Performance Evaluation of the Real-Q Cytomegalovirus (CMV) Quantification Kit Using Two Real-Time PCR Systems for Quantifying CMV DNA in Whole Blood

1

조회

0

다운로드

2

인용

0

참고문헌

키워드

주제분류병리학

초록

Standardized cytomegalovirus (CMV) DNA quantification is important for managing CMV disease. We evaluated the performance of the Real-Q CMV Quantification Kit (Real-Q assay; BioSewoom, Korea) using whole blood (WB), with nucleic acid extraction using MagNA Pure 96 (Roche Diagnostics, Germany). Real-time PCR was performed on two platforms: the 7500 Fast real-time PCR (7500 Fast; Applied Biosystems, USA) and CFX96 real-time PCR detection (CFX96; Bio-Rad, USA) systems. The WHO international standard, diluted with CMV-negative WB, was used to validate the analytical performance. We used 90 WB clinical samples for comparison with the artus CMV RG PCR kit (artus assay; Qiagen, Germany). Limits of detections (LODs) in 7500 Fast and CFX96 were 367 and 479 IU/mL, respectively. The assay was linear from the LOD to 106 IU/mL (R2 ≥0.9886). The conversion factors from copies to IU in 7500 Fast and CFX96 were 0.95 and 1.06, respectively. Compared with the artus assay, for values <1,000 copies/mL, 100% of the samples had a variation <0.7 log10 copies/mL; >1,000 copies/mL, 73.3% and 80.6% of samples in 7500 Fast and CFX96, respectively, had <0.5 log10 copies/mL. The Real-Q assay is useful for quantifying CMV in WB with the two real-time PCR platforms.

Abstract

Standardized cytomegalovirus (CMV) DNA quantification is important for managing CMV disease. We evaluated the performance of the Real-Q CMV Quantification Kit (Real-Q assay; BioSewoom, Korea) using whole blood (WB), with nucleic acid extraction using MagNA Pure 96 (Roche Diagnostics, Germany). Real-time PCR was performed on two platforms: the 7500 Fast real-time PCR (7500 Fast; Applied Biosystems, USA) and CFX96 real-time PCR detection (CFX96; Bio-Rad, USA) systems. The WHO international standard, diluted with CMV-negative WB, was used to validate the analytical performance. We used 90 WB clinical samples for comparison with the artus CMV RG PCR kit (artus assay; Qiagen, Germany). Limits of detections (LODs) in 7500 Fast and CFX96 were 367 and 479 IU/mL, respectively. The assay was linear from the LOD to 106 IU/mL (R2 ≥0.9886). The conversion factors from copies to IU in 7500 Fast and CFX96 were 0.95 and 1.06, respectively. Compared with the artus assay, for values <1,000 copies/mL, 100% of the samples had a variation <0.7 log10 copies/mL; >1,000 copies/mL, 73.3% and 80.6% of samples in 7500 Fast and CFX96, respectively, had <0.5 log10 copies/mL. The Real-Q assay is useful for quantifying CMV in WB with the two real-time PCR platforms.

참고문헌 (15)

