본문 바로가기
KERD
논문KCI 등재

Microbiological Investigation of Reusing Disposable Saliva Ejectors

1

조회

0

다운로드

0

인용

0

참고문헌

키워드

초록

Objective: Dental instruments are often designed for single use; however, some dental clinics resort to reusing instruments, including saliva ejectors. In this study, we aimed to investigate and compare various disinfection and sterilization methods for contaminated disposable salivary ejectors. Methods: Saliva ejectors contaminated with Streptococcus mutans were divided into f ive groups: negative control, positive control, cleaning, disinfection, and sterilization. Surface contamination and microbial culture tests were conducted using the adenosine triphosphate (ATP) test and colony-forming units (CFUs) counting, respectively. One-way ANOVA was used to assess the differences among groups. Results: There were no significant differences among the cleaning, disinfection, and sterilization groups based on immediate ATP test. However, CFU counts were significantly higher in the cleaning and disinfection groups compared to negative control group, while no microbial growth was observed in the sterilization group. Conclusions: This study demonstrated that saliva ejectors harbor bacteria capable of growth, even after cleaning with a toothbrush or disinfectant. Autoclaving proved to be the most effective sterilization method, eliminating microbial growth, and allowing for microbiological reuse. Future studies involving diverse microorganisms and different types of contaminants are recommended.

Abstract

Objective: Dental instruments are often designed for single use; however, some dental clinics resort to reusing instruments, including saliva ejectors. In this study, we aimed to investigate and compare various disinfection and sterilization methods for contaminated disposable salivary ejectors. Methods: Saliva ejectors contaminated with Streptococcus mutans were divided into f ive groups: negative control, positive control, cleaning, disinfection, and sterilization. Surface contamination and microbial culture tests were conducted using the adenosine triphosphate (ATP) test and colony-forming units (CFUs) counting, respectively. One-way ANOVA was used to assess the differences among groups. Results: There were no significant differences among the cleaning, disinfection, and sterilization groups based on immediate ATP test. However, CFU counts were significantly higher in the cleaning and disinfection groups compared to negative control group, while no microbial growth was observed in the sterilization group. Conclusions: This study demonstrated that saliva ejectors harbor bacteria capable of growth, even after cleaning with a toothbrush or disinfectant. Autoclaving proved to be the most effective sterilization method, eliminating microbial growth, and allowing for microbiological reuse. Future studies involving diverse microorganisms and different types of contaminants are recommended.

참고문헌 (20)

  1. Park BY, "Survey of the real-world re-use of disposable dental supplies in dental offices", J Korean Soc Dent Hyg, 19(1), pp.141~149 (2019)[DOI]
  2. Matys J, "Dental aerosol as a hazard risk for dental workers", Materials, 13(22), pp.1~13 (2020)[DOI]
  3. Rexhepi I, "Effects of natural ventilation and saliva standard ejectors during the COVID-19 pandemic : a quantitative analysis of aerosol produced during dental procedures", Int J Environ Res Public Health, 18(14), pp.1~9 (2021)[DOI]
  4. Harrel SK, "Aerosols and splatter in dentistry : a brief review of the literature and infection control implications", J Am Dent Assoc, 135(4), pp.429~437 (2004)[DOI]
  5. Engsomboon N, "Comparative dissemination of aerosol and splatter using suction device during ultrasonic scaling : a pilot study", Dent J, 10(8), pp.1~9 (2022)[DOI]
  6. Lin LW, "Saliva ejector assisted laryngoscopy(SEAL)for protective intubation", Am J Emerg Med, 43, pp.270~271 (2021)[DOI]
  7. Wu KY, "COVID-19's impact on private practice and academic dentistry in North America", Oral Dis, 27(S3), pp.684~687 (2020)[DOI]
  8. Peng X, "Transmission routes of 2019-nCoV and controls in dental practice", nt J Oral Sci, 12(1), pp.1~6 (2020)[DOI]
  9. Suwandi T, "The efficacy of high-volume evacuators and extraoral vacuum aspirators in reducing aerosol and droplet in ultrasonic scaling procedures during the COVID-19 pandemic", Eur J Dent, 16(4), pp.803~808 (2022)[DOI]
  10. McGain F, "Financial and environmental costs of reusable and single-use anesthetic equipment", Br J Anaesth, 118(6), pp.862~869 (2017)[DOI]
  11. Ghaffari T, "Comparison of the disinfectant effects of Nanosil D2 and Korsolex extra solutions on thermoset acrylic resin contaminated with Streptococcus mutans and Bacillus subtilis", J Adv Pharm Technol Res, 10(2), pp.90~94 (2019)[DOI]
  12. Lee YS, "Comparison of autofluorescence and pH changes in Acinetobacter baumannii and Candida albicans cultures based on incubation time", Korean J Clin Dent Hyg, 9(1), pp.11~19 (2021)[DOI]
  13. Florez FLE, "Real-time assessment of Streptococcus mutans biofilm metabolism on resin composite", Dent Mater, 32(10), pp.1263~1269 (2016)[DOI]
  14. Knape L, "The adenosine triphosphate method as a quality control tool to assess 'cleanliness' of frequently touched hospital surfaces", J Hosp Infect, 91(2), pp.166~170 (2015)[DOI]
  15. van Arkel A, "ATP measurement as an objective method to measure environmental contamination in 9 hospitals in the Dutch", Belgian border area, 2020, pp.9(77), 1~8
  16. Ferreira AM, "Assessment of disinfection of hospital surfaces using different monitoring methods", Rev Lat Am Enfermagem, 23(3), pp.466~474 (2015)[DOI]
  17. Nunes LKF, "In vitro comparison of peracetic acid and autoclave sterilization in the corrosion of orthodontic pliers : a pilot study", Dental Press J Orthod, 26(5), pp.1~26 (2021)[DOI]
  18. Sajjanshetty S, "Decontamination methods used for dental burs–a comparative study", J Clin Diagn Res, 8(6), pp.39~41 (2014)[DOI]
  19. Sogin JH, "Implementation of ATP and microbial indicator testing for hygiene monitoring in a Tofu production facility improves product quality and hygienic conditions of food contact surfaces : A case study", Appl Environ Microbiol, 87(5), pp.e02278-20~ (2021)[DOI]
  20. Omidbakhsh N, "How reliable are ATP bioluminescence meters in assessing decontamination of environmental surfaces in healthcare settings?", PLoS One, 9(6), pp.e99951~ (2014)[DOI]