Influence of abutment height and convergence angle on the retrievability of cement-retained implant prostheses with a lingual slot
2
조회
0
다운로드
0
인용
0
참고문헌
키워드
초록
PURPOSE. Cement-retained implant prostheses can lack proper retrievability during repair, and residual cement can cause peri-implantitis. The purpose of this in vitro study was to evaluate the influence of abutment height and convergence angle on the retrievability of cement-retained implant prostheses with lingual slots, known as retrievable cement-type slots (RCS). MATERIALS AND METHODS. We fabricated six types of titanium abutments (10 of each type) with two different heights (4 mm and 6 mm), three different convergence angles (8°, 10°, and 12°), a sloped shoulder margin (0.6 mm depth), a rectangular shape (6 mm × 6.5 mm) with rounded edges, and a rectangular ledge (2 mm × 1 mm) for the RCS. One monolithic zirconia crown was fabricated for each abutment using a dental computer-aided design/computer-aided manufacturing system. The abutments and crowns were permanently cemented together with dual-curing resin cement, followed by 24 hours in demineralized water at room temperature. Using a custom-made device with a slot driver and torque gauge, we recorded the torque (N·cm) required to remove the crowns. Statistical analysis was conducted using multiple regression analysis and Mann-Whitney U tests (α=.05). RESULTS. Removal torques significantly decreased as convergence angles increased. Multiple regression analysis showed no significant interaction between the abutment height and the convergence angle (Durbin-Watson ratio: 2.186). CONCLUSION. Within the limitations of this in vitro study, we suggest that the retrievability of cement-retained implant prostheses with RCS can be maintained by adjusting the abutment height and convergence angle, even when they are permanently cemented together.
Abstract
PURPOSE. Cement-retained implant prostheses can lack proper retrievability during repair, and residual cement can cause peri-implantitis. The purpose of this in vitro study was to evaluate the influence of abutment height and convergence angle on the retrievability of cement-retained implant prostheses with lingual slots, known as retrievable cement-type slots (RCS). MATERIALS AND METHODS. We fabricated six types of titanium abutments (10 of each type) with two different heights (4 mm and 6 mm), three different convergence angles (8°, 10°, and 12°), a sloped shoulder margin (0.6 mm depth), a rectangular shape (6 mm × 6.5 mm) with rounded edges, and a rectangular ledge (2 mm × 1 mm) for the RCS. One monolithic zirconia crown was fabricated for each abutment using a dental computer-aided design/computer-aided manufacturing system. The abutments and crowns were permanently cemented together with dual-curing resin cement, followed by 24 hours in demineralized water at room temperature. Using a custom-made device with a slot driver and torque gauge, we recorded the torque (N·cm) required to remove the crowns. Statistical analysis was conducted using multiple regression analysis and Mann-Whitney U tests (α=.05). RESULTS. Removal torques significantly decreased as convergence angles increased. Multiple regression analysis showed no significant interaction between the abutment height and the convergence angle (Durbin-Watson ratio: 2.186). CONCLUSION. Within the limitations of this in vitro study, we suggest that the retrievability of cement-retained implant prostheses with RCS can be maintained by adjusting the abutment height and convergence angle, even when they are permanently cemented together.
