The relationship between dental implant stability and trabecular bone structure using cone-beam computed tomography

2

조회

0

다운로드

2

인용

0

참고문헌

키워드

주제분류치의학

초록

Purpose: The objective of this study was to investigate the relationships between primary implant stability as measured by impact response frequency and the structural parameters of trabecular bone using cone-beam computed tomography(CBCT), excluding the effect of cortical bone thickness. Methods: We measured the impact response of a dental implant placed into swine bone specimens composed of only trabecular bone without the cortical bone layer using an inductive sensor. The peak frequency of the impact response spectrum was determined as an implant stability criterion (SPF). The 3D microstructural parameters were calculated from CT images of the bone specimens obtained using both micro-CT and CBCT. Results: SPF had significant positive correlations with trabecular bone structural parameters (BV/TV, BV, BS, BSD, Tb.Th, Tb.N, FD, and BS/BV) (P<0.01) while SPF demonstrated significant negative correlations with other microstructural parameters (Tb.Sp, Tb.Pf, and SMI) using micro-CT and CBCT (P<0.01). Conclusions: There was an increase in implant stability prediction by combining BV/TV and SMI in the stepwise forward regression analysis. Bone with high volume density and low surface density shows high implant stability. Well-connected thick bone with small marrow spaces also shows high implant stability. The combination of bone density and architectural parameters measured using CBCT can predict the implant stability more accurately than the density alone in clinical diagnoses.

Abstract

Purpose: The objective of this study was to investigate the relationships between primary implant stability as measured by impact response frequency and the structural parameters of trabecular bone using cone-beam computed tomography(CBCT), excluding the effect of cortical bone thickness. Methods: We measured the impact response of a dental implant placed into swine bone specimens composed of only trabecular bone without the cortical bone layer using an inductive sensor. The peak frequency of the impact response spectrum was determined as an implant stability criterion (SPF). The 3D microstructural parameters were calculated from CT images of the bone specimens obtained using both micro-CT and CBCT. Results: SPF had significant positive correlations with trabecular bone structural parameters (BV/TV, BV, BS, BSD, Tb.Th, Tb.N, FD, and BS/BV) (P<0.01) while SPF demonstrated significant negative correlations with other microstructural parameters (Tb.Sp, Tb.Pf, and SMI) using micro-CT and CBCT (P<0.01). Conclusions: There was an increase in implant stability prediction by combining BV/TV and SMI in the stepwise forward regression analysis. Bone with high volume density and low surface density shows high implant stability. Well-connected thick bone with small marrow spaces also shows high implant stability. The combination of bone density and architectural parameters measured using CBCT can predict the implant stability more accurately than the density alone in clinical diagnoses.

참고문헌 (40)

