Trueness Assessment for Serum Glucose Measurement Using Commercial Systems through the Preparation of Commutable Reference Materials
2
조회
0
다운로드
0
인용
0
참고문헌
키워드
초록
Background: Commutable reference materials (RMs) are suitable for end-users for evaluating the metrological traceability of values obtained using routine measurement systems. We assessed the performance of 6 routine measurement systems with validated secondary RMs. Methods: We tested the homogeneity, stability, and commutability of 5 minimally processed human serum pools according to the standard guidelines. The serum pools were assigned values as per the reference procedure of the United States Centers for Disease Control and were used to evaluate the trueness of results from 6 commercial measurement systems based on enzymatic methods: 3 glucose oxidase (GOD) and 3 hexokinase (HK) methods. Results: The prepared RMs were validated to be sufficiently homogenous, stable, and commutable with the patient samples. Method bias varied for different systems: GOD01, -0.17 to 2.88%; GOD02, 1.66 to 4.58%; GOD03, -0.17 to 3.14%; HK01, -3.48 to -0.85%; HK02, -3.83 to -0.11%, and HK03, -1.82 to -0.27%. Conclusions: We observed that the prepared serum glucose RMs were qualified for trueness assessment. Most of the measurement systems met the minimal quality specifications.
Abstract
Background: Commutable reference materials (RMs) are suitable for end-users for evaluating the metrological traceability of values obtained using routine measurement systems. We assessed the performance of 6 routine measurement systems with validated secondary RMs. Methods: We tested the homogeneity, stability, and commutability of 5 minimally processed human serum pools according to the standard guidelines. The serum pools were assigned values as per the reference procedure of the United States Centers for Disease Control and were used to evaluate the trueness of results from 6 commercial measurement systems based on enzymatic methods: 3 glucose oxidase (GOD) and 3 hexokinase (HK) methods. Results: The prepared RMs were validated to be sufficiently homogenous, stable, and commutable with the patient samples. Method bias varied for different systems: GOD01, -0.17 to 2.88%; GOD02, 1.66 to 4.58%; GOD03, -0.17 to 3.14%; HK01, -3.48 to -0.85%; HK02, -3.83 to -0.11%, and HK03, -1.82 to -0.27%. Conclusions: We observed that the prepared serum glucose RMs were qualified for trueness assessment. Most of the measurement systems met the minimal quality specifications.
참고문헌 (23)
- American Diabetes Association, "Standards of medical care in diabetes--2010", Diabetes Care, 33(S1), pp.S11~S61 (2010)
- Genter PM, "Accuracy of plasma glucose measurements in the hypoglycemic range", Diabetes Care, 17, pp.595~598 (1994)
- Thienpont LM, "Pilot external quality assessment survey for post-market vigilance of in vitro diagnostic medical devices and investigation of trueness of participants’ results", Clin Chem Lab Med, 41, pp.183~186 (2003)
- Gambino R, "Glucose: a simple molecule that is not simple to quantify", Clin Chem, 53, pp.2040~2041 (2007)
- Jansen R, "Trueness verification and traceability assessment of results from commercial systems for measurement of six enzyme activities in serum: an international study in the EC4 framework of the Calibration 2000 project", Clin Chim Acta, 368, pp.160~167 (2006)
- Xia C, "Application of five frozen human-pooled serum samples assigned by the International Federation of Clinical Chemistry and Laboratory Medicine reference procedure in a traceability investigation of gamma-glutamyltransferase catalytic concentration measurements in China", Ann Clin Biochem, 47, pp.189~194 (2010)
- Henriksen GM, "Minimally processed fresh frozen human reference sera: preparation, testing, and application to international external quality assurance", Scand J Clin Lab Invest, 64, pp.293~308 (2004)
- Linsinger T, "Homogeneity and stability of reference materials", Accred Qual Assur, 6, pp.20~25 (2001)
- Clinical and Laboratory Standards Institute (CLSI)., "Characterization and qualification of commutable reference materials for laboratory medicine, In Approved guideline, C53-A", Clinical and Laboratory Standards Institute, ~ (2010)
- Infusino I, "Commutability of the ERMDA470 reference material for two assays measuring serum albumin using immunochemical principles", Clin Chem Lab Med, 49, pp.1383~1384 (2011)
- glucose-6-phosphate dehydrogenase procedure for use as a national glucose reference method, "Development and evaluation of a hexokinase", pp.1976, Neese J
- "RELA-IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine"[URL]
- Snedecor G, "Statistical methods. 8th ed", pp.278~280 (1989), Iowa State University Press
- Pauwels J, "Evaluation of uncertatinty of reference materials", Accred Qual Assur, 5, pp.95~99 (2000)
- Passing H, "A new biometrical procedure for testing the equality of measurements from two different analytical methods. Application of linear regression procedures for method comparison studies in clinical chemistry, Part I", J Clin Chem Clin Biochem, 21, pp.709~720 (1983)
- Fraser C, "Biological Variation: From Principles to Practice" (2001), AACC Press
- Miller WG, "State of the art in trueness and interlaboratory harmonization for 10 analytes in general clinical chemistry", Arch Pathol Lab Med, 132, pp.838~846 (2008)
- Lee W, "Trueness assessment of Korean nation-wide glucose proficiency testing", Clin Chem Lab Med, 49, pp.1061~1064 (2011)
- Pedersen MM, "The Nordic Trueness Project 2002: use of reference measurement procedure values in a general clinical chemistry survey", Scand J Clin Lab Invest, 64, pp.309~320 (2004)
- Ricós C, "Current databases on biological variation: pros, cons and progress", Scand J Clin Lab Invest, 59, pp.491~500 (1999)
- Fraser CG, "Proposals for setting generally applicable quality goals solely based on biology", Ann Clin Biochem, 34, pp.8~12 (1997)
- Brion E, "Evaluation of commutability of control materials", Clin Chem Lab Med, 40, pp.625~630 (2002)
- Gubern G, "Determination of alpha-amylase activity: methods comparison and commutability study of several control materials", Clin Chem, 41, pp.435~438 (1995)