참고문헌 (32)
- Misch CE, "Contemporary implant dentistry", pp.549~593 (1999), Mosby Inc
- Misch CE, "Screw-retained versus cement-retained implant supported prostheses", Pract Periodontics Aesthet Dent, 7, pp.15~18 (1995)
- Hebel KS, "Cement-retained versus screw-retained implant restorations: achieving optimal occlusion and esthetics in implant dentistry", J Prosthet Dent, 77, pp.28~35 (1997)[DOI]
- Chee W, "Cemented versus screw-retained implant prostheses: which is better?", Int J Oral Maxillofac Implants, 14, pp.137~141 (1999)
- Ekfeldt A, "Clinical evaluation of single-tooth restorations supported by osseointegrated implants: a retrospective study", Int J Oral Maxillofac Implants, 9, pp.179~183 (1994)
- Breeding LC, "Use of luting agents with an implant system: Part I", J Prosthet Dent, 68, pp.737~741 (1992)[DOI]
- Valbao FP Jr, "Alternative method for retention and removal of cement-retained implant prostheses", J Prosthet Dent, 86, pp.181~183 (2001)[DOI]
- Clausen GF, "The lingual locking screw for implant-retained restorations--aesthetics and retrievability", Aust Prosthodont J, 9, pp.17~20 (1995)
- Doerr J, "Simplified technique for retrieving cemented implant restorations", J Prosthet Dent, 88, pp.352~353 (2002)[DOI]
- Daher T, "The use of digital photographs to locate implant abutment screws for implant-supported cement-retained restorations", J Prosthet Dent, 100, pp.238~239 (2008)[DOI]
- Wadhwani C, "Simple device for locating the abut- ment screw position of a cement-retained implant restoration", J Prosthet Dent, 109, pp.272~274 (2013)[DOI]
- Prestipino V, "A practical approach for retrieving cement-retained, implant-supported restorations", Quintessence Dent Technol, 24, pp.182~187 (2001)
- Schweitzer DM, "A technique for retrieval of cement-retained implant-supported prostheses", J Prosthet Dent, 106, pp.134~138 (2011)[DOI]
- Shareen H Elshiyab, "Fracture resistance of implantsupported monolithic crowns cemented to zirconia hybridabutments: zirconia-based crowns vs. lithium disilicate crowns", The Journal of Advanced of Prosthodontics, 10(1), pp.65~72 (2018), 대한치과보철학회[DOI]
- Matthias Rödiger, "Evaluation of removal forces of implant-supported zirconia copings depending on abutment geometry, luting agent and cleaning method during re-cementation", The Journal of Advanced of Prosthodontics, 6(3), pp.233~240 (2014), 대한치과보철학회[DOI]
- Mehl C, "Influence of abutment height and thermocycling on retrievability of cemented implant-supported crowns", Int J Oral Maxillofac Implants, 27, pp.1106~1115 (2012)
- Mehl C, "In vitro influence of ultrasonic stress, removal force preload and thermocycling on the retrievability of implant-retained crowns", Clin Oral Implants Res, 23, pp.930~937 (2012)[DOI]
- Pintinha M, "Effect of mechanical loading on the removal torque of different types of tapered connection abutments for dental implants", J Prosthet Dent, 110, pp.383~388 (2013)[DOI]
- Mehl C, "Influence of cement film thickness on the retention of implant-retained crowns", J Prosthodont, 22, pp.618~625 (2013)[DOI]
- Abbo B, "Resistance to dis- lodgement of zirconia copings cemented onto titanium abut- ments of different heights", J Prosthet Dent, 99, pp.25~29 (2008)[DOI]
- Sina Safari, "Effects of abutment diameter, luting agent type, and re-cementation on the retention of implant-supported CAD", CAM metal copings over short abutments, 2018, pp.10(1), 1~7[DOI]
- Covey DA, "Effects of abutment size and luting cement type on the uniaxial retention force of implant-supported crowns", J Prosthet Dent, 83, pp.344~348 (2000)[DOI]
- Bresciano M, "Retention of luting agents on implant abutments of different height and taper", Clin Oral Implants Res, 16, pp.594~598 (2005)[DOI]
- Mehl C, "Retrievability of implant-retained crowns following cementation", Clin Oral Implants Res, 19, pp.1304~1311 (2008)[DOI]
- Heinemann F, "Retrospective evaluation of temporary cemented, tooth and implant supported fixed partial dentures", J Craniomaxillofac Surg, 34, pp.86~90 (2006)
- Lim HP, "Fracture load of implant-supported zirconia all-ceramic crowns luted with various cements", Int J Prosthodont, 23, pp.361~363 (2010)
- Arrais CA, "Microtensile bond strength of dual-polymerizing cementing systems to dentin using different polymerizing modes", J Prosthet Dent, 97, pp.99~106 (2007)[DOI]
- Luhrs AK, "Composite cements benefit from light-curing", Dent Mater, 30, pp.292~301 (2014)[DOI]
- 염승현, "Effect of dentin pretreatment and curing mode on the microtensile bond strength of self-adhesive resin cements", The Journal of Advanced of Prosthodontics, 7(4), pp.317~322 (2015), 대한치과보철학회[DOI]
- Pan YH, "The effect of luting agents on the retention of dental implant-supported crowns", Chang Gung Med J, 28, pp.403~410 (2005)
- Harder S, "Molecular leakage at implant-abutment connection--in vitro investigation of tightness of internal conical implant-abutment connections against endotoxin penetration", Clin Oral Investig, 14, pp.427~432 (2010)[DOI]
- Chaar MS, "Prosthetic outcome of cementretained implant-supported fixed dental restorations: a systematic review", J Oral Rehabil, 38, pp.697~711 (2011)[DOI]