Evaluation of the VIDAS Anti-HCV Assay for Detection of Hepatitis C Virus Infection

1

조회

0

다운로드

1

인용

0

참고문헌

키워드

주제분류병리학

초록

Background: Anti-hepatitis C virus antibody (anti-HCV) assays are recommended for screening HCV-infected persons. The VIDAS Anti-HCV Assay (bioMérieux, France), based on the enzyme-linked fluorescence test principle, was recently introduced in Korea. We evaluated the clinical performance of the VIDAS assay. Methods: One hundred HCV-positive and 1,002 HCV-negative blood samples confirmed by Architect anti-HCV (Abbott Laboratories, USA) and COBAS TaqMan HCV real-time PCR (Roche Diagnostics, USA) or the Procleix Ultrio Plus Assay (Gen-Probe Incorporated, USA) were obtained from the Human Serum Bank (HSB) and tested by VIDAS. In case of discrepant results, we conducted a recombinant immunoblot assay (RIBA). Results: The agreement rates for known HCV-positive and HCV-negative samples between the VIDAS assay and the HSB testing were 100% (95% confidence interval [CI]: 96.4-100%) and 99.5% (95% CI: 98.8-99.8%), respectively. One of the five discrepant samples was positive for Core 2+ and NS3-2 2+ reactivity, two samples were negative, and the other two were indeterminate regarding NS4 2+ reactivity in RIBA. We observed a significant but weak positive correlation between the titers of VIDAS and Architect assays (r=0.315, P<0.001). Conclusions: The VIDAS anti-HCV assay, developed on the VIDAS automated immunoassay platform based on the ready-to-use, single-sample test concept may be useful in small-to-medium-sized laboratories. It showed good agreement with Architect anti-HCV and COBAS PCR assays and is therefore useful for detection of HCV infection. Weakly test-positive (ambiguous) samples require additional testing by another anti-HCV, RIBA, or HCV RNA assay.

Abstract

Background: Anti-hepatitis C virus antibody (anti-HCV) assays are recommended for screening HCV-infected persons. The VIDAS Anti-HCV Assay (bioMérieux, France), based on the enzyme-linked fluorescence test principle, was recently introduced in Korea. We evaluated the clinical performance of the VIDAS assay. Methods: One hundred HCV-positive and 1,002 HCV-negative blood samples confirmed by Architect anti-HCV (Abbott Laboratories, USA) and COBAS TaqMan HCV real-time PCR (Roche Diagnostics, USA) or the Procleix Ultrio Plus Assay (Gen-Probe Incorporated, USA) were obtained from the Human Serum Bank (HSB) and tested by VIDAS. In case of discrepant results, we conducted a recombinant immunoblot assay (RIBA). Results: The agreement rates for known HCV-positive and HCV-negative samples between the VIDAS assay and the HSB testing were 100% (95% confidence interval [CI]: 96.4-100%) and 99.5% (95% CI: 98.8-99.8%), respectively. One of the five discrepant samples was positive for Core 2+ and NS3-2 2+ reactivity, two samples were negative, and the other two were indeterminate regarding NS4 2+ reactivity in RIBA. We observed a significant but weak positive correlation between the titers of VIDAS and Architect assays (r=0.315, P<0.001). Conclusions: The VIDAS anti-HCV assay, developed on the VIDAS automated immunoassay platform based on the ready-to-use, single-sample test concept may be useful in small-to-medium-sized laboratories. It showed good agreement with Architect anti-HCV and COBAS PCR assays and is therefore useful for detection of HCV infection. Weakly test-positive (ambiguous) samples require additional testing by another anti-HCV, RIBA, or HCV RNA assay.

