Two Evaluation Budgets for the Measurement Uncertainty of Glucose in Clinical Chemistry

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류병리학

초록

Background: Measurement uncertainty characterizes the dispersion of the quantity values attributed to a measurand. Although this concept was introduced to medical laboratories some years ago, not all medical researchers are familiar with it. Therefore, the evaluation and expression of measurement uncertainty must be highlighted using a practical example. Methods: In accordance with the procedure for evaluating and expressing uncertainty, provided by the Joint Committee for Guides in Metrology (JCGM), we used plasma glucose (Glu) as an example and defined it as the measurand. We then analyzed the main sources of uncertainty, evaluated each component of uncertainty, and calculated the combined uncertainty and expanded uncertainty with 2 budgets for single measurements and continuous monitoring, respectively. Results: During the measurement of Glu, the main sources of uncertainty included imprecision, within-subject biological variance (BVw), calibrator uncertainty, and systematic bias. We evaluated the uncertainty of each component to be 1.26%, 1.91%, 5.70%, 0.42%, and -2.87% for within-run imprecision, between-day imprecision, BVw, calibrator uncertainty, and systematic bias, respectively. For a single specimen, the expanded uncertainty was 7.38% or 6.1±0.45 mmol/L (κ=2); in continuous monitoring of Glu, the expanded uncertainty was 13.58% or 6.1±0.83 mmol/L (κ=2). Conclusions: We have demonstrated the overall procedure for evaluating and reporting uncertainty with 2 different budgets. The uncertainty is not only related to the medical laboratory in which the measurement is undertaken, but is also associated with the calibrator uncertainty and the biological variation of the subject. Therefore, it is helpful in explaining the accuracy of test results.

Abstract

Background: Measurement uncertainty characterizes the dispersion of the quantity values attributed to a measurand. Although this concept was introduced to medical laboratories some years ago, not all medical researchers are familiar with it. Therefore, the evaluation and expression of measurement uncertainty must be highlighted using a practical example. Methods: In accordance with the procedure for evaluating and expressing uncertainty, provided by the Joint Committee for Guides in Metrology (JCGM), we used plasma glucose (Glu) as an example and defined it as the measurand. We then analyzed the main sources of uncertainty, evaluated each component of uncertainty, and calculated the combined uncertainty and expanded uncertainty with 2 budgets for single measurements and continuous monitoring, respectively. Results: During the measurement of Glu, the main sources of uncertainty included imprecision, within-subject biological variance (BVw), calibrator uncertainty, and systematic bias. We evaluated the uncertainty of each component to be 1.26%, 1.91%, 5.70%, 0.42%, and -2.87% for within-run imprecision, between-day imprecision, BVw, calibrator uncertainty, and systematic bias, respectively. For a single specimen, the expanded uncertainty was 7.38% or 6.1±0.45 mmol/L (κ=2); in continuous monitoring of Glu, the expanded uncertainty was 13.58% or 6.1±0.83 mmol/L (κ=2). Conclusions: We have demonstrated the overall procedure for evaluating and reporting uncertainty with 2 different budgets. The uncertainty is not only related to the medical laboratory in which the measurement is undertaken, but is also associated with the calibrator uncertainty and the biological variation of the subject. Therefore, it is helpful in explaining the accuracy of test results.

참고문헌 (10)

  1. Working Group 2 of the Joint Committee for Guides in Metrology, "International vocabulary of metrology: basic and general concepts and associated terms (VIM). 3rd ed" (2008), JCGM
  2. Ehrlich C, "Evolution of philosophy and description of measurement (preliminary rationale for VIM3)", Accred Qual Assur, 12, pp.201~218 (2007)
  3. "Estimation of uncertainty of measurement in medical laboratories"[URL]
  4. Krouwer JS, "Uncertainty intervals based on deleting data are not useful", Clin Chem, 52, pp.1204~ (2006)
  5. Krouwer JS, "Critique of the guide to the expression of uncertainty in measurement method of estimating and reporting uncertainty in diagnostic assays", Clin Chem, 49, pp.1818~1821 (2003)
  6. Kristiansen J, "The guide to expression of uncertainty in measurement approach for estimating uncertainty: an appraisal", Clin Chem, 49, pp.1822~1829 (2003)
  7. "Desirable specifications for total error, imprecision, and bias, derived from intra- and inter-individual biologic variation"[URL]
  8. National Committee for Clinical Laboratory Standards, "User demonstration of performance for precision and accuracy: approved guideline" (2001), National Committee for Clinical Laboratory Standards
  9. Working Group 1 of the Joint Committee for Guides in Metrology, "Evaluation of the measurement data: guide to the express of uncertainty in measurement" (2008), JCGM
  10. White GH, "Uncertainty of measurement in quantitative medical testing: a laboratory implementation guide", Clin Biochem Rev, 25, pp.1~24 (2004)

관련 논문AI추천

이 간행물의 다른 논문 (16건)