본문 바로가기
KERD
논문KCI 등재

성견에서 즉시 임플란트 식립시 국내 임플란트에서의 골 유착반응에 관한 연구

Study on Osseointegration of the Immediate Placement Using Korean Implant Systems in Beagle Dog

0

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류치의학

초록

This study was attempted to evaluate home-manufactured implants by placing Stage-1Ⓡ Implant (Lifecore. Co., USA) whose surface is treated with RBM that has already been varified clinically, ChaorumⓇ Implant(Chaorum Co., Korea) whose surface treatment is same as that of Stage-1 Implant and AtlasⓇ Implant(Cewellmedi Co., Korea) whose surface is treated with anodic oxidation immediately after the teeth of experimental animals were extracted to compare histological findings among them. Stage-1 Implant(diameter: 3.5mm, length: 10mm), Chaorum Implant(diameter: 3.3mm, length: 8.5mm) and Cowell medi Implant(diameter: 4.0mm, length: 8.0mm) were placed into the mandible premolars of 2 adult beagle dogs immediately after their teeth were extracted, and then histological findings were analyzed at 6 weeks. After those implants were inserted directly after their teeth were extracted, the results of periotest were recorded, radiography was done, the subjects went through thorough control for 6 weeks, and then comparison among periotest, radiography and histological finding was made. After comparison of those findings, the values of periotest were satisfactory and bone healing was relatively satisfactory on radiography at 6 weeks. For osseointegration with the bone tissue, Stage-1 was 45.3%, Chaorum 55.3%, and Cowellmedi 52.5%, which was a satisfactory result. Although implant surgery immediately after teeth were extracted involves difficulties among recent implant surgeries, it is being frequently used in that it may reduce surgery hours, the frequency of surgery, and bone loss for patients. This experiment was conducted to evaluate the technological levels of home-manufactured implants that have been remarkably developed in recent years and in conclusion, those implants showed nearly similar result.

Abstract

This study was attempted to evaluate home-manufactured implants by placing Stage-1Ⓡ Implant (Lifecore. Co., USA) whose surface is treated with RBM that has already been varified clinically, ChaorumⓇ Implant(Chaorum Co., Korea) whose surface treatment is same as that of Stage-1 Implant and AtlasⓇ Implant(Cewellmedi Co., Korea) whose surface is treated with anodic oxidation immediately after the teeth of experimental animals were extracted to compare histological findings among them. Stage-1 Implant(diameter: 3.5mm, length: 10mm), Chaorum Implant(diameter: 3.3mm, length: 8.5mm) and Cowell medi Implant(diameter: 4.0mm, length: 8.0mm) were placed into the mandible premolars of 2 adult beagle dogs immediately after their teeth were extracted, and then histological findings were analyzed at 6 weeks. After those implants were inserted directly after their teeth were extracted, the results of periotest were recorded, radiography was done, the subjects went through thorough control for 6 weeks, and then comparison among periotest, radiography and histological finding was made. After comparison of those findings, the values of periotest were satisfactory and bone healing was relatively satisfactory on radiography at 6 weeks. For osseointegration with the bone tissue, Stage-1 was 45.3%, Chaorum 55.3%, and Cowellmedi 52.5%, which was a satisfactory result. Although implant surgery immediately after teeth were extracted involves difficulties among recent implant surgeries, it is being frequently used in that it may reduce surgery hours, the frequency of surgery, and bone loss for patients. This experiment was conducted to evaluate the technological levels of home-manufactured implants that have been remarkably developed in recent years and in conclusion, those implants showed nearly similar result.

참고문헌 (37)

