Comparative study on stress distribution around internal tapered connection implants according to fit of cement-and screw-retained prostheses

2

조회

0

다운로드

2

인용

0

참고문헌

키워드

초록

PURPOSE. The aim of this study was to compare the passivity of implant superstructures by assessing the strain development around the internal tapered connection implants with strain gauges. MATERIALS AND METHODS. A polyurethane resin block in which two implants were embedded served as a measurement model. Two groups of implant restorations utilized cement-retained design and internal surface of the first group was adjusted until premature contact between the restoration and the abutment completely disappeared. In the second group, only nodules detectable to the naked eye were removed. The third group employed screw-retained design and specimens were generated by computer-aided design/computer-aided manufacturing system (n=10). Four strain gauges were fixed on the measurement model mesially and distally to the implants. The strains developed in each strain gauge were recorded during fixation of specimens. To compare the difference among groups, repeated measures 2-factor analysis was performed at a level of significance of α=.05. RESULTS. The absolute strain values were measured to analyze the magnitude of strain. The mean absolute strain value ranged from 29.53 to 412.94 μm/m at the different strain gauge locations. According to the result of overall comparison, the cement-retained prosthesis groups exhibited significant difference. No significant difference was detected between milled screw-retained prostheses group and cement-retained prosthesis groups. CONCLUSION. Within the limitations of the study, it was concluded that the cement-retained designs do not always exhibit lower levels of stress than screw-retained designs. The internal adjustment of a cement-retained implant restoration is essential to achieve passive fit.

Abstract

PURPOSE. The aim of this study was to compare the passivity of implant superstructures by assessing the strain development around the internal tapered connection implants with strain gauges. MATERIALS AND METHODS. A polyurethane resin block in which two implants were embedded served as a measurement model. Two groups of implant restorations utilized cement-retained design and internal surface of the first group was adjusted until premature contact between the restoration and the abutment completely disappeared. In the second group, only nodules detectable to the naked eye were removed. The third group employed screw-retained design and specimens were generated by computer-aided design/computer-aided manufacturing system (n=10). Four strain gauges were fixed on the measurement model mesially and distally to the implants. The strains developed in each strain gauge were recorded during fixation of specimens. To compare the difference among groups, repeated measures 2-factor analysis was performed at a level of significance of α=.05. RESULTS. The absolute strain values were measured to analyze the magnitude of strain. The mean absolute strain value ranged from 29.53 to 412.94 μm/m at the different strain gauge locations. According to the result of overall comparison, the cement-retained prosthesis groups exhibited significant difference. No significant difference was detected between milled screw-retained prostheses group and cement-retained prosthesis groups. CONCLUSION. Within the limitations of the study, it was concluded that the cement-retained designs do not always exhibit lower levels of stress than screw-retained designs. The internal adjustment of a cement-retained implant restoration is essential to achieve passive fit.

참고문헌 (44)