  1. Le Page AK, "Clinical aspects of cytomegalovirus antiviral resistance in solid organ transplant recipients", Clin Infect Dis, 56, pp.1018~1029 (2013)
  2. Razonable RR, "Management strategies for cytomegalovirus infection and disease in solid organ transplant recipients", Infect Dis Clin North Am, 27, pp.317~342 (2013)
  3. Baldanti F, "Monitoring human cytomegalovirus infection in transplant recipients", J Clin Virol, 41, pp.237~241 (2008)
  4. Gerna G, "Virologic and immunologic monitoring of cytomegalovirus to guide preemptive therapy in solid-organ transplantation", Am J Transplant, 11, pp.2463~2471 (2011)
  5. Kraft CS, "Interpreting quantitative cytomegalovirus DNA testing: understanding the laboratory perspective", Clin Infect Dis, 54, pp.1793~1797 (2012)
  6. Pang XL, "Interlaboratory comparison of cytomegalovirus viral load assays", Am J Transplant, 9, pp.258~268 (2009)
  7. Collaborative study to evaluate the proposed 1st WHO international standard for human cytomegalovirus (HCMV) for nucleic acid amplification (NAT)-based assays, "Fryer JF", World Health Organization, ~ (2010)
  8. CLSI, "Evaluation of detection capability for clinical laboratory measurement procedures; approved guideline-Second ed. CLSI document EP17-A2", Clinical and Laboratory Standards Institute, ~ (2012)
  9. Miller S, "Comparison of QIAGEN automated nucleic acid extraction methods for CMV quantitative PCR testing", Am J Clin Pathol, 133, pp.558~563 (2010)
  10. Lisboa LF, "The clinical utility of whole blood versus plasma cytomegalovirus viral load assays for monitoring therapeutic response", Transplantation, 91, pp.231~236 (2011)
  11. Schnepf N, "Fully automated quantification of cytomegalovirus (CMV) in whole blood with the new sensitive Abbott RealTime CMV assay in the era of the CMV international standard", J Clin Microbiol, 51, pp.2096~2102 (2013)
  12. Deback C, "Use of the Roche LightCycler 480 system in a routine laboratory setting for molecular diagnosis of opportunistic viral infections: evaluation on whole blood specimens and proficiency panels", J Virol Methods, 159, pp.291~294 (2009)
  13. Bravo D, "Comparative evaluation of three automated systems for DNA extraction in conjunction with three commercially available real-time PCR assays for quantification of plasma Cytomegalovirus DNAemia in allogeneic stem cell transplant recipients", J Clin Microbiol, 49, pp.2899~2904 (2011)
  14. Edelmann A, "Performance of the MagNA Pure 96 system for cytomegalovirus nucleic acid amplification testing in clinical samples", J Clin Microbiol, 51, pp.1600~1601 (2013)
  15. Mengelle C, "A new highly automated extraction system for quantitative real-time PCRs from whole blood samples: routine monitoring of opportunistic infections in immunosuppressed patients", J Clin Virol, 53, pp.314~319 (2012)

관련 논문AI추천

Comparison of Quantitation of Cytomegalovirus DNA by Real-Time PCR in Whole Blood with the Cytomegalovirus Antigenemia Assay고선영, 권선희, 이창규, 정보경, 조윤정대한진단검사의학회 · Annals of Laboratory Medicine2015 · 제35권 제1호Performance of the Real-Q EBV Quantification Kit for Epstein-Barr Virus DNA Quantification in Whole Blood기창석, 김종원, 김지연, 박종은, 윤선애, 이남용, 이명근, 허희재대한진단검사의학회 · Annals of Laboratory Medicine2017 · 제37권 제2호거대세포바이러스 정량 검출을 위한 BiosewoomTM Real-Q CytomegalovirusQuantification kit 평가김명희, 김유리, 서장수, 오흥범, 원동일, 허운보대한진단검사의학회 · Annals of Laboratory Medicine2007 · 제27권 제4호Real-Time PCR과 일반 PCR을 이용한 Cytomegalovirus 정성검사의 비교 평가김문정, 김윤정, 배일권, 정석훈, 정세리대한임상미생물학회 · Annals of Clinical Microbiology2013 · 제16권 제1호Automated Nucleic Acid Extraction Systems for Detecting Cytomegalovirus and Epstein-Barr Virus Using Real-Time PCR: A Comparison Study Between the QIAsymphony RGQ and QIAcube Systems김지영, 김한아, 문희원, 윤여민, 조현찬, 허미나대한진단검사의학회 · Annals of Laboratory Medicine2017 · 제37권 제2호인간거대세포바이러스 DNA 정량 중합효소연쇄반응 검사의 실제김미나, 성흥섭, 장정현, 황상현대한임상미생물학회 · Annals of Clinical Microbiology2017 · 제20권 제2호

이 간행물의 다른 논문 (19건)