Comparison of Two Clinical Devices for the Measurement of Distortion Product Otoacoustic Emissions in Normal-Hearing Adults

0

조회

0

다운로드

0

인용

0

참고문헌

키워드

초록

Background and Objectives: Otoacoustic emissions (OAEs) are low-intensity sounds generated by the cochlea and associated with the function of the outer hair cells. Since OAE measurements do not require active participation of a listener, OAEs are considered an objective measure of cochlear function. While distortion-product OAEs (DPOAEs) are commonly used in clinical practice, limited information is available on the performance of various clinical devices. This study compared two commercial clinical devices, Titan and Eclipse from Interacoustics, and collected normative data for DPOAEs in people with normal hearing. The data collection and analysis were focused on signal and noise amplitudes as well as signal-to-noise ratios (SNRs).Subjects and Methods: Sixty-three participants with normal hearing (age 23.2±2.0 years) were included in the study. DPOAEs were measured at 31 frequencies ranging from 500 to 10,000 Hz.Results: DPOAE amplitude differed across frequencies. Additionally, a high number of unreliable responses were observed at the edge frequencies (<828 Hz and >6,072 Hz). There were no significant differences between the two devices in terms of DPOAE amplitudes, but the recorded-noise levels (amplitude) differed significantly. SNRs of the recorded OAEs were frequency dependent (higher frequencies showed larger SNRs), and significant differences in terms of SNR were found between the two devices.Conclusions: Despite the above-mentioned differences, the OAEs recorded with both devices met the pass criteria for the SNR (≥6 dB) consistently across frequencies, and thus the differences do not compromise the test outcomes. The frequency dependence of the OAE amplitudes and the corresponding SNRs may be relevant for clinical practice.

Abstract

Background and Objectives: Otoacoustic emissions (OAEs) are low-intensity sounds generated by the cochlea and associated with the function of the outer hair cells. Since OAE measurements do not require active participation of a listener, OAEs are considered an objective measure of cochlear function. While distortion-product OAEs (DPOAEs) are commonly used in clinical practice, limited information is available on the performance of various clinical devices. This study compared two commercial clinical devices, Titan and Eclipse from Interacoustics, and collected normative data for DPOAEs in people with normal hearing. The data collection and analysis were focused on signal and noise amplitudes as well as signal-to-noise ratios (SNRs).Subjects and Methods: Sixty-three participants with normal hearing (age 23.2±2.0 years) were included in the study. DPOAEs were measured at 31 frequencies ranging from 500 to 10,000 Hz.Results: DPOAE amplitude differed across frequencies. Additionally, a high number of unreliable responses were observed at the edge frequencies (<828 Hz and >6,072 Hz). There were no significant differences between the two devices in terms of DPOAE amplitudes, but the recorded-noise levels (amplitude) differed significantly. SNRs of the recorded OAEs were frequency dependent (higher frequencies showed larger SNRs), and significant differences in terms of SNR were found between the two devices.Conclusions: Despite the above-mentioned differences, the OAEs recorded with both devices met the pass criteria for the SNR (≥6 dB) consistently across frequencies, and thus the differences do not compromise the test outcomes. The frequency dependence of the OAE amplitudes and the corresponding SNRs may be relevant for clinical practice.

참고문헌 (15)

  1. Prieve P, "Handbook of Clinical Audiology", pp.357~349 (2015), Wolters Kluwer Health
  2. Lonsbury-Martin BL, "Handbook of Clinical Neurophysiology", pp.115~135 (2013), Elsevier
  3. Dhar S, "Otoacoustic emissions: principles, procedures, and protocols" (2011), Plural Publishing
  4. Petersen L, "A systematic review of stimulus parameters for eliciting distortion product otoacoustic emissions from adult humans", Int J Audiol, 56, pp.382~391 (2017)
  5. Parthasarathy TK, "Similarities and differences in distortion-product otoacoustic emissions among four FDA-approved devices", J Am Acad Audiol, 12, pp.397~405 (2001)
  6. "Tympanometry:relevant paper" (1988)[URL]
  7. Carter L, "TE and DP otoacoustic emission data from an Australian cross-sectional hearing study", Int J Audiol, 54, pp.806~817 (2015)
  8. Vinck BM, "Distortion product otoacoustic emissions : a normative study", Int J Audiol, 35, pp.231~245 (1996)
  9. Dunckley KT, "Gender effects on high frequency distortion product otoacoustic emissions in humans", Ear Hear, 25, pp.554~564 (2004)
  10. McFadden D, "Sex differences in distortion-product and transient-evoked otoacoustic emissions compared", J Acoust Soc Am, 125, pp.239~246 (2009)
  11. Pinochet K, "Effects of varying the intensity of the primary tones on the amplitudes of distortion product otoacoustic emissions (OAEdp) in patients with normal hearing", Rev Otorrinolaringol Cir Cabeza Cuello, 72, pp.235~242 (2012)
  12. Petersen L, "Towards the preferred stimulus parameters for distortion product otoacoustic emissions in adults : a preliminary study", S Afr J Commun Disord, 65, pp.e1~e10 (2018)
  13. Ng IHY, "Test-retest reliability of distortion product otoacoustic emissions in the 1 to 7 kHz range", Audiol Med, 3, pp.108~115 (2005)
  14. Siegel JH, "Sound calibration and distortion product otoacoustic emissions at high frequencies", Hear Res, 80, pp.146~152 (1994)
  15. Dreisbach LE, "Distortion-product otoacoustic emissions measured at high frequencies in humans", J Acoust Soc Am, 110(5 Pt 1), pp.2456~2469 (2001)

관련 논문AI추천

이 간행물의 다른 논문 (10건)