참고문헌 (16)

  1. "Hepatitis C"[URL]
  2. "Guidelines for the screening care and treatment of persons with hepatitis c infection" (2014)[URL]
  3. European Association for Study of L, "EASL Clinical Practice Guidelines: management of hepatitis C virus infection", J Hepatol, 60, pp.392~420 (2014)
  4. Kamili S, "Laboratory diagnostics for hepatitis C virus infection", Clin Infect Dis, 55(Sl), pp.S43~S48 (2012)
  5. Seignères B, "Multicenter clinical evaluation of the new 3rd generation assay for detection of antibodies against hepatitis C virus on the VIDAS system", J Clin Virol, 78, pp.20~26 (2016)
  6. Alborino F, "Multicenter evaluation of a fully automated third-generation anti-HCV antibody screening test with excellent sensitivity and specificity", Med Microbiol Immunol, 200, pp.77~83 (2011)
  7. Kim S, "Clinical performance evaluation of four automated chemiluminescence immunoassays for hepatitis C virus antibody detection", J Clin Microbiol, 46, pp.3919~3923 (2008)
  8. Park Y, "Performance evaluation of the Vitros anti-hepatitis C virus antibody assay for use in clinical laboratories", Clin Biochem, 45, pp.175~177 (2012)
  9. Watterson JM, "Evaluation of the Ortho-Clinical Diagnostics Vitros ECi Anti-HCV test: comparison with three other methods", J Clin Lab Anal, 21, pp.162~166 (2007)
  10. Berger A, "Evaluation of the new ARCHITECT anti-HCV screening test under routine laboratory conditions", J Clin Virol, 43, pp.158~161 (2008)
  11. Kiely P, "The significance of third-generation HCV RIBA-indeterminate, RNA-negative results in voluntary blood donors screened with sequential third-generation immunoassays", Transfusion, 44, pp.349~358 (2004)
  12. Lemaire JM, "HCV RNA in blood donors with isolated reactivities by third-generation RIBA", Transfusion, 40, pp.867~870 (2000)
  13. Pawlotsky JM, "Significance of indeterminate third-generation hepatitis C virus recombinant immunoblot assay", J Clin Microbiol, 34, pp.80~83 (1996)
  14. Alter MJ, "Guidelines for laboratory testing and result reporting of antibody to hepatitis C virus. Centers for Disease Control and Prevention", MMWR Recomm Rep, 52, pp.1~13 (2003)
  15. Centers for Disease Control and Prevention, "Testing for HCV infection: an update of guidance for clinicians and laboratorians", MMWR Morb Mortal Wkly Rep, 62, pp.362~365 (2013)
  16. Yang R, "Performance evaluation and comparison of the newly developed Elecsys anti-HCV II assay with other widely used assays", Clin Chim Acta, 426, pp.95~101 (2013)

관련 논문AI추천

The Role of the Signal-to-Cutoff Ratio in Automated Anti-HCV Chemiluminescent Immunoassays by Referring to the Nucleic Acid Amplification Test and the Recombinant Immunoblot AssayYun Ji Hong대한진단검사의학회 · Annals of Laboratory Medicine2018 · 제38권 제5호세가지 효소면역측정 시약을 이용한 C형 간염 바이러스 항체 검사의 비교김경동, 문진영, 이채훈, 조희순영남대학교 의과대학 · Journal of Yeungnam Medical Science1998 · 제15권 제1호Evaluation of ARCHITECT HCV Core Antigen Assay강정은, 김신영, 김형회, 손한철, 송두열, 이은엽, 황상현대한진단검사의학회 · Annals of Laboratory Medicine2010 · 제30권 제6호The diagnostic significance of hepatitis C virus antibody levels for chronic hepatitis C virus infection김정희, Dong Joon Kim, Hyoung Su Kim, Jang Han Jung, Jin Gu Kang, Ji Won Park, Ki Tae Suk, Myoung Kuk Jang, Myung Seok Lee, Sang Hoon Park, Sung Eun Kim대한내과학회 · The Korean Journal of Internal Medicine2023 · 제38권 제3호Development of a Rapid Automated Fluorescent Lateral Flow Immunoassay to Detect Hepatitis B Surface Antigen (HBsAg), Antibody to HBsAg, and Antibody to Hepatitis C권민석, 김용구, 류지형, 문정대, 박기현, 배현진, 오은지, 윤소정, 이정민, 이혜영, 정봉수, 정주희, 최애란, 한경자, 황민웅대한진단검사의학회 · Annals of Laboratory Medicine2018 · 제38권 제6호Performance Evaluation of the OraQuick Hepatitis C Virus Rapid Antibody Test강은숙, 김명희, 김진욱, 박경운, 우병철, 차영주, 황유성, Quehn Park대한진단검사의학회 · Annals of Laboratory Medicine2013 · 제33권 제3호

이 간행물의 다른 논문 (19건)