Background: The application of image analysis technologies for the interpretation of microbiological cultures is evolving rapidly. The primary aim of this study was to establish whether the image analysis system named Automated Plate Assessment System (APAS; LBT Innovations Ltd., Australia) could be applied to screen urine cultures. A secondary aim was to evaluate differences between traditional plate reading (TPR) and the reading of cultures from images, or digital plate reading (DPR). Methods: A total of 9,224 urine samples submitted for culture to three clinical laboratories, two in Australia and one in the USA, were included in the study. Cultures were prepared on sheep blood and MacConkey agar plates and read by panels of three microbiologists. The plates were then presented to APAS for image capture and analysis, and the images and results were stored for later review. Results: Image analysis of cultures using APAS produced a diagnostic sensitivity and specificity of 99.0% and 84.5%, respectively. Colonies were detected by APAS on 99.0% of blood agar plates with growth and on 99.5% of MacConkey agar plates. DPR agreed with TPR for colony enumeration on 92.1% of the plates, with a sensitivity of 90.8% and specificity of 92.8% for case designation. However, several differences in the classification of colony morphologies using DPR were identified. Conclusions: APAS was shown to be a reliable screening system for urine cultures. The study also showed acceptable concordance between DPR and TPR for colony detection, enumeration, and case designation.

Abstract

Background: The application of image analysis technologies for the interpretation of microbiological cultures is evolving rapidly. The primary aim of this study was to establish whether the image analysis system named Automated Plate Assessment System (APAS; LBT Innovations Ltd., Australia) could be applied to screen urine cultures. A secondary aim was to evaluate differences between traditional plate reading (TPR) and the reading of cultures from images, or digital plate reading (DPR). Methods: A total of 9,224 urine samples submitted for culture to three clinical laboratories, two in Australia and one in the USA, were included in the study. Cultures were prepared on sheep blood and MacConkey agar plates and read by panels of three microbiologists. The plates were then presented to APAS for image capture and analysis, and the images and results were stored for later review. Results: Image analysis of cultures using APAS produced a diagnostic sensitivity and specificity of 99.0% and 84.5%, respectively. Colonies were detected by APAS on 99.0% of blood agar plates with growth and on 99.5% of MacConkey agar plates. DPR agreed with TPR for colony enumeration on 92.1% of the plates, with a sensitivity of 90.8% and specificity of 92.8% for case designation. However, several differences in the classification of colony morphologies using DPR were identified. Conclusions: APAS was shown to be a reliable screening system for urine cultures. The study also showed acceptable concordance between DPR and TPR for colony detection, enumeration, and case designation.

참고문헌 (19)

