Background: We evaluated the reliability and accuracy of the combined use of matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS) bacterial identification and Vitek 2 antimicrobial susceptibility testing (AST) for bacteria from positive blood culture bottles. Methods: Direct identification and AST were performed in parallel to the standard methods in monomicrobial positive blood culture bottles. In total, 254 isolates grown on aerobic and/or anaerobic bottles were identified with MALDI-TOF Vitek MS (bioMérieux, France), and 1,978 microorganism/antimicrobial agent combinations were assessed. For isolates from anaerobic bottles, an aliquot of the culture broth was centrifuged, washed, and filtered through a nylon mesh. For isolates from aerobic/pediatric bottles, a lysis step using 9.26% ammonium chloride solution and 2% saponin solution was included. Results: The overall correct identification rate was 81.8% (208/254) and that for gram-positive/gram-negative isolates was 73.9%/92.6%, respectively, and it was 81.8%, 87.6%, and 57.9% for isolates from aerobic, anaerobic, and pediatric bottles, respectively. Identification was not possible in 45 cases, and most of these isolates were streptococci (N=14) and coagulase-negative staphylococci (N=11). Misidentification occurred only in one case. Compared with standard methods, direct AST showed 97.9% (1,936/1,978) agreement with very major error of 0.25%, major error of 0.05%, and minor error of 1.8%. Conclusions: This simple and cost-effective sample preparation method gives reliable results for the direct identification and AST of bacteria. For the identification of streptococci and coagulase-negative staphylococci, the method should be further improved.
Abstract
Background: We evaluated the reliability and accuracy of the combined use of matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS) bacterial identification and Vitek 2 antimicrobial susceptibility testing (AST) for bacteria from positive blood culture bottles. Methods: Direct identification and AST were performed in parallel to the standard methods in monomicrobial positive blood culture bottles. In total, 254 isolates grown on aerobic and/or anaerobic bottles were identified with MALDI-TOF Vitek MS (bioMérieux, France), and 1,978 microorganism/antimicrobial agent combinations were assessed. For isolates from anaerobic bottles, an aliquot of the culture broth was centrifuged, washed, and filtered through a nylon mesh. For isolates from aerobic/pediatric bottles, a lysis step using 9.26% ammonium chloride solution and 2% saponin solution was included. Results: The overall correct identification rate was 81.8% (208/254) and that for gram-positive/gram-negative isolates was 73.9%/92.6%, respectively, and it was 81.8%, 87.6%, and 57.9% for isolates from aerobic, anaerobic, and pediatric bottles, respectively. Identification was not possible in 45 cases, and most of these isolates were streptococci (N=14) and coagulase-negative staphylococci (N=11). Misidentification occurred only in one case. Compared with standard methods, direct AST showed 97.9% (1,936/1,978) agreement with very major error of 0.25%, major error of 0.05%, and minor error of 1.8%. Conclusions: This simple and cost-effective sample preparation method gives reliable results for the direct identification and AST of bacteria. For the identification of streptococci and coagulase-negative staphylococci, the method should be further improved.
참고문헌 (15)
- Prod’hom G, "Matrix-assisted laser desorption ionization-time of flight mass spectrometry for direct bacterial identification from positive blood culture pellets", J Clin Microbiol, 48, pp.1481~1483 (2010)
- Romero-Gómez MP, "Identification and susceptibility testing of microorganism by direct inoculation from positive blood culture bottles by combining MALDI-TOF and Vitek-2 Compact is rapid and effective", J Infect, 65, pp.513~520 (2012)
- Machen A, "Same day identification and full panel antimicrobial susceptibility testing of bacteria from positive blood culture bottles made possible by a combined lysis-filtration method with MALDI-TOF VITEK mass spectrometry and the VITEK2 system", PLoS One, 9, pp.e87870~ (2014)
- Schneiderhan W, "Work flow analysis of around-the-clock processing of blood culture samples and integrated MALDI-TOF mass spectrometry analysis for the diagnosis of bloodstream infections", Clin Chem, 59, pp.1649~1656 (2013)
- Stevenson LG, "Rapid identification of bacteria in positive blood culture broths by matrix-assisted laser desorption ionization-time of flight mass spectrometry", J Clin Microbiol, 48, pp.444~447 (2009)
- Kok J, "Identification of bacteria in blood culture broths using matrix-assisted laser desorption-ionization Sepsityper™ and time of flight mass spectrometry", PLoS One, 6, pp.e23285~ (2011)
- Ling TK, "Evaluation of the VITEK 2 system for rapid direct identification and susceptibility testing of gram-negative bacilli from positive blood cultures", J Clin Microbiol, 41, pp.4705~4707 (2003)
- Fothergill A, "Rapid identification of bacteria and yeasts from positive-blood-culture bottles by using a lysis-filtration method and matrix-assisted laser desorption ionization-time of flight mass spectrum analysis with the SARAMIS database", J Clin Microbiol, 51, pp.805~809 (2013)
- Lavergne RA, "An extraction method of positive blood cultures for direct identification of Candida species by Vitek MS matrix-assisted laser desorption ionization time of flight mass spectrometry", Med Mycol, 51, pp.652~656 (2013)
- Ferreira L, "Microorganisms direct identification from blood culture by matrix-assisted laser desorption", ionization time-of flight mass spectrometry, 2011, pp.17, 546~551
- Monteiro J, "Fast and reliable bacterial identification direct from positive blood culture using a new TFA sample preparation protocol and the Vitek®MS system", J Microibiol Methods, 109, pp.157~159 (2015)
- Marinach-Patrice C, "Rapid species diagnosis for invasive candidiasis using mass spectrometry", PLos One, 5, pp.e8862~ (2010)
- Clinical and Laboratory Standards Institute, "Performance standards for antimicrobial susceptibility testing", National Committee for Clinical Laboratory Standards, ~ (2014)
- 김영식, "Direct Identification of Urinary Tract Pathogens From Urine Samples Using the Vitek MS System Based on Matrix-Assisted Laser Desorption Ionization–Time of Flight Mass Spectrometry", Annals of Laboratory Medicine, 35(4), pp.416~422 (2015), 대한진단검사의학회
- Coltella L, "Advancement in the routine identification of anaerobic bacteria by MALDI-TOF mass spectrometry", Eur J Clin Microbiol Infect Dis, 32, pp.1183~1192 (2013)