Assessment of the Trueness and Inter-Laboratory Precision of Routine Uric Acid Assays Using 4 Frozen Pooled Serum Samples Measured by the Japan Society of Clinical Chemistry’s HPLC Method

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류병리학

초록

Background: Reference procedures are required for evaluating the accuracy of routine analytical systems for uric acid (UA). External quality assessment (EQA) for UA has only been conducted with quality controls in China, and the results have not been published. This study was designed to investigate both the trueness and inter-laboratory precision of UA measurements among routine analytical systems using a candidate reference method. Methods: We performed the HPLC method recommended by the Japan Society of Clinical Chemistry (JSCC). Next, we evaluated its analytical performance and validated its true- ness. The performance of 4 routine analytical systems (5 instruments per system, n=20) for UA was assessed by using 4 frozen pooled serum samples measured by the HPLC method according to biologically relevant quality goals. Results: Within-run, between-run, inter-day, and total CV of the method were less than 0.3%, 0.4%, 1.8%, and 2.6%, respectively. The UA measurements were consistent with the target values of standard reference material (SRM) 909b, the sixth ring trial for Refer- ence Laboratories (RELA-2008) specimen, and national primary reference materials. The 4 frozen pooled serum samples were homogeneous, stable, and commutable. All routine systems achieved the desirable performance goal (total error <11.9%). Conclusions: We successfully reproduced the JSCC’s HPLC method, which was simple, specific, precise, and accurate. We recommend this method as a reference method for UA measurement in human serum. Four routine analytical systems for UA measurement had acceptable traceability, and their UA results showed good concordance.

Abstract

Background: Reference procedures are required for evaluating the accuracy of routine analytical systems for uric acid (UA). External quality assessment (EQA) for UA has only been conducted with quality controls in China, and the results have not been published. This study was designed to investigate both the trueness and inter-laboratory precision of UA measurements among routine analytical systems using a candidate reference method. Methods: We performed the HPLC method recommended by the Japan Society of Clinical Chemistry (JSCC). Next, we evaluated its analytical performance and validated its true- ness. The performance of 4 routine analytical systems (5 instruments per system, n=20) for UA was assessed by using 4 frozen pooled serum samples measured by the HPLC method according to biologically relevant quality goals. Results: Within-run, between-run, inter-day, and total CV of the method were less than 0.3%, 0.4%, 1.8%, and 2.6%, respectively. The UA measurements were consistent with the target values of standard reference material (SRM) 909b, the sixth ring trial for Refer- ence Laboratories (RELA-2008) specimen, and national primary reference materials. The 4 frozen pooled serum samples were homogeneous, stable, and commutable. All routine systems achieved the desirable performance goal (total error <11.9%). Conclusions: We successfully reproduced the JSCC’s HPLC method, which was simple, specific, precise, and accurate. We recommend this method as a reference method for UA measurement in human serum. Four routine analytical systems for UA measurement had acceptable traceability, and their UA results showed good concordance.

참고문헌 (28)

