A comparative study to measure the sagittal condylar inclination using mechanical articulator, virtual articulator and jaw tracking device

1

조회

0

다운로드

2

인용

0

참고문헌

키워드

초록

Purpose. To compare the sagittal condylar inclination (SCI) in dentate individuals measured by the different methods with mechanical articulator (MA), virtual articulator (VA), and a jaw tracking device (JTD) system. Materials and methods. A total of 22 healthy dentate participants were enrolled in this study. For MA workflow, the SCI was obtained by a semi-adjustable articulator with protrusive interocclusal records. The SCI was also set on a VA by aligning intraoral scan (IOS) with cone beam computed tomography (CBCT) and facial scan (FS), respectively. These virtual workflows were conducted in a dental design software, namely VAIOS-CBCT and VAIOS-FS. Meanwhile, a JTD system was also utilized to perform the measurement. Intraclass correlation was used to assess the repeatability within workflows. The bilateral SCI values were compared by Wilcoxon matched-pairs signed rank test for each workflow, and Kruskal-Wallis test and post hoc p-value Bonferroni correction were used to compare the differences among four workflows. The agreement of VAIOS-CBCT, VAIOS-FS, and JTD compared with MA was evaluated by Bland-Altman analysis. Results. Intraclass correlation of the SCI revealed a high degree of repeatability for each workflow. There were no significant differences between the left and right sides (P > .05), except for VAIOS-CBCT (P = .028). Significant differences were not found between MA and VAIOS-FS (P > .05). Bland-Altman plots indicated VAIOS-CBCT, VAIOS-FS, and JTD were considered to substitute MA with high 95% limits of agreement. Conclusion. The workflow of VAIOS-FS provided an alternative approach to measure the SCI compared with MA.

Abstract

Purpose. To compare the sagittal condylar inclination (SCI) in dentate individuals measured by the different methods with mechanical articulator (MA), virtual articulator (VA), and a jaw tracking device (JTD) system. Materials and methods. A total of 22 healthy dentate participants were enrolled in this study. For MA workflow, the SCI was obtained by a semi-adjustable articulator with protrusive interocclusal records. The SCI was also set on a VA by aligning intraoral scan (IOS) with cone beam computed tomography (CBCT) and facial scan (FS), respectively. These virtual workflows were conducted in a dental design software, namely VAIOS-CBCT and VAIOS-FS. Meanwhile, a JTD system was also utilized to perform the measurement. Intraclass correlation was used to assess the repeatability within workflows. The bilateral SCI values were compared by Wilcoxon matched-pairs signed rank test for each workflow, and Kruskal-Wallis test and post hoc p-value Bonferroni correction were used to compare the differences among four workflows. The agreement of VAIOS-CBCT, VAIOS-FS, and JTD compared with MA was evaluated by Bland-Altman analysis. Results. Intraclass correlation of the SCI revealed a high degree of repeatability for each workflow. There were no significant differences between the left and right sides (P > .05), except for VAIOS-CBCT (P = .028). Significant differences were not found between MA and VAIOS-FS (P > .05). Bland-Altman plots indicated VAIOS-CBCT, VAIOS-FS, and JTD were considered to substitute MA with high 95% limits of agreement. Conclusion. The workflow of VAIOS-FS provided an alternative approach to measure the SCI compared with MA.

참고문헌 (30)

