본문 바로가기
KERD
논문KCI 등재

Start-to-end modeling and transmission efficiency optimization for a cyclotron-based proton therapy beamline

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류원자력공학

초록

Utilizing first-order beam dynamics models is adequate for studying the beam properties during the conceptual design of a cyclotron-based proton therapy beamline. After finishing lattice design, particle-matter interaction simulations for passive elements (e.g., degrader, collimators, energy slit) are required. The cascade simulation is used for lattice updates in each iteration, which is complicated. In addition, when the models involve particle tracking and particle-matter interaction, their optimization process is time-consuming. Therefore, this study proposes a start-to-end modeling method using Monte Carlo Beam Delivery Simulation (BDSIM) software that considers more realistic factors, such as particle-matter interaction and the realistic vacuum chamber, to precisely evaluate working parameters, along with an efficient optimization method that utilizes multi-objective Bayesian optimization (MOBO) to improve transmission efficiency. Taking the Huazhong University of Science and Technology proton therapy facility (HUST-PTF) as an example, beam loss along the beamline is located, quantified, and subsequently reduced by tuning the quadrupole strengths based on MOBO. The results show that: (i) By considering the particle-matter interaction and the realistic vacuum chamber, the precision in the prediction of the beam properties is improved; (ii) After optimization, the transmission efficiency of the entire beamline is relatively increased by an average of 6.52 % under different energy settings, especially 11.39 % at 70 MeV.

Abstract

Utilizing first-order beam dynamics models is adequate for studying the beam properties during the conceptual design of a cyclotron-based proton therapy beamline. After finishing lattice design, particle-matter interaction simulations for passive elements (e.g., degrader, collimators, energy slit) are required. The cascade simulation is used for lattice updates in each iteration, which is complicated. In addition, when the models involve particle tracking and particle-matter interaction, their optimization process is time-consuming. Therefore, this study proposes a start-to-end modeling method using Monte Carlo Beam Delivery Simulation (BDSIM) software that considers more realistic factors, such as particle-matter interaction and the realistic vacuum chamber, to precisely evaluate working parameters, along with an efficient optimization method that utilizes multi-objective Bayesian optimization (MOBO) to improve transmission efficiency. Taking the Huazhong University of Science and Technology proton therapy facility (HUST-PTF) as an example, beam loss along the beamline is located, quantified, and subsequently reduced by tuning the quadrupole strengths based on MOBO. The results show that: (i) By considering the particle-matter interaction and the realistic vacuum chamber, the precision in the prediction of the beam properties is improved; (ii) After optimization, the transmission efficiency of the entire beamline is relatively increased by an average of 6.52 % under different energy settings, especially 11.39 % at 70 MeV.

참고문헌 (29)

