엑스선의 조사시간에 따른 형광유리선량계의 빌드업 특성
Buildup Characteristics of Radiophotoluminescent Glass Dosimeters with Exposure Time of X-ray
0
조회
0
다운로드
4
인용
0
참고문헌
키워드
초록
By using the buildup characteristics of the radiophotoluminescence glass dosimeter(RPLGD), it is aimed to help the measurement of the accurate dose by measuring the radiation dose according to the time of the glass element. Five glass elements were arranged on the table and the source to image receptor distance(SID) was set to 100 cm for the build-up radiation dose measurement of the fluorescent glass dosimeter glass element(GD-352M). Radiation doses and saturation rates were measured over time according to irradiation time, with the tube voltage (30, 60, 90 kVp) and tube current (50, 100 mAs) Repeatability test was repeated ten times to measure the coefficient of variation. The radiation dose increased from 0.182 mGy to 12.902 mGy and the saturation rate increased from 58.3% with increasing exposure condition and time. The coefficient of variation of the glass elements of the fluorescent glass dosimeter was ranged from 0.2 to 0.77 according to the X - ray exposure conditions. X - ray exposure showed that the radiation dose and saturation rate were increased with buildup characteristics, and degeneration of glass elements was not observed. The reproducibility of the variation coefficient of the radiation generator was included within the error range and the reproducibility of the radiation dose was excellent.
Abstract
By using the buildup characteristics of the radiophotoluminescence glass dosimeter(RPLGD), it is aimed to help the measurement of the accurate dose by measuring the radiation dose according to the time of the glass element. Five glass elements were arranged on the table and the source to image receptor distance(SID) was set to 100 cm for the build-up radiation dose measurement of the fluorescent glass dosimeter glass element(GD-352M). Radiation doses and saturation rates were measured over time according to irradiation time, with the tube voltage (30, 60, 90 kVp) and tube current (50, 100 mAs) Repeatability test was repeated ten times to measure the coefficient of variation. The radiation dose increased from 0.182 mGy to 12.902 mGy and the saturation rate increased from 58.3% with increasing exposure condition and time. The coefficient of variation of the glass elements of the fluorescent glass dosimeter was ranged from 0.2 to 0.77 according to the X - ray exposure conditions. X - ray exposure showed that the radiation dose and saturation rate were increased with buildup characteristics, and degeneration of glass elements was not observed. The reproducibility of the variation coefficient of the radiation generator was included within the error range and the reproducibility of the radiation dose was excellent.
참고문헌 (13)
- Piesch E, "Properties of radiophotoluminescencent glass dosemeter systems using pulsed laser UV excitation", Radiat Prot Dosim, 17, pp.293~297 (1986)
- Burgkhardt B, "New aspects of environmental monitoring using flat phosphate glass and thermoluminescence dosemeters", Radiat Prot Dosim, 66(1), pp.187~192 (1996)
- Asahi Techno Glass Corporation, "Technical information for RPL glass dosemeter type SC-1" (2000)
- Piesch E, "Photoluminescence dosimetry : progress and present state of art", Radiat Prot Dosim, 33(1-4), pp.215~226 (1990)
- Kneževi Ž, "Characterisation of RPL and TL dosimetry systems and comparison in medical dosimetry applications", Radiat Mes, 46, pp.1582~1585 (2011)
- Rah EJ, "A comparison of the dosimetric characteristics of a glass rod dosimeter and a thermoluminescent dosimeter for mailed dosimeter", Radiat Meas, 44(1), pp.18~22 (2009)
- Araki F, "Measurements of gamma-knife helmet output factors using a radiophotoluminescent glass rod dosimeter and a diode detector", Med Phys, 30(8), pp.1976~1981 (2003)
- Lee JH, "Dosimertry characteristics and performance comparisons : Environmental radiophotoluminescent glass dosemeters versus thermoluminescent dosemeters", Radiat Meas, 44, pp.86~91 (2009)
- Asahi Techno Glass Corporation, "New RPL glass dosemeter system for large scale personal monitoring" (2001)
- Hsu SM, "Comparison on characteristics of radiophotoluminescent glass dosemeters and thermoluminescent dosemeters", Radiat Prot Dosim, 119, pp.327~331 (2006)
- 라정은, "유리선량계의 선량 특성에 관한 연구", 방사선방어학회지, 31(4), pp.181~186 (2006), 대한방사선방어학회
- 손진현, "유리선량계의 단계별 관전류량 변화에 따른 특성연구", 방사선기술과학, 37(2), pp.135~141 (2014), 대한방사선과학회
- Lim SN, "Tumoricidal activity of low-energy 160-kV versus 6-MV Xrays against platinum-sensitized F98 glioma cells", J Radiat Res, 56(1), pp.77~89 (2015)