Purpose: We report on the 10-year clinical hip function and radiologic outcomes of patients who underwent hip arthroplasty using a COREN stem. Materials and Methods: A consecutive series of 224 primary cementless hip arthroplasty implantations were performed using a COREN stem between 2009 and 2011; among these, evaluation of 128 hips was performed during a minimum follow-up period of 10 years. The mean age of patients was 65.4 years (range, 40-82 years) and the mean duration of follow-up was 10.8 years (range, 10-12 years). Evaluation of clinical hip function and radiologic implant outcomes was performed according to clinical score, thigh pain, and radiologic analysis. Results: Dramatic improvement of the mean Harris hip score (HHS) from 59.4 preoperatively to 93.5 was observed at the final follow-up (P≤0.01). Stable fixation was demonstrated for all implants with no change in position except for one case of Vancouver type B2 periprosthetic femur fracture. A radiolucent line (RLL) was observed in 16 hips (12.5%). Thigh pain was observed in only two hips (1.6%) at the final follow-up. There were no cases of osteolysis around the stem. The survival rate for the COREN stem was 97.7%. Conclusion: Good long-term survival with excellent clinical and radiological outcomes can be achieved using the COREN femoral stem regardless of Dorr type.

Abstract

Purpose: We report on the 10-year clinical hip function and radiologic outcomes of patients who underwent hip arthroplasty using a COREN stem. Materials and Methods: A consecutive series of 224 primary cementless hip arthroplasty implantations were performed using a COREN stem between 2009 and 2011; among these, evaluation of 128 hips was performed during a minimum follow-up period of 10 years. The mean age of patients was 65.4 years (range, 40-82 years) and the mean duration of follow-up was 10.8 years (range, 10-12 years). Evaluation of clinical hip function and radiologic implant outcomes was performed according to clinical score, thigh pain, and radiologic analysis. Results: Dramatic improvement of the mean Harris hip score (HHS) from 59.4 preoperatively to 93.5 was observed at the final follow-up (P≤0.01). Stable fixation was demonstrated for all implants with no change in position except for one case of Vancouver type B2 periprosthetic femur fracture. A radiolucent line (RLL) was observed in 16 hips (12.5%). Thigh pain was observed in only two hips (1.6%) at the final follow-up. There were no cases of osteolysis around the stem. The survival rate for the COREN stem was 97.7%. Conclusion: Good long-term survival with excellent clinical and radiological outcomes can be achieved using the COREN femoral stem regardless of Dorr type.

참고문헌 (27)