  1. Bourbeau PP, "Automation in clinical microbiology", J Clin Microbiol, 51, pp.1658~1665 (2013)
  2. Ledeboer NA, "The automated clinical microbiology laboratory: fact or fantasy?", J Clin Microbiol, 52, pp.3140~3146 (2014)
  3. Nico T. Mutters, "Performance of Kiestra Total Laboratory Automation Combined with MS in Clinical Microbiology Practice", Annals of Laboratory Medicine, 34(2), pp.117~123 (2014), 대한진단검사의학회
  4. Rhoads DD, "A review of the current state of digital plate reading of cultures in clinical microbiology", J Pathol Inform, 6, pp.23~ (2015)
  5. Miyazawa K, "In situ detection and identification of microorganisms at single-colony resolution by spectral imaging", Opt Rev, 15, pp.285~291 (2008)
  6. Yoon SC, "Detection of vv colonies using hyperspectral imaging", Sens Instrumen Food Qual Saf, 4, pp.35~49 (2010)
  7. Faron ML, "Automated scoring of chromogenic media for detection of methicillin-resistant Staphylococcus aureus by use of WASPLab image analysis software", J Clin Microbiol, 54, pp.620~624 (2016)
  8. Glasson J, "Evaluation of an image analysis device (APAS) for screening urine cultures", J Clin Microbiol, 54, pp.300~304 (2016)
  9. McCarter YS, "Coordinating" (2009), ASM Press
  10. Kouri T, "ECLM. European urinalysis guidelines", Scand J Clin Lab Invest, 60(S231), pp.38~ (2000)
  11. Grude N, "Urinary tract infections in Norway: bacterial etiology and susceptibility. A retrospective study of clinical isolates", Clin Microbiol Infect, 7, pp.543~547 (2001)
  12. Karlowsky JA, "Antimicrobial resistance in urinary tract pathogens in Canada from 2007 to 2009: CANWARD Surveillance Study", Antimicrob Agents Chemother, 55, pp.3169~3175 (2011)
  13. Kahlmeter G, "An international survey of the antimicrobial susceptibility of pathogens from uncomplicated urinary tract infections: the ECO.SENS Project", J Antimicrob Chemother, 51, pp.69~76 (2003)
  14. Zhanel GG, "Antibiotic resistance in outpatient isolates: final results from the North American Urinary Tract Infection Collaborative Alliance (NAUTICA)", Int J Antimicrob Agents, 26, pp.380~388 (2005)
  15. Gupta K, "Increasing prevalence of antimicrobial resistance among uropathogens causing acute uncomplicated cystitis in women", JAMA, 281, pp.736~738 (1999)
  16. Murray PR, "Evaluation of the modified Bac-T-Screen and FiltraCheck-UTI urine screening systems for detection of clinically significant bacteriuira", J Clin Microbiol, 26, pp.2347~2350 (1988)
  17. Manoni F, "Cutoff values for bacteria and leukocytes for urine flow cytometer Sysmex UF-1000i in urinary tract infections", Diagn Microbiol and Infect Dis, 65, pp.103~107 (2009)
  18. Jolkkonen S, "Screening urine samples by flow cytometry reduces the need for culture", J Clin Microbiol, 48, pp.3117~3121 (2010)
  19. Dauwalder O, "Does bacteriology laboratory automation reduce time to results and increase quality management?", Clin Microbio Infect, 22, pp.236~243 (2016)

관련 논문AI추천

Shorter Incubation Times for Detecting Multi-drug Resistant Bacteria in Patient Samples: Defining Early Imaging Time Points Using Growth Kinetics and Total Laboratory AutomationIrene Burckhardt, Katharina Last,, Stefan Zimmermann,대한진단검사의학회 · Annals of Laboratory Medicine2019 · 제39권 제1호Evaluation of Five Automated Urine Analyzers as Screening Instruments for Enhancing Diagnostic Efficiency in Urinary Tract Infection김정호, 변정현, 오경진, 용동은, 이상국, 전범찬, 조주영대한임상미생물학회 · Annals of Clinical Microbiology2021 · 제24권 제3호Performance of Kiestra Total Laboratory Automation Combined with MS in Clinical Microbiology PracticeCaroline E.Visser, Caspar J. Hodiamont, Hendri P. J. Overmeijer, Mandy van den Boogaard, Menno D. de Jong, Nico T. Mutters대한진단검사의학회 · Annals of Laboratory Medicine2014 · 제34권 제2호Performance of the cobas u 701 Analyzer in Urinary Tract Infection ScreeningAna Infante, Coral Martin, Félix Gutiérrez, Fernando Buñuel, Gregoria Gázquez, Victoria Ortiz de la Tabla,대한진단검사의학회 · Annals of Laboratory Medicine2019 · 제39권 제5호체액 및 요검체 접종을 위한 자동 접종장비 PREVI Isola^®의 유용성 평가김윤정, 손영숙, 용동은, 윤서영, 이경원, 이양순, 이운형, 정석훈, 정윤섭, 정혜선대한진단검사의학회 · Laboratory Medicine Online2011 · 제1권 제2호Performance of Modified-EUCAST Rapid Direct Antimicrobial Susceptibility Testing on Clinical Urine Samples권미정, 김대원, 용동은, 이상국, 이혁민, 저스틴 무지라네자대한임상미생물학회 · Annals of Clinical Microbiology2020 · 제23권 제3호