  1. Alasia DD, "Association of lead exposure, serum uric acid and parameters of renal function in Nigerian lead-exposed workers", Int J Occup Environ Med, 1, pp.182~190 (2010)
  2. Pant SS, "Hyperuricemia in Down’s syndrome", J Clin Endocrinol Metab, 28, pp.472~478 (1968)
  3. Giordano N, "Hyperuricemia and gout in thyroid endocrine disorders", Clin Exp Rheumatol, 19, pp.661~665 (2001)
  4. Miguel AL, "Uric acid and metabolic syndrome: what is the relationship?", Curr Rheumatol Rev, 7, pp.162~169 (2011)
  5. Mankovsky B, "Is serum uric acid a risk factor for atherosclerotic cardiovascular disease? A review of the clinical evidence. Part 1", Diabetes Metab Syndr, 4, pp.176~184 (2010)
  6. Parvin S, "Study of serum uric acid in cerebrovascular disease", J Dhaka Med Coll, 18, pp.15~19 (2009)
  7. Wilson DM, "Renal urate excretion in patients with Wilson’s disease", Kidney Int, 4, pp.331~336 (1973)
  8. Passeron A, "Hypo-uricemia in the syndrome of inappropriate secretion of antidiuretic hormone: a prospective study", Rev Med Interne, 31, pp.665~669 (2010)
  9. Duncan PH, "A candidate reference method for uric acid in serum. I. Optimization and evaluation", Clin Chem, 28, pp.284~290 (1982)
  10. Siekmann L, "Determination of uric acid in human serum by isotope dilution-mass spectrometry. Definitive methods in clinical chemistry, III", J Clin Chem Clin Biochem, 23, pp.129~135 (1985)
  11. Ellerbe P, "Determination of serum uric acid by isotope dilution mass spectrometry as a new candidate definitive method", Anal Chem, 62, pp.2173~2177 (1990)
  12. Stöckl D, "Candidate reference methods for determining target values for cholesterol, creatinine, uric acid, and glucose in external quality assessment and internal accuracy control. I. Method setup", Clin Chem, 39, pp.993~1000 (1993)
  13. Japan Society of Clinical Chemistry, "Recommended method for determination of uric acid in serum by HPLC", Jpn J Clin Chem, 22, pp.300~307 (1993)
  14. Dai X, "Determination of serum uric acid using high-performance liquid chromatography (HPLC)", isotope dilution mass spectrometry (ID-MS) as a candidate reference method, 2007, pp.857, 287~295
  15. Zhang C, "Determination of serum uric acid by isotope dilution liquid chromatography tandem mass spectrometry", Lab Med, 24, pp.878~882 (2009)
  16. Duncan P, "A candidate reference method for uric acid in serum. II. Interlaboratory testing", Clin Chem, 28, pp.291~293 (1982)
  17. Kock R, "A candidate reference method for the determination of uric acid in serum based on high performance liquid chromatography, compared with an isotope dilution-gas chromatography-mass spectrometer method", J Clin Chem Clin Biochem, 27, pp.157~162 (1989)
  18. Sakuma R, "Deproteinizing methods evaluated for determination of uric acid in serum by reversed-phase liquid chromatography with ultraviolet detection", Clin Chem, 33, pp.1427~1430 (1987)
  19. Yasushi C, "Comparative study on a high performance liquid chromatographic method and an enzymatic (uricase-UV) method for determination of serum uric acid", Bull Ginkyo Coll Med Sci, 17, pp.21~28 (1993)
  20. Clinical and Laboratory Standards Institute, "Evaluation of precision performance of quantitative measurement methods; approved guideline-second edition. EP5-A2", CLSI, ~ (2004)
  21. Clinical and Laboratory Standards Institute, "Method comparison and bias estimation using patient samples; approved guideline-second edition. EP9-A2", CLSI, ~ (2002)
  22. International Organization for Standardization, "ISO Guide 35: Reference materials-General and statistical principles for certification" (2006), ISO
  23. Clinical and Laboratory Standards Institute, "Evaluation of matrix effects; approved guideline-second edition. EP14-A2", CLSI, ~ (2005)
  24. Fraser CG, "Proposals for setting generally applicable quality goals solely based on biology", Ann Clin Biochem, 34, pp.8~12 (1997)
  25. Ricós C, "Current databases on biological variation: pros, cons and progress", Scand J Clin Lab Invest, 59, pp.491~500 (1999)
  26. "National Institute of Standards and Technology" (2011)[URL]
  27. "National information center for Certified Reference Materials" (2012)[URL]
  28. Miller WG, "State of the art in trueness and interlaboratory harmonization for 10 analytes in general clinical chemistry", Arch Pathol Lab Med, 132, pp.838~846 (2008)

관련 논문AI추천

Trueness Assessment for Serum Glucose Measurement Using Commercial Systems through the Preparation of Commutable Reference MaterialsChangYu Xia, GuoBing Xu, LanZhen Wang, Ou Liu대한진단검사의학회 · Annals of Laboratory Medicine2012 · 제32권 제4호Two Evaluation Budgets for the Measurement Uncertainty of Glucose in Clinical ChemistryHui Chen, Ling Zhang, Xiaoling Deng, Xiaoyun Bi대한진단검사의학회 · Annals of Laboratory Medicine2011 · 제31권 제3호Practical Considerations for Clinical Laboratories in Top-down Approach for Assessing the Measurement Uncertainty of Clinical Chemistry Analytes전사일, Gu Hyunjung, Hong Jinyoung, Lee Juhee, Lee Woochang, Lee Woo-In, Min Won-Ki, Yun Yeo-Min대한진단검사의학회 · Annals of Laboratory Medicine2022 · 제42권 제6호Can Reference Materials Prepared Following CLSI C37-A Be Utilized Without Commutability Assessment? Perspectives Based on Lipid MeasurementsAn Dongheui, Cho Chan-Ik, Choi Qute, Kim Jeong-Ho, Kim Sollip, Kwon Gye Cheol, Lee Sang-Guk, Park Pil-Whan, Seo Jong Do, Song Junghan, Yun Yeo Min대한진단검사의학회 · Annals of Laboratory Medicine2025 · 제45권 제6호Performance Evaluation of the JEOL BioMajesty JCA-BM6010/C Automated Clinical Chemistry Analyzer김형년, 문희원, 윤여민, 지미숙, 허미나대한진단검사의학회 · Laboratory Medicine Online2017 · 제7권 제3호Standardization Status of Total Cholesterol Concentration Measurement: Analysis of Korean External Quality Assessment DataChun Sail, Kim Sollip, Lee Kyunghoon, Lee Yong-Wha, Min Won-Ki, Park Hyung-Doo, Yoon Young Ahn대한진단검사의학회 · Annals of Laboratory Medicine2021 · 제41권 제4호

이 간행물의 다른 논문 (17건)