  1. Javed F, "Impact of diabetes mellitus and glycemic control on the osseointegration of dental implants: a systematic literature review", J Periodontol, 80, pp.1719~1730 (2009)
  2. Javed F, "The role of primary stability for successful immediate loading of dental implants. A literature review", J Dent, 38, pp.612~620 (2010)
  3. Meredith N, "Assessment of implant stability as a prognostic determinant", Int J Prosthodont, 11, pp.491~501 (1998)
  4. Sakka S, "Bone quality: a reality for the process of osseointegration", Implant Dent, 18, pp.480~485 (2009)
  5. Song YD, "Correlation between bone quality evaluated by cone-beam computerized tomography and implant primary stability", Int J Oral Maxillofac Implants, 24, pp.59~64 (2009)
  6. Felsenberg D, "The bone quality framework: determinants of bone strength and their interrelationships, and implications for osteoporosis management", Clin Ther, 27, pp.1~11 (2005)
  7. Bouxsein ML, "Bone quality: where do we go from here?", Osteoporos Int, 14(Suppl 5), pp.S118~S127 (2003)
  8. Friberg B, "Stability measurements of one-stage Brånemark implants during healing in mandibles. A clinical resonance frequency analysis study", Int J Oral Maxillofac Surg, 28, pp.266~272 (1999)
  9. van Steenberghe D, "Impact of systemic diseases and medication on osseointegration", Periodontol 2000, 33, pp.163~171 (2003)
  10. Kaneko T, "Pulsed oscillation technique for assessing the mechanical state of the dental implant-bone interface", Biomaterials, 12, pp.555~560 (1991)
  11. Kaneko T, "Acoustoelectric technique for assessing the mechanical state of the dental implant-bone interface", J Biomed Mater Res, 20, pp.169~176 (1986)
  12. Noyes DH, "Measurement of mechanical mobility of human incisors with sinusoidal forces", J Biomech, 6, pp.439~442 (1973)
  13. Turkyilmaz I, "Influence of bone density on implant stability parameters and implant success: a retrospective clinical study", BMC Oral Health, 8, pp.32~ (2008)
  14. Isoda K, "Relationship between the bone density estimated by cone-beam computed tomography and the primary stability of dental implants", Clin Oral Implants Res, 23, pp.832~836 (2012)
  15. Rozé J, "Correlating implant stability to bone structure", Clin Oral Implants Res, 20, pp.1140~1145 (2009)
  16. Hsu JT, "Effects of the 3D bone-to-implant contact and bone stiffness on the initial stability of a dental implant: micro-CT and resonance frequency analyses", Int J Oral Maxillofac Surg, 42, pp.276~280 (2013)
  17. Kabel J, "Connectivity and the elastic properties of cancellous bone", Bone, 24, pp.115~120 (1999)
  18. Pothuaud L, "Combination of topological parameters and bone volume fraction better predicts the mechanical properties of trabecular bone", J Biomech, 35, pp.1091~1099 (2002)
  19. Rüegsegger P, "A microtomographic system for the nondestructive evaluation of bone architecture", Calcif Tissue Int, 58, pp.24~29 (1996)
  20. Uchiyama T, "A morphometric comparison of trabecular structure of human ilium between microcomputed tomography and conventional histomorphometry", Calcif Tissue Int, 61, pp.493~498 (1997)
  21. Ho JT, "Trabecular bone structural parameters evaluated using dental cone-beam computed tomography: cellular synthetic bones", Biomed Eng Online, 12, pp.115~ (2013)
  22. Panmekiate S, "Comparison of mandibular bone microarchitecture between micro-CT and CBCT images", Dentomaxillofac Radiol, 44, pp.20140322~ (2015)
  23. Kim DS, "A new method for the evaluation of dental implant stability using an inductive sensor", Med Eng Phys, 34, pp.1247~1252 (2012)
  24. Kim DS, "Comparison of dental implant stabilities by impact response and resonance frequencies using artificial bone", Med Eng Phys, 36, pp.715~720 (2014)
  25. Greenstein G, "Clinical recommendations for avoiding and managing surgical complications associated with implant dentistry: a review", J Periodontol, 79, pp.1317~1329 (2008)
  26. Natali AN, "Investigation of viscoelastoplastic response of bone tissue in oral implants press fit process", J Biomed Mater Res B Appl Biomater, 91, pp.868~875 (2009)
  27. Bischof M, "Implant stability measurement of delayed and immediately loaded implants during healing", Clin Oral Implants Res, 15, pp.529~539 (2004)
  28. Miyamoto I, "Influence of cortical bone thickness and implant length on implant stability at the time of surgery--clinical, prospective, biomechanical, and imaging st", Bone, 37, pp.776~780 (2005)
  29. Johns RB, "A multicenter study of overdentures supported by Brånemark implants", Int J Oral Maxillofac Implants, 7, pp.513~522 (1992)
  30. Saadoun AP, "Clinical results and guidelines on Steri-Oss endosseous implants", Int J Periodontics Restorative Dent, 12, pp.486~495 (1992)
  31. Atsumi M, "Methods used to assess implant stability: current status", Int J Oral Maxillofac Implants, 22, pp.743~754 (2007)
  32. Huwiler MA, "Resonance frequency analysis in relation to jawbone characteristics and during early healing of implant installation", Clin Oral Implants Res, 18, pp.275~280 (2007)
  33. Ito Y, "Relevance of resonance frequency analysis to evaluate dental implant stability: simulation and histomorphometrical animal experiments", Clin Oral Implants Res, 19, pp.9~14 (2008)
  34. Nkenke E, "Implant stability and histomorphometry: a correlation study in human cadavers using stepped cylinder implants", Clin Oral Implants Res, 14, pp.601~609 (2003)
  35. de Oliveira RC, "Bone tissue microarchitectural characteristics at dental implant sites. Part 1: identification of clinical-related parameters", Clin Oral Implants Res, 23, pp.981~986 (2012)
  36. Ab-Lazid R, "Pullout strength of cancellous screws in human femoral heads depends on applied insertion torque, trabecular bone microarchitecture and areal bone mineral density", J Mech Behav Biomed Mater, 40, pp.354~361 (2014)
  37. Ab-Lazid R, "Does cancellous screw insertion torque depend on bone mineral density and", or microarchitecture?, 2014, pp.47, 347~353
  38. Baum T, "Automated 3D trabecular bone structure analysis of the proximal femur--prediction of biomechanical strength by CT and DXA", Osteoporos Int, 21, pp.1553~1564 (2010)
  39. Huh KH, "Relationship between two-dimensional and three-dimensional bone architecture in predicting the mechanical strength of the pig mandible", Oral Surg Oral Med Oral Pathol Oral Radiol Endod, 101, pp.363~373 (2006)
  40. Syahrom A, "Mechanical and microarchitectural analyses of cancellous bone through experiment and computer simulation", Med Biol Eng Comput, 49, pp.1393~1403 (2011)

관련 논문AI추천

Three-dimensional microstructure of human alveolar trabecular bone: a micro-computed tomography study김희진, 윤정호, 이지현대한치주과학회 · Journal of Periodontal & Implant Science2017 · 제47권 제1호Relationship between cortical bone thickness and implant stability at the time of surgery and secondary stability after osseointegration measured using resonance frequency analysisIrena Sailer, Kenko Tanaka, Kensuke Yamauchi, Nobuhiro Yoda, Ryosuke Iwama, Shinnosuke Nogami, Tetsu Takahashi대한치주과학회 · Journal of Periodontal & Implant Science2018 · 제48권 제6호Effect of bone quality and implant surgical technique on implant stability quotient (ISQ) value곽재영, 김성균, 윤홍기, 이수영, 허성주대한치과보철학회 · The Journal of Advanced of Prosthodontics2011 · 제3권 제1호A clinical study of alveolar bone quality using the fractal dimension and the implant stability quotient구영, 류인철, 허경회, 허민석, Cheol-Woo Lee, Dae-Hyun Lee, Jeong-Ug Hong대한치주과학회 · Journal of Periodontal & Implant Science2010 · 제40권 제1호Maxillary Sinus의 inferior wall의 피질골 두께가 임플란트의 초기안정성에 미치는 영향에 대한 in vitro 연구강봉서, 권택가, 김석규, 양성은, 이철원대한통합치과학회 · 대한통합치과학회지2013 · 제2권 제2호The relationship between initial implant stability quotient values and bone-to-implant contact ratio in the rabbit tibia김성균, 박인필, 이신재, 이주희대한치과보철학회 · The Journal of Advanced of Prosthodontics2011 · 제3권 제2호

이 간행물의 다른 논문 (6건)