  1. "Osseointegrated implants in the treatment of the edentulous jaw Experience from a 10-years period", pp.1~132 (1977), Branemark
  2. "A 15-year study of osseointegrated implants in the treatment of edentulous jaw", pp.387~416 (1981), and Branemark
  3. "The longitudinal clinical effectiveness of osseointegrated dental implants", pp.451~457 (1990)
  4. "A clinical evaluation of fixed bridge restorations supported by the combination of tooth and osseointegrated tianinum implant", pp.307~312 (1986), Branemark
  5. "Osseointegration in clinical dentistry" (1985), Branemark Quintessence Publishing Co inc
  6. "Histomorphometric studies of hydroxyapatite coated and uncoated CP titanium threaded implants in bone", pp.399~404 (1991)
  7. "Histomorphometric study of unthreaded hydroxyapatite coated and titanium coated implants in rabbit bone", pp.485~490 (1992)
  8. "Mandibular bone response to plasma-sprayed coatings of hydroxyapatite Int Prosthodont", pp.53~58 (1990)
  9. "The Effect of Hydroxyapatite coating on bone Growth into porous titanium alloy implants", pp.71 213~216 (1989)
  10. "A 1-year follow-up of implants of differing surface roughness placed in rabbit bone", 12, pp.486~494 (1997)
  11. "Attachment of bone to threaded implants by ingrowth and mechanical interlocking", pp.401~412
  12. "Optimal implant stabilization in low density bone", 12, pp.423~432 (2001)
  13. "interface mechanics and histology of titanium and hydroxyapatite-coated titanium for dental implant applications", pp.15~22
  14. "Evaluation of hydroxyapatite-coated titanium dental implants in dogs", pp.601~607
  15. "Bone implant contact and bone quality: Evaluation of expected and actual bone contact on machined and osseotite implant surface", 22, pp.535~545 (2002)
  16. "Bone response to machined and resorbable blast material titanium implants: An experimental study in rabbits", 28, pp.2~8 (2002)
  17. "Experimental study of bone response to a new surface treatment of endosseous titanium implants", 10, pp.126~131 (2001)
  18. "The use of reduced healing times on ITI implants with a sandblasted and acid- etched (SLA) surface : early results from clinical trials on ITI SLA implants", 28, pp.2~8 (2002)
  19. "Influence of surface characteritics on bone integration of titanium implants A histomorphometric study in miniature pigs", pp.889~902, Steinemann S
  20. "Three-year retrospective evaluation of 50 consecutive cases" (1993)
  21. "Immediate implantation of pure titanium implants into extraction sockets of Macaca facicularis. Part Ⅱ : Histologic observations", 11, pp.489~ (1996)
  22. "Osseous adaptation to continuous loading of rigid endosseous implant", pp.95~111
  23. "Direct bone anchorage of oral implants Clinical and experimental considerations of the concept of osseointegration", pp.30~41
  24. "Influence of surface characteritics on bone integration of titanium implants A histomorphometric study in miniature pigs", pp.889~902, Steinemann S
  25. "The applicability of osseointegrated implants in the treatmentof partially edentulous patients Three ye ars result of a prospectivemulticenter study", pp.123~129 (1993)
  26. "Integration of screw implants in the rabbit a 1-year follow-up of removal torque of titanium implants", pp.69~75 (1987)
  27. "Reconstruction of severely resorbed edentulous maxillae using osseointegrated fixturesin immediate autogenous bone grafts", pp.233~246 (1990), nemark
  28. "Characteerization of titanium implant surface 3", pp.45~158 (1988)
  29. "Effect of titaium surface roughness on proliferation, differentitaion, and protein synthesis of human osteoblast-like cell(MG63)", 29, pp.389~401 (1995)
  30. "Kineticsof bone growth into cylindrical channels in aluminum oxide and titanium", pp.375~400 (1972)
  31. "An evaluationof variables influencing implant fixation by direct bone apposition", pp.875~901 (1985)
  32. "Influence of surface characteristics on bone integration of titaniumimplants A histomorphometric study in miniature pigs", pp.889~902, Steinemann S
  33. "Optimization of surface micromorphology for enhanced osteoblastresponses in vitro", pp.302~310
  34. "Bone response to machined and resorbable blast material titaniumimplants: An experimental study in rabbits", 28, pp.2~8 (2002)
  35. "A 1-year follow-up of implants of differing surface roughnessplaced in rabbit bone", 12, pp.486~494 (1997)
  36. "Bone response to machined and resorbable blast material titaniumimplants: An experimental study in rabbits", 28, pp.2~8 (2002)
  37. "Experimental study of bone response to a new surface treatment of endosseous titanium implants", 10, pp.126~131 (2001)

관련 논문AI추천

성견에 발치 후 즉시 임플란트 식립시 RBM 처리된 임플란트의 골융합에 관한 연구김윤상, 박재영, 선기종, 신미란, 신형식, 유형근, 정은경, 피성희대한치주과학회 · Journal of Periodontal & Implant Science2007 · 제37권 제2호성견 하악골에 식립한 다양한 임플란트 표면에서의 골 치유 양상에 관한 연구김용학, 박상원, 방몽숙, 양홍서, 임현필, 최홍란대한턱관절교합학회 · 구강회복응용과학지2009 · 제25권 제2호성견에서 SLA®와 Joy 1 implant®의 골-임플란트 계면의 조직형태계측학적 연구김병옥, 박도영, 유상준, 이원표, 정명진, 정민구치의학연구원 · Oral Biology Research2014 · 제38권 제2호Comparative study on the osseointegration of implants in dog mandibles according to the implant surface treatment김수관, 오지수, 유재식, 윤욱재, 임성철, 정경인, 정미애대한구강악안면외과학회 · 대한구강악안면외과학회지2016 · 제42권 제6호성견에서 국산 임시 임플란트의 골 유착반응에 관한 연구김윤상, 신형식, 유형근, 이욱재, 정종원, 피성희대한치주과학회 · Journal of Periodontal & Implant Science2006 · 제36권 제1호Comparative Evaluation of Osseointegration Parameters between Titanium and Zirconia Implants Placed in Beagle Dogs김기현, 설양조, 여인성, 이준범, Nguyen Pham-Hanh Luong대한구강악안면임플란트학회 · 대한구강악안면임플란트학회지2024 · 제28권 제3호

이 간행물의 다른 논문 (11건)