Quantitative Assessment and Ligament Traceability of Volume Isotropic Turbo Spin Echo Acquisition (VISTA) Ankle Magnetic Resonance Imaging: Fat Suppression versus without Fat Suppression
0
조회
0
다운로드
3
인용
0
참고문헌
키워드
초록
Purpose : To compare the image quality and ligament traceability in ankle images obtained using Volume Isotropic Turbo Spin Echo Acquisition (VISTA) MRI with and without fat suppression. Materials and Methods: The signal-to-noise ratios (SNRs) in images from a phantom and from the ankle of a volunteer were compared. Ten ankles from 10 non-symptomatic volunteers were imaged for comparisons of contrast ratio (CR) and ligament traceability. All examinations were performed using VISTA sequences with and without fat suppression on a 3T MRI scanner. The SNRs were obtained from images with subjects and without subjects (noise-only). Contrast ratios from images of the 10 ankles were acquired between fluid and tendon (F-T), F-cartilage (C), F-ligament (L), fat (f)-T, f-C and f-L. Two musculoskeletal radiologists independently scored the traceability of 7 ligaments, in sagittal, axial and coronal images respectively, based on a 4-point scale (1 as not traceable through 4 as clearly traceable). The Wilcoxon signed-rank test was used to compare the CR. Fisher’s exact test and Pearson’s chi-squared test were used to compare the ligament traceability. Results: The SNRs did not differ significantly between the two sequences except in bone marrow. VISTA SPAIR showed the higher CR only in F-T (p = 0.04), whereas VISTA showed higher CR in f-T (p = 0.005), f-C (p = 0.005) and f-L (p =0.005). The calcaneofibular ligament traceability with VISTA was superior to that obtained with VISTA SPAIR (p < 0.05) in all planes. Conclusion: VISTA showed significant superiority to VISTA SPAIR in tracing CFL due to the superior CR between fat and ligament.
Abstract
Purpose : To compare the image quality and ligament traceability in ankle images obtained using Volume Isotropic Turbo Spin Echo Acquisition (VISTA) MRI with and without fat suppression. Materials and Methods: The signal-to-noise ratios (SNRs) in images from a phantom and from the ankle of a volunteer were compared. Ten ankles from 10 non-symptomatic volunteers were imaged for comparisons of contrast ratio (CR) and ligament traceability. All examinations were performed using VISTA sequences with and without fat suppression on a 3T MRI scanner. The SNRs were obtained from images with subjects and without subjects (noise-only). Contrast ratios from images of the 10 ankles were acquired between fluid and tendon (F-T), F-cartilage (C), F-ligament (L), fat (f)-T, f-C and f-L. Two musculoskeletal radiologists independently scored the traceability of 7 ligaments, in sagittal, axial and coronal images respectively, based on a 4-point scale (1 as not traceable through 4 as clearly traceable). The Wilcoxon signed-rank test was used to compare the CR. Fisher’s exact test and Pearson’s chi-squared test were used to compare the ligament traceability. Results: The SNRs did not differ significantly between the two sequences except in bone marrow. VISTA SPAIR showed the higher CR only in F-T (p = 0.04), whereas VISTA showed higher CR in f-T (p = 0.005), f-C (p = 0.005) and f-L (p =0.005). The calcaneofibular ligament traceability with VISTA was superior to that obtained with VISTA SPAIR (p < 0.05) in all planes. Conclusion: VISTA showed significant superiority to VISTA SPAIR in tracing CFL due to the superior CR between fat and ligament.
참고문헌 (36)
- Fong DT, "A systematic review on ankle injury and ankle sprain in sports", Sports Medicine, 37, pp.73~94 (2007)
- 김현수, "Qualitative and Quantitative Assessment of Isotropic Ankle Magnetic Resonance Imaging: Three-Dimensional Isotropic Intermediate-Weighted Turbo Spin Echo versus Three-Dimensional Isotropic Fast Field Echo Sequences", Korean Journal of Radiology, 13(4), pp.443~449 (2012), 대한영상의학회
- Gold GE, "Recent advances in MRI of articular cartilage", AJR Am J Roentgenol, 193, pp.628~638 (2009)
- Stevens KJ, "Ankle: isotropic MR imaging with 3D-FSE-cube--initial experience in healthy volunteers", Radiology, 249, pp.1026~1033 (2008)
- Yao L, "Isotropic 3D fast spin-echo with proton-density-like contrast: a comprehensive approach to musculoskeletal MRI", AJR Am J Roentgenol, 188, pp.W199~W201 (2007)
- Gold GE, "Isotropic MRI of the knee with 3D fast spin-echo extended echo-train acquisition (XETA): initial experience", AJR Am J Roentgenol, 188, pp.1287~1293 (2007)
