PURPOSE. Autoclaves and UV sterilizers have been commonly used to prevent cross-infections between dental patients and dental instruments or materials contaminated by saliva and blood. To develop a dental sterilizer which can sterilize most materials, such as metals, rubbers, and plastics, the sterilization effect of an atmospheric pressure non-thermal air plasma device was evaluated. MATERIALS AND METHODS. After inoculating E. coli and B. subtilis the diamond burs and polyvinyl siloxane materials were sterilized by exposing them to the plasma for different lengths of time (30, 60, 90, 120, 180 and, 240 seconds). The diamond burs and polyvinyl siloxane materials were immersed in PBS solutions, cultured on agar plates and quantified by counting the colony forming units. The data were analyzed using one-way ANOVA and significance was assessed by the LSD post hoc test (α=0.05). RESULTS. The device was effective in killing E. coli contained in the plasma device compared with the UV sterilizer. The atmospheric pressure non-thermal air plasma device contributed greatly to the sterilization of diamond burs and polyvinyl siloxane materials inoculated with E. coli and B. subtilis. Diamond burs and polyvinyl siloxane materials inoculated with E. coli was effective after 60 and 90 seconds. The diamond burs and polyvinyl siloxane materials inoculated with B. subtilis was effective after 120 and 180 seconds. CONCLUSION. The atmospheric pressure non-thermal air plasma device was effective in killing both E. coli and B. subtilis, and was more effective in killing E. coli than the UV sterilizer.

Abstract

PURPOSE. Autoclaves and UV sterilizers have been commonly used to prevent cross-infections between dental patients and dental instruments or materials contaminated by saliva and blood. To develop a dental sterilizer which can sterilize most materials, such as metals, rubbers, and plastics, the sterilization effect of an atmospheric pressure non-thermal air plasma device was evaluated. MATERIALS AND METHODS. After inoculating E. coli and B. subtilis the diamond burs and polyvinyl siloxane materials were sterilized by exposing them to the plasma for different lengths of time (30, 60, 90, 120, 180 and, 240 seconds). The diamond burs and polyvinyl siloxane materials were immersed in PBS solutions, cultured on agar plates and quantified by counting the colony forming units. The data were analyzed using one-way ANOVA and significance was assessed by the LSD post hoc test (α=0.05). RESULTS. The device was effective in killing E. coli contained in the plasma device compared with the UV sterilizer. The atmospheric pressure non-thermal air plasma device contributed greatly to the sterilization of diamond burs and polyvinyl siloxane materials inoculated with E. coli and B. subtilis. Diamond burs and polyvinyl siloxane materials inoculated with E. coli was effective after 60 and 90 seconds. The diamond burs and polyvinyl siloxane materials inoculated with B. subtilis was effective after 120 and 180 seconds. CONCLUSION. The atmospheric pressure non-thermal air plasma device was effective in killing both E. coli and B. subtilis, and was more effective in killing E. coli than the UV sterilizer.

참고문헌 (24)

  1. 최하나, "치과 감염 관리에 관한 국내 문헌고찰(1988~2009년)", 치위생과학회지, 10(4), pp.199~209 (2010), 한국치위생과학회
  2. Hutchings ML, "Immersion disinfection of irreversible hydrocolloid impressions in pH-adjusted sodium hypochlorite. Part 2: Effect on gypsum casts", Int J Prosthodont, 9, pp.223~229 (1996)
  3. Leung RL, "Gypsum casts as a potential source of microbial cross-contamination", J Prosthet Dent, 49, pp.210~211 (1983)
  4. McNeill MR, "Disinfection of irreversible hydrocolloid impressions: a comparative study", Int J Prosthodont, 5, pp.563~567 (1992)
  5. Runnells RR, "An overview of infection control in dental practice", J Prosthet Dent, 59, pp.625~629 (1988)
  6. Jennings KJ, "The persistence of microorganisms on impression materials following disinfection", Int J Prosthodont, 4, pp.382~387 (1991)
  7. Morfill GE, "Nosocomial infections-a new approach towards preventive medicine using plasmas", New J phys, 11, pp.115019~ (2009)
  8. Samaranayake L, "Essential Microbiology for dentistry.", pp.385~412 (2007), Churchill Livingstone
  9. "Guideline of sterilization of appliance and article at medical institution", Ministry of Health and Welfare Notification No. 2010-61, 2010. 8. 13 enactment, ~ (2010)
  10. Moreau M, "Non-thermal plasma technologies: new tools for bio-decontamination", Biotechnol Adv, 26, pp.610~617 (2008)
  11. Kyenam Lee, "Sterilization of Bacteria, Yeast, and Bacterial Endospores by Atmospheric-Pressure Cold Plasma using Helium and Oxygen", The Journal of Microbiology, 44(3), pp.269~275 (2006), 한국미생물학회
  12. Dobrynin D, "Physical and biological mechanisms of direct plasma interaction with living tissue", New J Phys, 11, pp.115020~ (2009)
  13. Fridman G, "Applied plasma Medicine", Plasma Process Polym, 5, pp.503~533 (2008)
  14. McCombs GB, "New discoveries and directions for medical, dental and dental hygiene research: low temperature atmospheric pressure plasma", Int J Dent Hyg, 8, pp.10~15 (2010)
  15. Sladek REJ, "Deactivation of Escherichia coli by the plasma needle", J Phys D Appl Phys, 38, pp.1716~1721 (2005)
  16. Lee HW, "Tooth bleaching with nonthermal atmospheric pressure plasma", J Endod, 35, pp.587~591 (2009)
  17. Samaranayake LP, "Carriage of oral flora on irreversible hydrocolloid and elastomeric impression materials", J Prosthet Dent, 65, pp.244~249 (1991)
  18. Smith A, "Management of infection control in dental practice", J Hosp Infect, 71, pp.353~358 (2009)
  19. Seo YS, "Comparative studies of atmospheric pressure plasma characteristics between He and Ar working gases for sterilization", IEEE Trans Plasma Sci IEEE Nucl Plasm, 38, pp.1954~1962 (2010)
  20. Iza F, "Microplasmas: Sources, particle kinetics, and biomedical applications", Plasma Process Polym, 5, pp.322~344 (2008)
  21. Laroussi M, "Low Temperature Plasma-Based Sterilization: Overview and state-of-the Art", Plasma Process Polym, 2, pp.391~400 (2005)
  22. Fridman G, "Comparison of direct and indirect effects of non-thermal atmosphericpressure plasma on bacteria", Plasma Process Polym, 4, pp.370~375 (2007)
  23. Laroussi M, "Nonthermal decontamination of biological media by atmospheric-pressure plasma: review, analysis, and prospects", IEEE Trans Plasma Sci, 30, pp.1409~1415 (2002)
  24. Suzuki T, "Influence of ozone on oxidation of dental alloys", Int J Prosthodont, 12, pp.179~183 (1999)