본문 바로가기
KERD
논문KCI 등재

Comparison of Precision between Optical and Electromagnetic Navigation Systems in Total Knee Arthroplasty

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

초록

purpose: The purpose of this study is to compare and analyze the precision of optical and electromagnetic navigation systems in total knee arthroplasty (TKA). materials and methods: We retrospectively reviewed 60 patients who underwent TKA using an optical navigation system and 60 patients who underwent TKA using an electromagnetic navigation system from June 2010 to March 2012. The mechanical axis that was measured on preoperative radiographs and by the intraoperative navigation systems were compared between the groups. The postoperative positions of the femoral and tibial components in the sagittal and coronal plane were assessed. results: The difference of the mechanical axis measured on the preoperative radiograph and by the intraoperative navigation systems was 0.6 degrees more varus in the electromagnetic navigation system group than in the optical navigation system group, but showed no statistically significant difference between the two groups (p>0.05). The positions of the femoral and tibial components in the sagittal and coronal planes on the postoperative radiographs also showed no statistically significant difference between the two groups (p>0.05).conclusions: In TKA, both optical and electromagnetic navigation systems showed high accuracy and reproducibility, and the measurements from the postoperative radiographs showed no significant difference between the two groups.

Abstract

purpose: The purpose of this study is to compare and analyze the precision of optical and electromagnetic navigation systems in total knee arthroplasty (TKA). materials and methods: We retrospectively reviewed 60 patients who underwent TKA using an optical navigation system and 60 patients who underwent TKA using an electromagnetic navigation system from June 2010 to March 2012. The mechanical axis that was measured on preoperative radiographs and by the intraoperative navigation systems were compared between the groups. The postoperative positions of the femoral and tibial components in the sagittal and coronal plane were assessed. results: The difference of the mechanical axis measured on the preoperative radiograph and by the intraoperative navigation systems was 0.6 degrees more varus in the electromagnetic navigation system group than in the optical navigation system group, but showed no statistically significant difference between the two groups (p>0.05). The positions of the femoral and tibial components in the sagittal and coronal planes on the postoperative radiographs also showed no statistically significant difference between the two groups (p>0.05).conclusions: In TKA, both optical and electromagnetic navigation systems showed high accuracy and reproducibility, and the measurements from the postoperative radiographs showed no significant difference between the two groups.

참고문헌 (12)

  1. Lotke PA, "Influence of positioning of prosthesis in total knee replacement", J Bone Joint Surg Am, 59, pp.77~79 (1977)
  2. Moreland JR, "Mechanisms of failure in total knee arthroplasty", Orthop Relat Res, 226, pp.49~64 (1988)
  3. Ritter MA, "Postoperative alignment of total knee replacement. Its effect on survival", Clin Orthop Relat Res, 299, pp.153~156 (1994)
  4. Amin DV, "Ultrasound registration of the bone surface for surgical navigation", Comput Aided Surg, 8, pp.1~16 (2003)
  5. Pitto RP, "Accuracy of a computer-assisted navigation system for total knee replacement", J Bone Joint Surg Br, 88, pp.601~605 (2006)
  6. Song EK, "Accuracy of navigation:a comparative study of infrared optical and electromagnetic navigation", Orthopedics, 31(10) (2008)
  7. Chen TK, "A system for ultrasound-guided computer-assisted orthopaedic surgery", Comput Aided Surg, 10, pp.281~292 (2005)
  8. Hummel J, "Evaluation of a new electromagnetic tracking system using a standardized assessment protocol", Phys Med Biol, 51, pp.N205~N210 (2006)
  9. Khadem R, "Comparative tracking error analysis of five different optical tracking systems", Comput Aided Surg, 5, pp.98~107 (2000)
  10. Schicho K, "Stability of miniature electromagnetic tracking systems", Phys Med Biol, 50, pp.2089~2098 (2005)
  11. Stiehl JB, "Six sigma analysis of computer-assisted surgery tracking protocols in TKA", Clin Orthop Relat Res, 464, pp.105~110 (2007)
  12. Yau WP, "Intraobserver errors in obtaining visually selected anatomic landmarks during registration process in nonimage-based navigationassisted total knee arthroplasty : a cadaveric experiment", J Arthroplasty, 20, pp.591~601 (2005)

관련 논문AI추천

Comparison of an Accelerometer-Based Portable Navigation System, Patient-Specific Instrumentation, and Conventional Instrumentation for Femoral Alignment in Total Knee ArthroplastyJunya Higuchi, Kazuhiko Michishita, Kohei Kawaguchi, Takeshi Manabe, Yoshiyuki Akasaka대한슬관절학회 · Knee Surgery and Related Research2017 · 제29권 제4호The Accuracy of Alignment Determined by Patient-Specific Instrumentation System in Total Knee Arthroplasty김광균, 이도경, Jiyoon Eom, Jungsung Kim, Sohui Lim대한슬관절학회 · Knee Surgery and Related Research2019 · 제31권 제1호Reliability and Validity of the Femorotibial Mechanical Axis Angle in Primary Total Knee Arthroplasty: Navigation versus Weight Bearing or Supine Whole Leg Radiographs박용범, 이한준, Jung­Won Lim, Min­Ku Song, Seong Hwan Kim대한슬관절학회 · Knee Surgery and Related Research2018 · 제30권 제4호Effects of Femoral Lateral Bowing on Coronal Alignment and Component Position after Total Knee Arthroplasty: A Comparison of Conventional and Navigation­-Assisted Surgery이창락, Chang-Wan Kim대한슬관절학회 · Knee Surgery and Related Research2018 · 제30권 제1호Navigation-Assisted Total Knee Arthroplasty for Patients with Extra-Articular Deformity서정탁, 서창효, 이승준대한슬관절학회 · Knee Surgery and Related Research2013 · 제25권 제4호해부학적 내반 변형을 가진 퇴행성 슬관절염 환자에서 슬관절 전치환술 시술 시 고식적 방법과 항법장치를 이용한 삽입물 위치와 하지 정렬의 비교양재혁, 오광준, 이호준대한슬관절학회 · Knee Surgery and Related Research2010 · 제22권 제1호