  1. Reicheneder C, "Variations in the inclination of the condylar path in children and adults", Angle Orthod, 79, pp.958~963 (2009)
  2. Cimić S, "Articulator-related registration and analysis of sagittal condylar inclination", Acta Clin Croat, 54, pp.432~437 (2015)
  3. Hobo S, "Articulator selection for restorative dentistry", J Prosthet Dent, 36, pp.35~43 (1976)
  4. Oancea L, "The influence of temporomandibular joint movement parameters on dental morphology", Ann Anat, 218, pp.49~58 (2018)
  5. Farias-Neto A, "Face-bow transfer in prosthodontics : a systematic review of the literature", J Oral Rehabil, 40, pp.686~692 (2013)
  6. Hangai K, "Clinical evaluation of semi-adjustable articulators : reproducibility of sagittal condylar path inclination assessed by a jaw-tracking system with six degrees of freedom", Nihon Hotetsu Shika Gakkai Zasshi, 52, pp.360~365 (2008)
  7. Donegan SJ, "Sagittal condylar guidance as determined by protrusion records and wear facets of teeth", Int J Prosthodont, 4, pp.469~472 (1991)
  8. Christensen C, "The problem of the bite", Dent Cosmos, 47, pp.1184~1905 (1905)
  9. Li Q, "The creation of a virtual dental patient with dynamic occlusion and its application in esthetic dentistry", J Prosthet Dent, 126, pp.14~18 (2021)
  10. Ahlers MO, "Motion analysis of the mandible : guidelines for standardized analysis of computer-assisted recording of condylar movements", Int J Comput Dent, 18, pp.201~223 (2015)
  11. Aslanidou K, "The fabrication of a customized occlusal splint based on the merging of dynamic jaw tracking records, cone beam computed tomography, and CAD-CAM digital impression", J Orthod Sci, 6, pp.104~109 (2017)
  12. Celar AG, "Accuracy of recording horizontal condylar inclination and Bennett angle with the Cadiax compact", J Oral Rehabil, 29, pp.1076~1081 (2002)
  13. Hong SJ, "Setting the sagittal condylar inclination on a virtual articulator by using a facial and intraoral scan of the protrusive interocclusal position : A dental technique", J Prosthet Dent, 125, pp.392~395 (2021)
  14. Yang S, "A novel technique to align the intraoral scans to the virtual articulator and set the patient-specific sagittal condylar inclination", J Prosthodont, 31, pp.79~84 (2022)
  15. Hong SJ, "Setting the sagittal condylar inclination on a virtual articulator using intraoral scan of protrusive interocclusal position and cone beam computed tomography", J Prosthodont, 29, pp.185~189 (2020)
  16. Solaberrieta E, "Virtual facebow technique", J Prosthet Dent, 114, pp.751~755 (2015)
  17. Lepidi L, "A full-digital technique to mount a maxillary arch scan on a virtual articulator", J Prosthodont, 28, pp.335~338 (2019)
  18. Raghav D, "Intricate relations and concepts of reference points in prosthodontics : A literature review", Eur J Prosthodont, 4, pp.1~6 (2016)
  19. Wilkie ND, "The anterior point of reference", J Prosthet Dent, 41, pp.488~496 (1979)
  20. Kuralt M, "Evaluation of gingival recessions with conventional versus digital methods", J Dent, 120, pp.104093~ (2022)
  21. Ahangari AH, "Evaluation of the Cadiax Compact® II accuracy in recording preadjusted condylar inclinations on fully adjustable articulator", J Contemp Dent Pract, 13, pp.504~508 (2012)
  22. Anderson GC, "An in vitro study of an electronic pantograph", J Prosthet Dent, 57, pp.577~580 (1987)
  23. Chang WS, "An in vitro evaluation of the reliability and validity of an electronic pantograph by testing with five different articulators", J Prosthet Dent, 92, pp.83~89 (2004)
  24. Ercoli C, "Face-bow record without a third point of reference : theoretical considerations and an alternative technique", J Prosthet Dent, 82, pp.237~241 (1999)
  25. Cimić S, "Measurements of the sagittal condylar inclination : intraindividual variations", Cranio, 32, pp.104~109 (2014)
  26. Matsumura H, "Analysis of sagittal condylar path inclination in consideration of Fischer’s angle", J Oral Rehabil, 33, pp.514~519 (2006)
  27. McCullock AJ, "Making occlusion work: 2. Practical considerations", Dent Update, 30, pp.211~216 (2003)
  28. Kattadiyil MT, "The relationship between centric occlusion and the maximal intercuspal position and their use as treatment positions for complete mouth rehabilitation : best evidence consensus statement", J Prosthodont, 30, pp.26~33 (2021)
  29. Weinberg LA, "An evaluation of basic articulators and their concepts: Part II. Arbitrary, positional, semi adjustable articulators", J Prosthet Dent, 13, pp.645~663 (1963)
  30. Tejo SK, "A comparative evaluation of dimensional stability of three types of interocclusal recording materials-an in-vitro multi-centre study", Head Face Med, 8, pp.27~ (2012)

관련 논문AI추천

Impact of digital jaw tracking systems on dynamic occlusal surface morphology and condylar inclination measurementsAlper Aktosun, Ayşenur Özcan-Sezgin, Berk Bilgen, Sina Saygılı, Tonguç Sülün대한치과보철학회 · The Journal of Advanced of Prosthodontics2025 · 제17권 제4호Correlation between sagittal condylar guidance angles obtained using radiographic and protrusive occlusal record methods권오균, 김지환, 양승원대한치과보철학회 · The Journal of Advanced of Prosthodontics2017 · 제9권 제4호Evaluation of relative deviations in virtual articulators for simulating occlusal contacts at increased vertical dimensionsBerk Bilgen, Elvan Biren, Sina Saygılı, Tonguç Sülün대한치과보철학회 · The Journal of Advanced of Prosthodontics2025 · 제17권 제6호Recording maximal intercuspation and border positions of the mandible with intraoral scanner using the acquisition software’s multi-occlusion functionIhab Hammad, Noha Morsy대한치과보철학회 · The Journal of Advanced of Prosthodontics2024 · 제16권 제4호A comparative study to measure the condylar guidance by the radiographic and clinical methodsAshu Bhalla, Gunjan Pruthi, Pragya Shreshta, Veena Jain대한치과보철학회 · The Journal of Advanced of Prosthodontics2012 · 제4권 제3호Three-dimensional analysis of the outcome of different scanning strategies in virtual interocclusal registrationFei Liu, Jiansong Mei, Jiarui Chao, Jiefei Shen, Liya Ma대한치과보철학회 · The Journal of Advanced of Prosthodontics2022 · 제14권 제6호

이 간행물의 다른 논문 (4건)