Accurate Localization of Metal Electrodes Using Magnetic Resonance Imaging

1

조회

0

다운로드

1

인용

0

참고문헌

키워드

주제분류방사선과학

초록

Purpose : Localization using MRI is difficult due to susceptibility induced artifacts caused by metal electrodes. Here we took an advantage of the B0 pattern induced by the metal electrodes by using an oblique-view imaging method. Materials and Methods : Metal electrode models with various diameters and susceptibilities were simulated to understand the aspect of field distortion. We set localization criteria for a turbo spin-echo (TSE) sequence using conventional (90view) and 45oblique-view imaging method through simulation of images with various resolutions and validated the criteria usingphantom images acquired by a 3.0T clinical MRI system. For a gradient-refocused echo (GRE) sequence, which is relatively more sensitive to field inhomogeneity, we used phase images to find the center of electrode. Results : There was least field inhomogeneity along the 45line that penetrated the center of the electrode. Therefore, our criteria for the TSE sequence with 45oblique view was coincided regardless of susceptibility. And with 45oblique-view angle images, pixel shifts were bidirectional so we can detect the location of electrodes even in low resolution. For the GRE sequence, the 45oblique-view angle method made the lines where field polarity changes become coincident to the Cartesian grid so the localization of the center coordinates was more facilitated. Conclusion : We suggested the method for accurate localization of electrode using 45 oblique-view angle imaging. It is expected to be a novel method to monitoring an electrophysiological brain study and brain neurosurgery.

Abstract

Purpose : Localization using MRI is difficult due to susceptibility induced artifacts caused by metal electrodes. Here we took an advantage of the B0 pattern induced by the metal electrodes by using an oblique-view imaging method. Materials and Methods : Metal electrode models with various diameters and susceptibilities were simulated to understand the aspect of field distortion. We set localization criteria for a turbo spin-echo (TSE) sequence using conventional (90view) and 45oblique-view imaging method through simulation of images with various resolutions and validated the criteria usingphantom images acquired by a 3.0T clinical MRI system. For a gradient-refocused echo (GRE) sequence, which is relatively more sensitive to field inhomogeneity, we used phase images to find the center of electrode. Results : There was least field inhomogeneity along the 45line that penetrated the center of the electrode. Therefore, our criteria for the TSE sequence with 45oblique view was coincided regardless of susceptibility. And with 45oblique-view angle images, pixel shifts were bidirectional so we can detect the location of electrodes even in low resolution. For the GRE sequence, the 45oblique-view angle method made the lines where field polarity changes become coincident to the Cartesian grid so the localization of the center coordinates was more facilitated. Conclusion : We suggested the method for accurate localization of electrode using 45 oblique-view angle imaging. It is expected to be a novel method to monitoring an electrophysiological brain study and brain neurosurgery.

참고문헌 (14)

  1. Schrader B, "Documentation of electrode localization", Mov Disord, 17, pp.S167~S174 (2002)
  2. Lüdeke KM, "Susceptibility artifacts in NMR imaging", Magn Reson Imaging, 3, pp.329~343 (1985)
  3. Ladd ME, "Biopsy needle susceptibility artifacts", Magn Reson Med, 36, pp.646~651 (1996)
  4. Müller-Bierl B, "Numerical modeling of needle tip artifacts in MR gradient echo imaging", Med Phys, 31, pp.579~587 (2004)
  5. Martinez-Santiesteban FM, "Magnetic field perturbation of neural recording and stimulating microelectrodes", Phys Med Biol, 52, pp.2073~2088 (2007)
  6. Liu H, "Biopsy needle tip artifact in MR-Guided Neurosurgery", J Magn Reson Imaging, 13, pp.16~22 (2001)
  7. Matsui T, "MRI-based localization of electrophysiological recording sites within the cerebral cortex at single-voxel accuracy", Nat Methods, 4, pp.161~168 (2007)
  8. Truwit CL, "Prospective stereotaxy: a novel method of trajectory alignment using real-time image guidance", J Magn Reson Imaging, 13, pp.452~457 (2001)
  9. Martin AJ, "Placement of deep brain stimulator electrodes using real-time high-field interventional magnetic resonance imaging", Magn Reson Med, 54, pp.1107~1114 (2005)
  10. Zrinzo L, "Stereotactic localization of the human pedunculopontine nucleus: atlas-based coordinates and validation of a magnetic resonance imaging protocol for direct localization", Brain, 131, pp.1588~1598 (2008)
  11. Oya H, "Stereotactic atlas-based depth electrode localization in the human amygdala", Stereotact Funct Neurosurg, 87, pp.219~228 (2009)
  12. Salomir R, "fast calculation method for magnetic field inhomogeneity due to an arbitrary distribution of bulk susceptibility", Concept Magn Reson B, 19B, pp.26~34 (2003)
  13. Koch KM, "Rapid calculations of susceptibility-induced magnetostatic field perturbations for in vivo magnetic resonance", Phys Med Biol, 51, pp.6381~6402 (2006)
  14. Hopper TAJ, "Experimental and computational analyses of the effects of slice distortion from a metallic sphere in an MRI phantom", Magn Reson Imaging, 24, pp.1077~1085 (2006)

관련 논문AI추천

이 간행물의 다른 논문 (9건)