Late Onset Disassembly of a Modular Neck-stem Component after Cementless Hip Replacement without Dislocation: A Case Report and Review of Literatures

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

초록

Modular femoral prostheses are characterized by a second neck-stem junction. This modularity provides many clinical benefits including hip offset restoration, intraoperative leg length and anteversion adjustment. Although, this extra junction in modular femoral prostheses can contribute to catastrophic consequences like fracture, cold welding, corrosion and fretting of the modularity. However, only few complications related to the modularity itself have been reported in the literature. We report a unique case of neck-stem component dissociation without dislocation of the R-120PCTM Modular Stem (DJO Surgical). Our 71-year-old obese female patient underwent cementless hip replacement 5 years ago. Following radiographic confirmation of neck-stem dissociation open reduction was performed and wiring fixation was applied to secure the neck to the stem. After reduction and fixation, hip joint was stable, and our patient returned to her daily routine 2.5 months postoperatively. The last follow up was at 12 months after surgery with excellent radiographic and clinical evaluation.

Abstract

Modular femoral prostheses are characterized by a second neck-stem junction. This modularity provides many clinical benefits including hip offset restoration, intraoperative leg length and anteversion adjustment. Although, this extra junction in modular femoral prostheses can contribute to catastrophic consequences like fracture, cold welding, corrosion and fretting of the modularity. However, only few complications related to the modularity itself have been reported in the literature. We report a unique case of neck-stem component dissociation without dislocation of the R-120PCTM Modular Stem (DJO Surgical). Our 71-year-old obese female patient underwent cementless hip replacement 5 years ago. Following radiographic confirmation of neck-stem dissociation open reduction was performed and wiring fixation was applied to secure the neck to the stem. After reduction and fixation, hip joint was stable, and our patient returned to her daily routine 2.5 months postoperatively. The last follow up was at 12 months after surgery with excellent radiographic and clinical evaluation.

참고문헌 (10)

  1. Kop AM, "Proximal component modularity in THA--at what cost? An implant retrieval study", Clin Orthop Relat Res, 470, pp.1885~1894 (2012)[DOI]
  2. Ahmed P, "Late disassembly of femoral head and neck of a modular primary total hip arthroplasty", J Orthop Case Rep, 5, pp.8~10 (2015)
  3. Barrack RL, "Complications related to modularity of total hip components", J Bone Joint Surg Br, 75, pp.688~692 (1993)
  4. Kouzelis A, "Dissociation of modular total hip arthroplasty at the neck-stem interface without dislocation", J Orthop Traumatol, 13, pp.221~224 (2012)[DOI]
  5. Shiga T, "Disassembly of a modular femoral component after femoral head prosthetic replacement", J Arthroplasty, 25, pp.659.e17~659.e19 (2010)[DOI]
  6. Star MJ, "Dissociation of modular hip arthroplasty components after dislocation. A report of three cases at differing dissociation levels", Clin Orthop Relat Res, 278, pp.111~115 (1992)
  7. Grupp TM, "Modular titanium alloy neck adapter failures in hip replacement--failure mode analysis and influence of implant material", BMC Musculoskelet Disord, 11, pp.3~ (2010)[DOI]
  8. Bunn A, "Bony impingement limits design-related increases in hip range of motion", Clin Orthop Relat Res, 470, pp.418~427 (2012)[DOI]
  9. Pallini F, "Modular hip stems: determination of disassembly force of a neck-stem coupling", Artif Organs, 31, pp.166~170 (2007)[DOI]
  10. Badarudeen S, "Complications after revision total hip arthroplasty in the medicare population", J Arthroplasty, 32, pp.1954~1958 (2017)[DOI]

관련 논문AI추천

이 간행물의 다른 논문 (8건)