Background: Respiratory muscle strength in patients with an artificial airway is commonly assessed as the maximal inspiratory pressure (MIP) and is measured using analogue or digital manometers. Recently, new electronic loading devices have been proposed to measure respiratory muscle strength. This study evaluates the agreement between the MIPs measured by a digital manometer and those according to an electronic loading device in patients being weaned from mechanical ventilation.Methods: In this prospective study, the standard MIP was obtained using a protocol adapted from Marini, in which repetitive inspiratory efforts were performed against an occluded airway with a one-way valve and were recorded with a digital manometer for 40 seconds (MIPDM). The MIP measured using the electronic loading device (MIPELD) was obtained from repetitively tapered flow resistive inspirations sustained for at least 2 seconds during a 40-second test. The agreement between the results was verified by a Bland-Altman analysis.Results: A total of 39 subjects (17 men, 55.4±17.7 years) was enrolled. Although a strong correlation between MIPDM and MIPELD (R=0.73, P<0.001) was observed, the Bland-Altman analysis showed a high bias of –47.4 (standard deviation, 22.3 cm H2O; 95% confidence interval, –54.7 to –40.2 cm H2O).Conclusions: The protocol of repetitively tapering flow resistive inspirations to measure the MIP with the electronic loading device is not in agreement with the standard protocol using one-way valve inspiratory occlusion when applied in poorly cooperative patients being weaned from mechanical ventilation.

Abstract

Background: Respiratory muscle strength in patients with an artificial airway is commonly assessed as the maximal inspiratory pressure (MIP) and is measured using analogue or digital manometers. Recently, new electronic loading devices have been proposed to measure respiratory muscle strength. This study evaluates the agreement between the MIPs measured by a digital manometer and those according to an electronic loading device in patients being weaned from mechanical ventilation.Methods: In this prospective study, the standard MIP was obtained using a protocol adapted from Marini, in which repetitive inspiratory efforts were performed against an occluded airway with a one-way valve and were recorded with a digital manometer for 40 seconds (MIPDM). The MIP measured using the electronic loading device (MIPELD) was obtained from repetitively tapered flow resistive inspirations sustained for at least 2 seconds during a 40-second test. The agreement between the results was verified by a Bland-Altman analysis.Results: A total of 39 subjects (17 men, 55.4±17.7 years) was enrolled. Although a strong correlation between MIPDM and MIPELD (R=0.73, P<0.001) was observed, the Bland-Altman analysis showed a high bias of –47.4 (standard deviation, 22.3 cm H2O; 95% confidence interval, –54.7 to –40.2 cm H2O).Conclusions: The protocol of repetitively tapering flow resistive inspirations to measure the MIP with the electronic loading device is not in agreement with the standard protocol using one-way valve inspiratory occlusion when applied in poorly cooperative patients being weaned from mechanical ventilation.

참고문헌 (29)