  1. R. R. Wilson, "Radiological use of fast protons", Radiology, 47(5), pp.487~491 (1946)
  2. V. Rizzoglio, "On the accuracy of Monte Carlo based beam dynamics models for the degrader in proton therapy facilities", Nucl. Instrum. Methods Phys. Res. Sect. A Accel. Spectrom. Detect. Assoc. Equip, 898, pp.1~10 (2018)
  3. U. Rohrer, "PSI Graphic Transport Framework", ~
  4. U. Rohrer, "PSI Graphic Turtle Framework", ~
  5. S. Agostinelli, "Geant4—a simulation toolkit", Nucl. Instrum. Methods Phys. Res. Sect. A Accel. Spectrom. Detect. Assoc. Equip, 506(3), pp.250~303 (2003)
  6. P. R. S. A. Ferrari, "CERN Yellow Report No. CERN-2005-10", ~
  7. M. K. Hamad, "Bragg-curve simulation of carbon-ion beams for particle-therapy applications : a study with the GEANT4 toolkit", Nucl. Eng. Technol, 53(8), pp.2767~2773 (2021)
  8. B. Qin, "Design of gantry beamline for HUST proton therapy facility", IEEE Trans. Appl. Supercond, 28(3), pp.1~5 (2018)
  9. Z. Liang, "Design and optimization of an energy degrader with a multi-wedge scheme based on Geant4", Nucl. Instrum. Methods Phys. Res. Sect. A Accel. Spectrom. Detect. Assoc. Equip, 890, pp.112~118 (2018)
  10. W. Chen, "Transmission calculation and intensity suppression for a proton therapy system", Nucl. Instrum. Methods Phys. Res. Sect. A Accel. Spectrom. Detect. Assoc. Equip, 881, pp.82~87 (2018)
  11. C. Hernalsteens, "A novel approach to seamless simulations of compact hadron therapy systems for self-consistent evaluation of dosimetric and radiation protection quantities", Europhys. Lett, 132(5), pp.50004~ (2020)
  12. E. Ramoisiaux, "Hybrid monitoring and measurement of concrete shielding activation at the ProtherWal proton therapy centre", EPJ Tech. Instrum, 10(1), pp.9~ (2023)
  13. E. Ramoisiaux, "Self-consistent numerical evaluation of concrete shielding activation for proton therapy systems", Europ. Phys. J. Plus, 137(8), pp.889~ (2022)
  14. V. Maradia, "A new emittance selection system to maximize beam transmission for low-energy beams in cyclotron-based proton therapy facilities with gantry", Med. Phys, 48(12), pp.7613~7622 (2021)
  15. V. Maradia, "Increase of the transmission and emittance acceptance through a cyclotron-based proton therapy gantry", Med. Phys, 49(4), pp.2183~2192 (2022)
  16. V. Rizzoglio, "Evolution of a beam dynamics model for the transport line in a proton therapy facility", Phys. Rev. Accelerat. Beams, 20(12), pp.124702~ (2017)
  17. R. Tesse, "Georges : a modular Python library for seamless beam dynamics simulations and optimization", SoftwareX, 24, pp.101579~ (2023)
  18. A. Edelen, "Machine learning for orders of magnitude speedup in multiobjective optimization of particle accelerator systems", Phys. Rev. Accelerat. Beams, 23(4), pp.044601~ (2020)
  19. B. Qin, "Design and development of the beamline for a proton therapy system", Nucl. Sci. Tech, 32(12), pp.138~ (2021)
  20. P. Li, "Design of HUST-PTF beamline control system for fast energy changing", Nucl. Eng. Technol, 54(8), pp.2852~2858 (2022)
  21. L. J. Nevay, "BDSIM : an accelerator tracking code with particle-matter interactions", Comput. Phys. Commun, 252, pp.107200~ (2018)
  22. A. V. Ivantchenko, "Geant4 hadronic physics for space radiation environment", Int. J. Radiat. Biol, 88(1–2), pp.171~175 (2012)
  23. V. Nguyen, "Bayesian optimization for accelerating hyper-parameter tuning", 2019 IEEE Second International Conference on Artificial Intelligence and Knowledge Engineering (AIKE), pp.302~305 (2019)
  24. D. Eriksson, "Latency-Aware Neural Architecture Search with Multi-Objective Bayesian Optimization, abs", 2106, 2021
  25. D. Eriksson, "High-dimensional Bayesian optimization with sparse axis-aligned subspaces", Presented at the Proceedings of the Thirty-Seventh Conference on Uncertainty in Artificial Intelligence, Proceedings of Machine Learning Research (2021)
  26. S. Daulton, "Neural Information Processing Systems" (2021)

관련 논문AI추천

A study of multi-parameter impact on beam position accuracy for scanning proton therapyLiao Sheng, Shen Junsong, Wang Ming, Xie Guoyu, Zeng Xianhu한국원자력학회 · Nuclear Engineering and Technology2026 · 제58권 제2호Beamline analysis for a laser-driven proton therapy accelerator using superconducting bendsCai Shixian, Easton Matthew, Huang Wei, Jiang Xiaopeng, Liang Yu, Li Jie, Lin Chen, Wang Kai, Wang Kedong, Wu Wei, Yan Xueqing, Zhang Caijie, Zhu Kun, Zhu Tingru한국원자력학회 · Nuclear Engineering and Technology2025 · 제57권 제3호Improved HUST-PTF beamline control system for beam commissioningChen Qushan, Chen Yu, Li Dong, Liu Chengyong, Qin Bin, Shu Lingfeng, Yang Zhenyi, Zhou Shaokun한국원자력학회 · Nuclear Engineering and Technology2025 · 제57권 제12호Conceptual RF design of 750 MHz IH cavities for β  = 0.10–0.15 ion beams in medical acceleratorsAlvarado María, Calvo Pedro, Carmona José Miguel, Gavela Daniel, Giner Navarro Jorge, León Miguel, López Ricardo, Moreno Gabriela, Oliver Concepción, Rodríguez Ángel한국원자력학회 · Nuclear Engineering and Technology2024 · 제56권 제9호Analysis of fringe field effects on measurement errors for Q-scanning-based emittance diagnosis systemHu H., Hu T.N., Wang Y., Zeng Y.F.한국원자력학회 · Nuclear Engineering and Technology2025 · 제57권 제5호Analysis and optimizations of macro-pulse performances for a constructed high-power beam injector with a compact layoutFeng G.Y., Hu H., Hu T.N., Pei Y.J., Wang H.M., Wang Y., Xu H.J., Zeng Y.F.한국원자력학회 · Nuclear Engineering and Technology2025 · 제57권 제10호

이 간행물의 다른 논문 (49건)