resolution (R=8.3), high sensitivity, reproducibility, and linear dynamic range greater than three orders of magnitude. These results indicate that the portable GC module is expected to be used for a wide range of applications,particularly for in situ environmental monitoring, chemical processes, and regulation of contaminant emission.
참고문헌 (12)
- Hodgson, M., , Occup. Med, 15, pp.571~ (2000)
- Smedbold, H. T, , Indoor Air, 11, pp.23~ (2001)
- Wolkoff, P., , Indoor Air, (3), pp.9~73 (1995)
- J. J. Isbell, , J. Am. Soc. Mass Spectrom, 7, pp.565~ (1996)
- Burton, B., "Volatile Organic Compounds In Indoor Air Pollution and Health", pp.127~153 (1997), Marcel Dekker
- Peng, C. Y., , J. Environ. Monit, 2, pp.213~ (2000)
- Wahl, J. H., , J. Sep. Sci, 26, pp.1083~ (2003)
- Smith, P. A., , J. Chromatogr. A, 1067, pp.285~ (2005)
- Agah, M., , J. Microelectromech. Syst, 14, pp.1039~ (2005)
- Whiting, J. J., , J. Sep. Sci, 29, pp.218~ (2006)
- Kim, M. G., , J. Korean Chem. Soc, 38, pp.151~ (1994)
- Lee, J.-H., , Bull. Korean Chem. Soc, 23, pp.488~ (2002)