Purpose: In this study, we aimed to evaluate the degree of heat generation when a novel drill design with an irrigation slot was used with metal sleeve-free (MF) and metal sleeveincorporated (MI) surgical guides in an environment similar to that of the actual oral cavity. Methods: A typodont with a missing mandibular right first molar and 21 bovine rib blocks were used. Three-dimensional-printed MF and MI surgical guides, designed for the placement of internal tapered implant fixtures, were used with slot and non-slot drills. The following groups were compared: group 1, MI surgical guide with slot drill; group 2, MI surgical guide with a non-slot drill; and group 3, MF surgical guide with a slot drill. A constant-temperature water bath at 36°C was used. The drilling was performed in 6 stages, and the initial, highest, and lowest temperatures of the cortical bone were measured at each stage using a non-contact infrared thermometer. Results: There were no temperature increases above the initial temperature in any drilling procedure. The only significant difference between the non-slot and slot groups was observed with the use of the first drill in the MI group, with a higher temperature in the non-slot group (P=0.012). When the heat generation during the first and the second drilling was compared in the non-slot group, the heat generation during the first drilling was significantly higher (P<0.001), and there was no significant difference in heat generation between the drills in the slot group. Conclusions: Within the limitations of this study, implant-site preparation with the surgical guide showed no critical increase in the temperature of the cortical bone, regardless of whether there was a slot in the drill. In particular, the slotted drill had a cooling effect during the initial drilling.
Abstract
Purpose: In this study, we aimed to evaluate the degree of heat generation when a novel drill design with an irrigation slot was used with metal sleeve-free (MF) and metal sleeveincorporated (MI) surgical guides in an environment similar to that of the actual oral cavity. Methods: A typodont with a missing mandibular right first molar and 21 bovine rib blocks were used. Three-dimensional-printed MF and MI surgical guides, designed for the placement of internal tapered implant fixtures, were used with slot and non-slot drills. The following groups were compared: group 1, MI surgical guide with slot drill; group 2, MI surgical guide with a non-slot drill; and group 3, MF surgical guide with a slot drill. A constant-temperature water bath at 36°C was used. The drilling was performed in 6 stages, and the initial, highest, and lowest temperatures of the cortical bone were measured at each stage using a non-contact infrared thermometer. Results: There were no temperature increases above the initial temperature in any drilling procedure. The only significant difference between the non-slot and slot groups was observed with the use of the first drill in the MI group, with a higher temperature in the non-slot group (P=0.012). When the heat generation during the first and the second drilling was compared in the non-slot group, the heat generation during the first drilling was significantly higher (P<0.001), and there was no significant difference in heat generation between the drills in the slot group. Conclusions: Within the limitations of this study, implant-site preparation with the surgical guide showed no critical increase in the temperature of the cortical bone, regardless of whether there was a slot in the drill. In particular, the slotted drill had a cooling effect during the initial drilling.
참고문헌 (33)
- Smitkarn P, "The accuracy of single-tooth implants placed using fully digital-guided surgery and freehand implant surgery", J Clin Periodontol, 46, pp.949~957 (2019)
- D’Souza KM, "Types of implant surgical guides in dentistry : a review", J Oral Implantol, 38, pp.643~652 (2012)
- Scherer U, "Template-guided vs. non-guided drilling in site preparation of dental implants", Clin Oral Investig, 19, pp.1339~1346 (2015)
- dos Santos PL, "Evaluation of bone heating, drill deformation, and drill roughness after implant osteotomy: guided surgery and classic drilling procedure", Int J Oral Maxillofac Implants, 29, pp.51~58 (2014)
- Misir AF, "Effect of surgical drill guide on heat generated from implant drilling", J Oral Maxillofac Surg, 67, pp.2663~2668 (2009)
- Oh KC, "In vitro comparison between metal sleeve-free and metal sleeve-incorporated 3D-printed computer-assisted implant surgical guides", Materials (Basel), 14, pp.615~ (2021)