  1. Sanfilippo JA, "Implants for total hip arthroplasty", Expert Rev Med Devices, 3, pp.769~776 (2006)[DOI]
  2. Suckel A, "Longterm results for the uncemented Zweymuller", Alloclassic hip endoprosthesis. A 15-year minimum follow-up of 320 hip operations, 2009, pp.24, 846~853
  3. Korovessis P, "High medium-term survival of Zweymüller SLR-plus stem used in femoral revision", Clin Orthop Relat Res, 467, pp.2032~2040 (2009)[DOI]
  4. Korovessis P, "High medium-term survivorship and durability of Zweymüller-Plus total hip arthroplasty", Arch Orthop Trauma Surg, 131, pp.603~611 (2011)[DOI]
  5. 김희중, "Cementless Total Hip Arthroplasty Using the COREN Hip System: A Minimum Five-Year Follow-up Study", Hip and Pelvis, 30(3), pp.162~167 (2018), 대한고관절학회[DOI]
  6. Zweymüller KA, "Radiolucent lines and osteolysis along tapered straight cementless titanium hip stems : a comparison of 6-year and 10-year follow-up results in 95 patients", Acta Orthop, 77, pp.871~876 (2006)[DOI]
  7. 박윤수, "COREN HIP System을 이용한 무시멘트 고관절 전치환술", Hip and Pelvis, 19(4), pp.457~462 (2007), 대한고관절학회
  8. 김상민, "COREN Hip System을 이용한 무시멘트 고관절 전치환술의 최소 7년 추시 결과", Hip and Pelvis, 25(3), pp.173~181 (2013), 대한고관절학회[DOI]
  9. Lim YW, "Enhanced cell integration to titanium alloy by surface treatment with microarc oxidation : a pilot study", Clin Orthop Relat Res, 467, pp.2251~2258 (2009)[DOI]
  10. 이중명, "Bencox® Hip Stem을 이용한 인공 고관절 치환술의 최소 5년 추시 결과", 대한정형외과학회지, 49(2), pp.126~132 (2014), 대한정형외과학회
  11. Harris WH, "Traumatic arthritis of the hip after dislocation and acetabular fractures : treatment by mold arthroplasty. An endresult study using a new method of result evaluation", J Bone Joint Surg Am, 51, pp.737~755 (1969)
  12. Dorr LD, "Structural and cellular assessment of bone quality of proximal femur", Bone, 14, pp.231~242 (1993)[DOI]
  13. Engh CA, "The influence of stem size and extent of porous coating on femoral bone resorption after primary cementless hip arthroplasty", Clin Orthop Relat Res, (231), pp.7~28 (1988)[DOI]
  14. Gruen TA, ""Modes of failure"of cemented stem-type femoral components : a radiographic analysis of loosening", Clin Orthop Relat Res, (141), pp.17~27 (1979)[DOI]
  15. Davies MS, "Migration of the uncemented CLS femoral component", Orthopedics, 22, pp.225~228 (1999)[DOI]
  16. Ström H, "Early migration pattern of the uncemented CLS stem in total hip arthroplasties", Clin Orthop Relat Res, 454, pp.127~132 (2007)[DOI]
  17. Kubo T, "Cementless Lord total hip arthroplasty : cup loosening common after minimum 10-year follow-up of 103 hips", Acta Orthop Scand, 72, pp.585~590 (2001)[DOI]
  18. 강준순, "Clinical and Radiological Outcomes of Rectangular Tapered Cementless Stem According to Proximal Femoral Geometry in Elderly Asian Patients", Hip and Pelvis, 31(4), pp.224~231 (2019), 대한고관절학회
  19. Belmont PJ Jr, "Results of the anatomic medullary locking total hip arthroplasty at a minimum of twenty years. A concise follow-up of previous reports", J Bone Joint Surg Am, 90, pp.1524~1530 (2008)[DOI]
  20. Grant P, "Total hip arthroplasty with the Lord prosthesis. A long-term follow-up study", J Bone Joint Surg Am, 86, pp.2636~2641 (2004)[DOI]
  21. Vresilovic EJ, "Incidence of thigh pain after uncemented total hip arthroplasty as a function of femoral stem size", J Arthroplasty, 11, pp.304~311 (1996)[DOI]
  22. Engh CA, "Porous-coated hip replacement. The factors governing bone ingrowth, stress shielding, and clinical results", J Bone Joint Surg Br, 69, pp.45~55 (1987)[DOI]
  23. Biemond JE, "Longterm survivorship analysis of the cementless Spotorno femoral component in patients less than 50 years of age", J Arthroplasty, 26, pp.386~390 (2011)[DOI]
  24. Müller LA, "Seventeen-year survival of the cementless CLS Spotorno stem", Arch Orthop Trauma Surg, 130, pp.269~275 (2010)[DOI]
  25. Keeling P, "Long-term survival of the cemented exeter universal stem in patients 50 years and younger : an update on 130 hips", J Arthroplasty, 35, pp.1042~1047 (2020)[DOI]
  26. Aldinger PR, "Survival of the cementless Spotorno stem in the second decade", Clin Orthop Relat Res, 467, pp.2297~2304 (2009)[DOI]
  27. Lim SJ, "High incidence of stem loosening in association with periprosthetic femur fractures in previously well-fixed cementless gritblasted tapered-wedge stems", Int Orthop, 39, pp.1689~1693 (2015)[DOI]

관련 논문AI추천

이 간행물의 다른 논문 (8건)