  1. Rudd KD, "Horizontal tooth mobility in carefully screened subjects", Periodontology, 4, pp.65~70 (1964)
  2. Sekine H, "Mobility characteristics and tactile sensitivity of osseointegrated fixture-supporting systems. In Tissue integration in oral and maxillofacial reconstruction", pp.326~329 (1986), Excerpta Medica
  3. Adell R, "A 15-year study of osseointegrated implants in the treatment of the edentulous jaw", Int J Oral Surg, 10, pp.387~416 (1981)
  4. Lundqvist S, "Maxillary fixed prostheses on osseointegrated dental implants", J Prosthet Dent, 50, pp.262~270 (1983)
  5. Lekholm U, "Complications. in Tissue-Integrated Prostheses: Osseointegration in Clinical Dentistry", pp.233~240 (1985), Quintessence
  6. Rangert B, "Forces and moments on Branemark implants", Int J Oral Maxillofac Implants, 4, pp.241~247 (1989)
  7. Spector MR, "An evaluation of impression techniques for osseointegrated implants", J Prosthet Dent, 63, pp.444~447 (1990)
  8. Michalakis KX, "Cement-retained versus screw-retained implant restorations: a critical review", Int J Oral Maxillofac Implants, 18, pp.719~728 (2003)
  9. Taylor TD, "Prosthodontic problems and limitations associated with osseointegration", J Prosthet Dent, 79, pp.74~78 (1998)
  10. Goodacre CJ, "Clinical complications of osseointegrated implants", J Prosthet Dent, 81, pp.537~552 (1999)
  11. Kallus T, "Loose gold screws frequently occur in full-arch fixed prostheses supported by osseointegrated implants after 5 years", Int J Oral Maxillofac Implants, 9, pp.169~178 (1994)
  12. Jemt T, "Failures and complications in 391 consecutively inserted fixed prostheses supported by Brånemark implants in edentulous jaws: a study of treatment from the time of prosthesis placement to the first annual checkup", Int J Oral Maxillofac Implants, 6, pp.270~276 (1991)
  13. Naert I, "A study of 589 consecutive implants supporting complete fixed prostheses. Part II: Prosthetic aspects", J Prosthet Dent, 68, pp.949~956 (1992)
  14. Adell R, "Long-term follow-up study of osseointegrated implants in the treatment of totally edentulous jaws", Int J Oral Maxillofac Implants, 5, pp.347~359 (1990)
  15. Jemt T, "Prosthesis misfit and marginal bone loss in edentulous implant patients", Int J Oral Maxillofac Implants, 11, pp.620~625 (1996)
  16. Rangert B, "Bending overload and implant fracture: a retrospective clinical analysis", Int J Oral Maxillofac Implants, 10, pp.326~334 (1995)
  17. Tan KB, "The clinical significance of distortion in implant prosthodontics: is there such a thing as passive fit", Ann Acad Med Singapore, 24, pp.138~157 (1995)
  18. Heckmann SM, "Cement fixation and screw retention: parameters of passive fit. An in vitro study of three-unit implant-supported fixed partial dentures", Clin Oral Implants Res, 15, pp.466~473 (2004)
  19. Karl M, "In vitro study on passive fit in implant-supported 5-unit fixed partial dentures", Int J Oral Maxillofac Implants, 19, pp.30~37 (2004)
  20. Taylor TD, "Implant prosthodontics: current perspective and future directions", Int J Oral Maxillofac Implants, 15, pp.66~75 (2000)
  21. Chiche GJ, "Considerations for fabrication of implant-supported posterior restorations", Int J Prosthodont, 4, pp.37~44 (1991)
  22. Pietrabissa R, "An in vitro study on compensation of mismatch of screw versus cement-retained implant supported fixed prostheses", Clin Oral Implants Res, 11, pp.448~457 (2000)
  23. Guichet DL, "Passivity of fit and marginal opening in screw- or cement-retained implant fixed partial denture designs", Int J Oral Maxillofac Implants, 15, pp.239~246 (2000)
  24. Guichet DL, "Load transfer in screw- and cement-retained implant fixed partial denture designs", J Prosthet Dent, 72, pp.631~ (1994)
  25. Misch CE, "Screw-retained versus cement-retained implantsupported prostheses", Pract Periodontics Aesthet Dent, 7, pp.15~18 (1995)
  26. Kim WD, "In vitro stress analyses of dental implants supporting screw-retained and cement-retained prostheses", Implant Dent, 8, pp.141~151 (1999)
  27. Williams WN, "Human discrimination of different bite forces", J Oral Rehabil, 11, pp.407~413 (1984)
  28. Kikuchi M, "The association among occlusal contacts, clenching effort, and bite force distribution in man", J Dent Res, 76, pp.1316~1325 (1997)
  29. Watanabe F, "Analysis of stress distribution in a screw-retained implant prosthesis", Int J Oral Maxillofac Implants, 15, pp.209~218 (2000)
  30. Iplikçioğlu H, "Comparison of non-linear finite element stress analysis with in vitro strain gauge measurements on a Morse taper implant", Int J Oral Maxillofac Implants, 18, pp.258~265 (2003)
  31. Karl M, "Static implant loading caused by as-cast metal and ceramic-ve-neered superstructures", J Prosthet Dent, 93, pp.324~330 (2005)
  32. Karl M, "Influence of fixation mode and superstructure span upon strain development of implant fixed partial dentures", J Prosthodont, 17, pp.3~8 (2008)
  33. Karl M, "Strain development in 3-unit implant-supported CAD", CAM restorations, 2008, pp.23, 648~652
  34. Stafford GD, "Intraoral strain gauge measurements on complete dentures: a methodological study", J Dent, 19, pp.80~84 (1991)
  35. Bernal G, "The effects of abutment taper, length and cement type on resistance to dislodgement of cement-retained, implant-supported restorations", J Prosthodont, 12, pp.111~115 (2003)
  36. Clelland NL, "Comparison of strains transferred to a bone simulant among implant overdenture bars with various levels of misfit", J Prosthodont, 4, pp.243~250 (1995)
  37. Clelland NL, "Comparison of strains transferred to a bone simulant between as-cast and postsoldered implant frameworks for a five-implant-supported fixed prosthesis", J Prosthodont, 5, pp.193~200 (1996)
  38. Clelland NL, "Comparison of strains produced in a bone simulant between conventional cast and resin-luted implant frameworks", Int J Oral Maxillofac Implants, 12, pp.793~799 (1997)
  39. Pietrabissa R, "Experimental and computational approach for the evaluation of the biomechanical effects of dental bridge misfit", J Biomech, 33, pp.1489~1495 (2000)
  40. Nishioka RS, "Machined and plastic copings in three-element prostheses with different types of implant-abutment joints: a strain gauge comparative analysis", J Appl Oral Sci, 18, pp.225~230 (2010)
  41. Schwarz MS, "Mechanical complications of dental implants", Clin Oral Implants Res, 11, pp.156~158 (2000)
  42. Jemt T, "Prosthesis misfit and marginal bone loss in edentulous implant patients", Int J Oral Maxillofac Implants, 11, pp.620~625 (1996)
  43. Carr AB, "The response of bone in primates around unloaded dental implants supporting prostheses with different levels of fit", J Prosthet Dent, 76, pp.500~509 (1996)
  44. Eser A, "Nonlinear finite element analysis versus ex vivo strain gauge measurements on immediately loaded implants", Int J Oral Maxillofac Implants, 24, pp.439~446 (2009)