  1. Medrinal C, "Respiratory weakness after mechanical ventilation is associated with one-year mortality : a prospective study", Crit Care, 20, pp.231~ (2016)
  2. De Jonghe B, "Respiratory weakness is associated with limb weakness and delayed weaning in critical illness", Crit Care Med, 35, pp.2007~2015 (2007)
  3. Dres M, "Diaphragm dysfunction during weaning from mechanical ventilation : an underestimated phenomenon with clinical implications", Crit Care, 22, pp.73~ (2018)
  4. Laveneziana P, "ERS statement on respiratory muscle testing at rest and during exercise", Eur Respir J, 53, pp.1801214~ (2019)
  5. Caruso P, "Diagnostic methods to assess inspiratory and expiratory muscle strength", J Bras Pneumol, 41, pp.110~123 (2015)
  6. Marini JJ, "Estimation of inspiratory muscle strength in mechanically ventilated patients : the measurement of maximal inspiratory pressure", J Crit Care, 1, pp.32~38 (1986)
  7. Truwit JD, "Validation of a technique to assess maximal inspiratory pressure in poorly cooperative patients", Chest, 102, pp.1216~1219 (1992)
  8. American Thoracic Society, "ATS", ERS Statement on respiratory muscle testing, 2002, pp.166, 518~624
  9. Medrinal C, "Reliability of respiratory pressure measurements in ventilated and non-ventilated patients in ICU : an observational study", Ann Intensive Care, 8, pp.14~ (2018)
  10. Pinheiro Filho GR, "Comparison and effects of two different airway occlusion times during measurement of maximal inspiratory pressure in adult intensive care unit neurological patients", Rev Bras Ter Intensiva, 22, pp.33~39 (2010)
  11. de Souza LC, "Evaluation of the inspiratory pressure using a digital vacuometer in mechanically ventilated patients : analysis of the time to achieve the inspiratory peak", Respir Care, 57, pp.257~262 (2012)
  12. Guimarães FS, "Maximal inspiratory pressure evaluation among non-cooperative critical patients : comparison between two methods", Braz J Phys Ther, 11, pp.233~238 (2007)
  13. Minahan C, "Repeated-sprint cycling does not induce respiratory muscle fatigue in active adults : measurements from the powerbreathe® inspiratory muscle trainer", J Sports Sci Med, 14, pp.233~238 (2015)
  14. Spurling DA, "Respiratory muscle training device", United States patent US 8.459.255 B2. 2013 Jun 11, ~
  15. Langhan ML, "Implementation of newly adopted technology in acute care settings : a qualitative analysis of clinical staff", J Med Eng Technol, 39, pp.44~53 (2015)
  16. Hendra KP, "Development and prospective validation of a model for predicting weaning in chronic ventilator dependent patients", BMC Pulm Med, 3, pp.3~ (2003)
  17. Zhang JB, "Use of the modified Glasgow Coma Scale score to guide sequential invasive-noninvasive mechanical ventilation weaning in patients with AECOPD and respiratory failure", Exp Ther Med, 20, pp.1441~1446 (2020)
  18. de Souza LC, "Evaluation of a new index of mechanical ventilation weaning : the timed inspiratory effort", J Intensive Care Med, 30, pp.37~43 (2015)
  19. Caruso P, "The unidirectional valve is the best method to determine maximal inspiratory pressure during weaning", Chest, 115, pp.1096~1101 (1999)
  20. Giavarina D, "Understanding Bland Altman analysis", Biochem Med(Zagreb), 25, pp.141~151 (2015)
  21. Bland JM, "Statistical methods for assessing agreement between two methods of clinical measurement", Lancet, 327, pp.307~310 (1986)
  22. Ratnovsky A, "Mechanics of respiratory muscles", Respir Physiol Neurobiol, 163, pp.82~89 (2008)
  23. Whitelaw WA, "Occlusion pressure as a measure of respiratory center output in conscious man", Respir Physiol, 23, pp.181~199 (1975)
  24. Silva PE, "Assessment of maximum dynamic inspiratory pressure", Respir Care, 63, pp.1231~1238 (2018)
  25. Marshall R, "Relationships between stimulus and work of breathing at different lung volumes", J Appl Physiol, 17, pp.917~921 (1962)
  26. McDonald T, "Inspiratory muscle training is feasible and safe for patients with acute spinal cord injury", J Spinal Cord Med, 42, pp.220~227 (2019)
  27. Langer D, "Measurement validity of an electronic inspiratory loading device during a loaded breathing task in patients with COPD", Respir Med, 107, pp.633~635 (2013)
  28. Bissett BM, "Inspiratory muscle training to enhance recovery from mechanical ventilation : a randomised trial", Thorax, 71, pp.812~819 (2016)
  29. Vorona S, "Inspiratory muscle rehabilitation in critically ill adults : a systematic review and meta-analysis", Ann Am Thorac Soc, 15, pp.735~744 (2018)

관련 논문AI추천

Thoracic fluid content by electrical cardiometry versus diaphragmatic excursion by ultrasound for the prediction of weaning success in patients with lung congestionAhmed Mostafa Abdelhamid, Enas Wageh Mahdy, Samar Rafik Amin, Shawky Meselhy Elshaer대한중환자의학회 · Acute and Critical Care2025 · 제40권 제4호Predictive value of the negative inspiratory force index as a predictor of weaning success: a crosssectional studyCong Quyet Thang, Duong Minh Cuong, Nguyen Thu, Tran Viet Duc, Vu Phuong Hoang대한중환자의학회 · Acute and Critical Care2020 · 제35권 제4호Comparison of clinical utility between diaphragm excursion and thickening change using ultrasonography to predict extubation success유정완, Ho Cheol Kim, Jong Deog Lee, Seung Jun Lee대한내과학회 · The Korean Journal of Internal Medicine2018 · 제33권 제2호Clinical Application of Modified Burns Wean Assessment Program Scores at First Spontaneous Breathing Trial in Weaning Patients from Mechanical Ventilation이광하, Eun Suk Jeong대한중환자의학회 · Acute and Critical Care2018 · 제33권 제4호Diaphragm ultrasound in weaning from mechanical ventilation: a last step to predict successful extubation?Costanzo Domenica Di, Esquinas Antonio, Mazza Mariano대한중환자의학회 · Acute and Critical Care2022 · 제37권 제4호The role of diaphragmatic thickness measurement in weaning prediction and its comparison with rapid shallow breathing index: a single-center experienceDhruva Chaudhry, Lokesh Kumar Lalwani, Manjunath B Govindagoudar, Mukesh Sharma, Pawan Kumar Singh대한중환자의학회 · Acute and Critical Care2022 · 제37권 제3호

이 간행물의 다른 논문 (29건)