The effect of low-speed drilling without irrigation on heat generation: an experimental study

1

조회

0

다운로드

2

인용

0

참고문헌

키워드

주제분류치의학

초록

Objectives: In this study we evaluated heat generation during the low-speed drilling procedure without irrigation. Materials and Methods: Ten artificial bone blocks that were similar to human D1 bone were used in this study. The baseline temperature was 37.0oC. We drilled into 5 artificial bone blocks 60 times at the speed of 50 rpm without irrigation. As a control group, we drilled into an additional 5 artificial bone blocks 60 times at the speed of 1,500 rpm with irrigation. The temperature changes during diameter 2 mm drilling were measured using thermocouples. Results: The mean maximum temperatures during drilling were 40.9oC in the test group and 39.7oC in the control group. Even though a statistically significant difference existed between the two groups, the low-speed drilling did not produce overheating. Conclusion: These findings suggest that low-speed drilling without irrigation may not lead to overheating during drilling.

Abstract

Objectives: In this study we evaluated heat generation during the low-speed drilling procedure without irrigation. Materials and Methods: Ten artificial bone blocks that were similar to human D1 bone were used in this study. The baseline temperature was 37.0oC. We drilled into 5 artificial bone blocks 60 times at the speed of 50 rpm without irrigation. As a control group, we drilled into an additional 5 artificial bone blocks 60 times at the speed of 1,500 rpm with irrigation. The temperature changes during diameter 2 mm drilling were measured using thermocouples. Results: The mean maximum temperatures during drilling were 40.9oC in the test group and 39.7oC in the control group. Even though a statistically significant difference existed between the two groups, the low-speed drilling did not produce overheating. Conclusion: These findings suggest that low-speed drilling without irrigation may not lead to overheating during drilling.

참고문헌 (12)

  1. Kim SJ, "Temperature change in pig rib bone during implant site preparation by low-speed drilling", J Appl Oral Sci, 18, pp.522~527 (2010)
  2. Giro G, "Effect of drilling technique on the early integration of plateau root form endosteal implants: an experimental study in dogs", J Oral Maxillofac Surg, 69, pp.2158~2163 (2011)
  3. Gaspar J, "Osteotomy at low-speed drilling without irrigation versus highspeed drilling with irrigation: an experimental study", Acta Med Port, 26, pp.231~236 (2013)
  4. Misir AF, "Effect of surgical drill guide on heat generated from implant drilling", J Oral Maxillofac Surg, 67, pp.2663~2668 (2009)
  5. Jeong SM, "The effect of guided flapless implant procedure on heat generation from implant drilling", J Craniomaxillofac Surg, 42, pp.725~729 (2014)
  6. Anitua E, "A novel drilling procedure and subsequent bone autograft preparation: a technical note", Int J Oral Maxillofac Implants, 22, pp.138~145 (2007)
  7. Eriksson RA, "The effect of heat on bone regeneration: an experimental study in the rabbit using the bone growth chamber", J Oral Maxillofac Surg, 42, pp.705~711 (1984)
  8. Eriksson RA, "Temperatures during drilling for the placement of implants using the osseointegration technique", J Oral Maxillofac Surg, 44, pp.4~7 (1986)
  9. Harris BH, "Effects of mechanical and thermal fatigue on dental drill performance", Int J Oral Maxillofac Implants, 16, pp.819~826 (2001)
  10. Chacon GE, "Heat production by 3 implant drill systems after repeated drilling and sterilization", J Oral Maxillofac Surg, 64, pp.265~269 (2006)
  11. Sharawy M, "Heat generation during implant drilling: the significance of motor speed", J Oral Maxillofac Surg, 60, pp.1160~1169 (2002)
  12. Carvalho AC, "Evaluation of bone heating, immediate bone cell viability, and wear of high-resistance drills after the creation of implant osteotomies in rabbit tibias", Int J Oral Maxillofac Implants, 26, pp.1193~1201 (2011)

관련 논문AI추천

이 간행물의 다른 논문 (11건)