- Jung JY, "Diagnosis of internal derangement of the knee at 3.0-T MR imaging: 3D isotropic intermediate-weighted versus 2D sequences", Radiology, 253, pp.780~787 (2009)
- Ristow O, "Isotropic 3-dimensional fast spin echo imaging versus standard 2-dimensional imaging at 3.0 T of the knee: artificial cartilage and meniscal lesions in a porcine model", J Comput Assist Tomogr, 34, pp.260~269 (2010)
- Notohamiprodjo M, "MRI of the knee at 3T: first clinical results with an isotropic PDfsweighted 3D-TSE-sequence", Invest Radiol, 44, pp.585~597 (2009)
- Jung JY, "Threedimensional isotropic shoulder MR arthrography: comparison with two-dimensional MR arthrography for the diagnosis of labral lesions at 3.0 T", Radiology, 250, pp.498~505 (2009)
- Choo HJ, "Comparison of threedimensional isotropic T1-weighted fast spin-echo MR arthrography with two-dimensional MR arthrography of the shoulder", Radiology, 262, pp.921~931 (2012)
- Notohamiprodjo M, "3D-MRI of the ankle with optimized 3D-SPACE", Invest Radiol, 47, pp.231~239 (2012)
- Delfaut EM, "Fat suppression in MR imaging: techniques and pitfalls", Radiographics, 19, pp.373~382 (1999)
- Lee S, "Proton density-weighted MR imaging of the knee: fat suppression versus without fat suppression", Skeletal Radiology, 40, pp.189~195 (2011)
- Sch¨afer FK, "Value of fat-suppressed proton-density-weighted turbo spin-echo sequences in detecting meniscal lesions: comparison with arthroscopy", Acta Radiol, 47, pp.385~390 (2006)
- Seo JM, "3D isotropic turbo spin-echo intermediate-weighted sequence with refocusing control in knee imaging: comparison study with 3D isotropic fast-field echo sequence", Acta Radiol, 52, pp.1119~1124 (2011)
- Dietrich O, "Measurement of signal-to-noise ratios in MR images: influence of multichannel coils, parallel imaging, and reconstruction filters", J Magn Reson Imaging, 26, pp.375~385 (2007)
- Viera AJ, "Understanding interobserver agreement: the kappa statistic", Fam Med, 37, pp.360~363 (2005)
- Gudbjartsson H, "The Rician distribution of noisy MRI data", Magn Reson Med, 34, pp.910~914 (1995)
- Nelles M, "Dual-source parallel RF transmission for clinical MR imaging of the spine at 3.0 T: intraindividual comparison with conventional single-source transmission", Radiology, 257, pp.743~753 (2010)
- Turetschek K, "Double inversion recovery imaging of the brain: initial experience and comparison with fluid attenuated inversion recovery imaging", Magn Reson Imaging, 16, pp.127~135 (1998)
- Crewson PE, "Reader agreement studies", AJR Am J Roentgenol, 184, pp.1391~1397 (2005)
- Kundel HL, "Measurement of observer agreement", Radiology, 228, pp.303~308 (2003)
- Wang Z, "2D partially parallel imaging with k-space surrounding neighbors-based data reconstruction", Magn Reson Med, 56, pp.1389~1396 (2006)
- Noll DC, "Homodyne detection in magnetic resonance imaging", IEEE Transactions on Medical Imaging, 10, pp.154~163 (1991)
- Singson RD, "MR evaluation of rotator cuff pathology using T2-weighted fast spin-echo technique with and without fat suppression", AJR Am J Roentgenol, 166, pp.1061~1065 (1996)
- Chimich D, "The effects of initial end contact on medial collateral ligament healing: a morphological and biomechanical study in a rabbit model", J Orthop Res, 9, pp.37~47 (1991)
- Brostrm L, "Sprained ankles. IV. Histologic changes in recent and “chronic” ligament ruptures", Acta Chir Scand, 132, pp.248~253 (1966)
- Datir A, "Imaging of impingement lesions in the ankle", Top Magn Reson Imaging, 21, pp.15~23 (2010)
- Kitsoulis P, "Morphological study of the calcaneofibular ligament in cadavers", Folia Morphol, 70, pp.180~184 (2011)
- Dimmick S, "Evaluation of thickness and appearance of anterior talofibular and calcaneofibular ligaments in normal versus abnormal ankles with MRI", J Med Imaging Radiat Oncol, 52, pp.559~563 (2008)
- Mirowitz SA, "MR imaging evaluation of knee collateral ligaments and related injuries: comparison of T1-weighted, T2-weighted, and fat-saturated T2-weighted sequences--correlation with clinical findings", J Magn Reson Imaging, 4, pp.725~732 (1994)
- Schneck CD, "MR imaging of the most commonly injured ankle ligaments. Part II. Ligament injuries", Radiology, 184, pp.507~512 (1992)
- Cass JR, "Ankle instability: current concepts, diagnosis, and treatment", Mayo Clinic proceedings, 59, pp.165~170 (1984)
- Kumar V, "Deficiencies of MRI in the diagnosis of chronic symptomatic lateral ankle ligament injuries", Foot and ankle surgery, 13, pp.171~176 (2007)
- Park HJ, "Accuracy of MRI findings in chronic lateral ankle ligament injury: comparison with surgical findings", Clin Radiol, 67, pp.313~318 (2012)