Background and Objectives We compared functional MRI acquisition methods of sparse temporal acquisition (STA) and continuous acquisition (CA) to estimate the effect of MRI scanner background noise (SBN) on blood oxygen level-dependent (BOLD) activation of cortical and subcortical auditory centers during auditory stimulation. Subjects and Method Fourteen healthy subjects (eight males, age 30.6±4.7 years) were presented with classical music in a block paradigm (36 s on/off) in two STA [repitition time (TR)=12 s, 60 volumes] and two CA (TR=2 s, 360 volumes) functional MRI sessions. To account for the sample size difference, an additional volume-matched continuous dataset (CAm) was generated by matching CA to 60 volumes of STA. A group-level analysis based on BOLD activation maps was performed. Percent signal change (PSC), T-statistic values and signal variability in cortical and subcortical auditory regions of interest (ROIs) were calculated from individual activation maps and compared between the STA, CA, and CAm. Results The group analysis showed activation in the primary and secondary auditory cortices in all datasets. However, the activation of subcortical auditory centers above the accepted threshold was only observed in STA. STA (less SBN) showed higher PSCs and T-statistic values in all ROIs except planum temporale when compared to CAm. However, there was no difference in signal variability among the datasets. Conclusion Our results suggest that SBN should be considered as a significant confounder in auditory-evoked functional MRI studies particularly in the activation of subcortical auditory centers, and that STA can be an effective imaging method for reducing the effect of SBN.
Abstract
Background and Objectives We compared functional MRI acquisition methods of sparse temporal acquisition (STA) and continuous acquisition (CA) to estimate the effect of MRI scanner background noise (SBN) on blood oxygen level-dependent (BOLD) activation of cortical and subcortical auditory centers during auditory stimulation. Subjects and Method Fourteen healthy subjects (eight males, age 30.6±4.7 years) were presented with classical music in a block paradigm (36 s on/off) in two STA [repitition time (TR)=12 s, 60 volumes] and two CA (TR=2 s, 360 volumes) functional MRI sessions. To account for the sample size difference, an additional volume-matched continuous dataset (CAm) was generated by matching CA to 60 volumes of STA. A group-level analysis based on BOLD activation maps was performed. Percent signal change (PSC), T-statistic values and signal variability in cortical and subcortical auditory regions of interest (ROIs) were calculated from individual activation maps and compared between the STA, CA, and CAm. Results The group analysis showed activation in the primary and secondary auditory cortices in all datasets. However, the activation of subcortical auditory centers above the accepted threshold was only observed in STA. STA (less SBN) showed higher PSCs and T-statistic values in all ROIs except planum temporale when compared to CAm. However, there was no difference in signal variability among the datasets. Conclusion Our results suggest that SBN should be considered as a significant confounder in auditory-evoked functional MRI studies particularly in the activation of subcortical auditory centers, and that STA can be an effective imaging method for reducing the effect of SBN.
참고문헌 (20)
- Price DL, "Investigation of acoustic noise on 15 MRI scanners from 0.2 T to 3T", J Magn Reson Imaging, 13(2), pp.288~293 (2001)
- Van de Moortele PF, "Latencies in fMRI time-series: effect of slice acquisition order and perception", NMR Biomed, 10(4-5), pp.230~236 (1997)
- Nam EC, "Development of sound measurement systems for auditory functional magnetic resonance imaging", Magn Reson Imaging, 26(5), pp.715~720 (2008)
- Hall DA, "“Sparse” temporal sampling in auditory fMRI", Hum Brain Mapp, 7(3), pp.213~223 (1999)
- Patterson RD, "The processing of temporal pitch and melody information in auditory cortex", Neuron, 36(4), pp.767~776 (2002)
- Friston KJ, "A multivariate analysis of PET activation studies", Hum Brain Mapp, 4(2), pp.140~151 (1996)
- Eden GF, "Utilizing hemodynamic delay and dispersion to detect fMRI signal change without auditory interference: the behavior interleaved gradients technique", Magn Reson Med, 41(1), pp.13~20 (1999)
- Arthurs OJ, "Linear coupling between functional magnetic resonance imaging and evoked potential amplitude in human somatosensory cortex", Neuroscience, 101(4), pp.803~806 (2000)
- Arthurs OJ, "What aspect of the fMRI BOLD signal best reflects the underlying electrophysiology in human somatosensory cortex?", Clin Neurophysiol, 114(7), pp.1203~1209 (2003)
- Gaab N, "Assessing the influence of scanner background noise on auditory processing. I. An fMRI study comparing three experimental designs with varying degrees of scanner noise", Hum Brain Mapp, 28(8), pp.703~720 (2007)
- Blackman GA, "Reducing the effects of background noise during auditory functional magnetic resonance imaging of speech processing: qualitative and quantitative comparisons between two image acquisition schemes and noise cancellation", J Speech Lang Hear Res, 54(2), pp.693~704 (2011)
- Peelle JE, "Evaluating an acoustically quiet EPI sequence for use in fMRI studies of speech and auditory processing", Neuroimage, 52(4), pp.1410~1419 (2010)
- 태우석, "피질 및 피질하청각중추 활성화 검출을 위한 기능적 자기공명영상 연구: 1.5-T MRI 스캐너를 위한 표준 프로토콜 개발", 대한영상의학회지, 61(5), pp.279~289 (2009), 대한영상의학회
- Robson MD, "Measurements of the temporal fMRI response of the human auditory cortex to trains of tones", Neuroimage, 7(3), pp.185~198 (1998)
- Schwarzbauer C, "Interleaved silent steady state (ISSS) imaging: a new sparse imaging method applied to auditory fMRI", Neuroimage, 29(3), pp.774~782 (2006)
- Rodd JM, "Dissociating frontotemporal contributions to semantic ambiguity resolution in spoken sentences", Cereb Cortex, 22(8), pp.1761~1773 (2012)
- Overath T, "Sensitivity to temporal modulation rate and spectral bandwidth in the human auditory system: fMRI evidence", J Neurophysiol, 107(8), pp.2042~2056 (2012)
- Kovacs S, "Activation of cortical and subcortical auditory structures at 3 T by means of a functional magnetic resonance imaging paradigm suitable for clinical use", Invest Radiol, 41(2), pp.87~96 (2006)
- Langers DR, "Lateralization, connectivity and plasticity in the human central auditory system", Neuroimage, 28(2), pp.490~499 (2005)
- Schlaug G, "In vivo evidence of structural brain asymmetry in musicians", Science, 267(5198), pp.699~701 (1995)