관련 논문AI추천

Impact of type and position of abutment connection on microstrain distribution: an in vitro studyBasel Mahardawi, Chamaiporn Sukjamsri, Jekita Siripru, Napapa Aimjirakul, Usanee Puengpaiboon대한치과보철학회 · The Journal of Advanced of Prosthodontics2024 · 제16권 제5호Effects of cementless fixation of implant prosthesis: A finite element study권긍록, 노건우, 박소연, 이현종대한치과보철학회 · The Journal of Advanced of Prosthodontics2019 · 제11권 제6호Effect of cement washout on loosening of abutment screws and vice versa in screw- and cement- retained implant-supported dental prosthesis김석규, 손미경, 정재헌대한치과보철학회 · The Journal of Advanced of Prosthodontics2015 · 제7권 제3호3본 나사 유지형 임플란트 보철물의 고정 방식에 따른 임플란트 고정체 치경부에 발생하는 변형율 비교분석김상범, 이두형, 이청희대한치과보철학회 · 대한치과보철학회지2020 · 제58권 제4호Influence of misfit direction and abutment type on torque loss and fracture strength of implant-supported fixed dental prosthesesDaniel Kules, Ieva Gendviliene, Ingrida Mischitz, Liudas Auškalnis, Maisa Haiek, Mutlu Özcan, Tan Fırat Eyüboğlu, Vygandas Rutkūnas대한치과보철학회 · The Journal of Advanced of Prosthodontics2025 · 제17권 제6호Effect of cyclic loading on axial displacement of abutment into implant with internal tapered connection: a pilot study곽재영, 김성균, 설현우, 한종현, 허성주대한치과보철학회 · 대한치과보철학회지2013 · 제51권 제4호

이 간행물의 다른 논문 (19건)