프로그램과 실이 측정을 이용한 보청기 적합의 임상적 유용성의 비교
Comparison of Clinical Usefulness of Program-Assisted and Real Ear Measurement-Assisted Hearing Aids Fitting
1
조회
0
다운로드
1
인용
0
참고문헌
키워드
초록
Background and Objectives The main objectives of this study were to determine the clinicalusefulness of the program-assisted and real ear measurement (REM)-assisted fitting ofhearing aids. Subjects and Method Fifteen participants with moderate to moderately severe hearing losswere enrolled in this study. Objective and subjective fitting results were assessed to comparethe benefits between the program-assisted fitting (using a software fitting program) and theREM-assisted fitting. Real ear insertion gain (REIG), sound-field audiometry using warble tone,and Korean Hearing in Noise Test (K-HINT) were performed as objective tests. Sound qualityrating was performed as a subjective test. Results In the program fitting, 48.89% of fitting points failed to come within ±10 dB of theREIG target. In the REM fitting, however, the percentage of failure significantly decreased to23.33% (p=0.013). In K-HINT test, the reception threshold for speech in quiet situation significantlydecreased from 50.1 dB HL with the program fitting to 44.7 dB HL after the REM fitting(p<0.001). In front noise condition, signal-to-noise ratio improved from 4.53 dB to 3.50 dBwith the REM fitting without statistical significance (p=0.099). In the sound quality rating, theREM fitting (4.27±0.56) showed a significantly better sound quality ratings than the programfitting (3.69±0.74) (p=0.017). Conclusion The REM fitting showed better results in both subjective and objective measurementsthan the program fitting
Abstract
Background and Objectives The main objectives of this study were to determine the clinicalusefulness of the program-assisted and real ear measurement (REM)-assisted fitting ofhearing aids. Subjects and Method Fifteen participants with moderate to moderately severe hearing losswere enrolled in this study. Objective and subjective fitting results were assessed to comparethe benefits between the program-assisted fitting (using a software fitting program) and theREM-assisted fitting. Real ear insertion gain (REIG), sound-field audiometry using warble tone,and Korean Hearing in Noise Test (K-HINT) were performed as objective tests. Sound qualityrating was performed as a subjective test. Results In the program fitting, 48.89% of fitting points failed to come within ±10 dB of theREIG target. In the REM fitting, however, the percentage of failure significantly decreased to23.33% (p=0.013). In K-HINT test, the reception threshold for speech in quiet situation significantlydecreased from 50.1 dB HL with the program fitting to 44.7 dB HL after the REM fitting(p<0.001). In front noise condition, signal-to-noise ratio improved from 4.53 dB to 3.50 dBwith the REM fitting without statistical significance (p=0.099). In the sound quality rating, theREM fitting (4.27±0.56) showed a significantly better sound quality ratings than the programfitting (3.69±0.74) (p=0.017). Conclusion The REM fitting showed better results in both subjective and objective measurementsthan the program fitting
참고문헌 (13)
- Gatehouse S, "The time course and magnitude of perceptual acclimatization to frequency responses: evidence from monaural fitting of hearing aids", J Acoust Soc Am, 92(3), pp.1258~1268 (1992)
- Campos PD, "Comparison: real and simulated ear insertion gain", Braz J Otorhinolaryngol, 77(5), pp.555~558 (2011)
- Aazh H, "The value of routine real ear measurement of the gain of digital hearing aids", J Am Acad Audiol, 18(8), pp.653~664 (2007)
- Kuk F, "Verifying the output of digital nonlinear hearing instruments", Hear Rev, 6(11), pp.35~36 (1999)
- Abrams HB, "Initial-fit approach versus verified prescription: comparing self-perceived hearing aid benefit", J Am Acad Audiol, 23(10), pp.768~778 (2012)
- Kirkwood DH, "Survey: dispensers fitted more hearing aids in 2005 at higher prices", Hear J, 59(4), pp.40~ (2006)
- Mueller HG, "Survey examines popularity of real-ear probemicrophone measures", Hear J, 63(5), pp.27~28 (2010)
- Aarts NL, "Manufacturer predicted and measured REAR values in adult hearing aid fitting: accuracy and clinical usefulness", Int J Audiol, 44(5), pp.293~301 (2005)
- Nilsson M, "Development of the hearing in noise test for the measurement of speech reception thresholds in quiet and in noise", J Acoust Soc Am, 95(2), pp.1085~1099 (1994)
- 문성균, "한국어 Hearing in Noise Test(HINT) 문장의 개발", 대한이비인후-두경부외과학회지, 48(6), pp.724~728 (2005), 대한이비인후과학회
- Aazh H, "The accuracy of matching target insertion gains with open-fit hearing aids", Am J Audiol, 21(2), pp.175~180 (2012)
- Hawkins DB, "Hearing aid software predictive gain values: how accurate are they?", Hear J, 56(7), pp.26~28 (2003)