세가지 효소면역측정 시약을 이용한 C형 간염 바이러스 항체 검사의 비교
Comparison of Three Third-generation Anti-HCV Enzyme Immunoassay Tests
1
조회
0
다운로드
0
인용
0
참고문헌
키워드
초록
The aim of this study was to evaluate domestic enzyme immunoassay(EIA) kit ?LG HCD 3.0?(LG) for the detection of antibody to hepatitis C virus(anti-HCV) in comparision with Axsym HCV version 3.0(Axsym), Cobas Core anti-HCV EIA(Cobas). Cobas kit shows better clear distinction between positive and negative by signal/cutoff ratio(S/C), but it also reveal relatively high false positive rate. The concordance rate of test results between LG and Axsym was 96.2%, between LG and Cobas was 95.5%, and total agreement between three EIA kit was 93.9%. LG were relative poor distinction between positive and negative results, but it could be applied clinically as a screening tool for hepatitis C in general population. The S/C of one false negative result by LG was 0.91, and false positive were less than 4.0, therefore we concluded it is necessary to confirm by immunoblotting assay when S/C were between 0.8 and 4.0.
Abstract
The aim of this study was to evaluate domestic enzyme immunoassay(EIA) kit ?LG HCD 3.0?(LG) for the detection of antibody to hepatitis C virus(anti-HCV) in comparision with Axsym HCV version 3.0(Axsym), Cobas Core anti-HCV EIA(Cobas). Cobas kit shows better clear distinction between positive and negative by signal/cutoff ratio(S/C), but it also reveal relatively high false positive rate. The concordance rate of test results between LG and Axsym was 96.2%, between LG and Cobas was 95.5%, and total agreement between three EIA kit was 93.9%. LG were relative poor distinction between positive and negative results, but it could be applied clinically as a screening tool for hepatitis C in general population. The S/C of one false negative result by LG was 0.91, and false positive were less than 4.0, therefore we concluded it is necessary to confirm by immunoblotting assay when S/C were between 0.8 and 4.0.