- Boa K, "Intraosseous generation of heat during guided surgical drilling : an ex vivo study of the effect of the temperature of the irrigating fluid", Br J Oral Maxillofac Surg, 54, pp.904~908 (2016)
- Eriksson A, "Thermal injury to bone. A vital-microscopic description of heat effects", Int J Oral Surg, 11, pp.115~121 (1982)
- Ashry A, "Effect of guiding sleeve design on intraosseous heat generation during implant site preparation (in vitro study)", J Prosthodont, 31, pp.147~154 (2021)
- Alhroob KH, "Effect of the use of a surgical guide on heat generation during implant placement : a comparative in vitro study", Dent Med Probl, 58, pp.55~59 (2021)
- Orgev A, "Addition of an irrigation channel to a surgical template to facilitate cooling during implant osteotomy", J Prosthet Dent, 126, pp.164~166 (2021)
- Liu YF, "Numerical and experimental analyses on the temperature distribution in the dental implant preparation area when using a surgical guide", J Prosthodont, 27, pp.42~51 (2018)
- Sener BC, "Effects of irrigation temperature on heat control in vitro at different drilling depths", Clin Oral Implants Res, 20, pp.294~298 (2009)
- Augustin G, "Temperature changes during cortical bone drilling with a newly designed step drill and an internally cooled drill", Int Orthop, 36, pp.1449~1456 (2012)
- Salimov F, "The effects of repeated usage of implant drills on cortical bone temperature, primary", secondary stability and bone healing: a preclinical in vivo micro-CT study, 2020, pp.31, 687~693
- Lajolo C, "Bone heat generated using conventional implant drills versus piezosurgery unit during apical cortical plate perforation", J Periodontol, 89, pp.661~668 (2018)
- Tehemar SH, "Factors affecting heat generation during implant site preparation : a review of biologic observations and future considerations", Int J Oral Maxillofac Implants, 14, pp.127~136 (1999)
- Davidson SR, "Measurement of thermal conductivity of bovine cortical bone", Med Eng Phys, 22, pp.741~747 (2000)
- Ercoli C, "The influence of drill wear on cutting efficiency and heat production during osteotomy preparation for dental implants : a study of drill durability", Int J Oral Maxillofac Implants, 19, pp.335~349 (2004)
- Scarano A, "Infrared thermographic evaluation of temperature modifications induced during implant site preparation with steel vs. zirconia implant drill", J Clin Med, 9, pp.148~ (2020)
- Norton MR, "Bone classification : an objective scale of bone density using the computerized tomography scan", Clin Oral Implants Res, 12, pp.79~84 (2001)
- Shahlaie M, "Bone density assessments of dental implant sites: 1. Quantitative computed tomography", Int J Oral Maxillofac Implants, 18, pp.224~231 (2003)
- Möhlhenrich SC, "Influence of bone density and implant drill diameter on the resulting axial force and temperature development in implant burs and artificial bone: an in vitro study", Oral Maxillofac Surg, 20, pp.135~142 (2016)
- Gehrke SA, "Development of a new drill design to improve the temperature control during the osteotomy for dental implants: a comparative in vitro analysis", Biology (Basel), 9, pp.9~ (2020)
- Zhou M, "Comparison of the accuracy of dental implant placement using static and dynamic computer-assisted systems : an in vitro study", J Stomatol Oral Maxillofac Surg, 122, pp.343~348 (2021)
- Oh HJ, "Effect of implant drill characteristics on heat generation in osteotomy sites : a pilot study", Clin Oral Implants Res, 22, pp.722~726 (2011)
- Marković A, "Effect of surgical drill guide and irrigans temperature on thermal bone changes during drilling implant sites - thermographic analysis on bovine ribs", Vojnosanit Pregl, 73, pp.744~750 (2016)
- Stocchero M, "Bone temperature variation using a 3D-printed surgical guide with internal irrigation", Appl Sci, 11, pp.2588~ (2021)
- Kondo S, "Thermological study of drilling bone tissue with a high-speed drill", Neurosurgery, 46, pp.1162~1168 (2000)
- Isler SC, "The effect of irrigation temperature on bone healing", Int J Med Sci, 8, pp.704~708 (2011)
- Laurito D, "An alternative method to record rising temperatures during dental implant site preparation : a preliminary study using bovine bone", Ann Ist Super Sanita, 46, pp.405~410 (2010)
- Rubayo DD, "Influences of build angle on the accuracy, printing time, and material consumption of additively manufactured surgical templates", J Prosthet Dent, 126, pp.658~663 (2021)
- Smith PN, "Microbial contamination and the sterilization", disinfection of surgical guides used in the placement of endosteal implants, 2011, pp.26, 274~281