이동식 생물안전 연구시설 내 공조공기 급·배기구의 위치선정에 관한 해석적 연구
Analytical Study on Location of Supply and Exhaust Air in Mobile Biosafety Research Facility
0
조회
0
다운로드
0
인용
0
참고문헌
키워드
초록
Recently a biosafety level-3(BL-3) mobile laboratory has been set up for the virus scanning and vaccine development because of the COVID-19 pandemic. The study on air flow inlet and outlet location and its flow direction with ventilation in the mobile laboratory needs to prevent spread of COVID-19 virus because the COVID-19 virus is primarily transmitted to people through respiratory droplets and aerosol coming out as their coughing. This study is conducted on the air flow pattern optimization in BL-3 mobile laboratory with various design specifications of position of air supply & exhaust port and particle source. Air flow patterns of ceiling supply-exhaust and ceiling supply-bottom side exhaust with particle source were determined to compare the impact of the infection prevention. CFD simulation was used to analyze for two air flow patterns and particle source position. Numerical results showed that air flow pattern of air conditioning system with ceiling supply-exhaust in a row is more effective than that of ceiling supply-bottom side exhaust air flow pattern in terms of infection prevention in biosafety mobile laboratory.
Abstract
Recently a biosafety level-3(BL-3) mobile laboratory has been set up for the virus scanning and vaccine development because of the COVID-19 pandemic. The study on air flow inlet and outlet location and its flow direction with ventilation in the mobile laboratory needs to prevent spread of COVID-19 virus because the COVID-19 virus is primarily transmitted to people through respiratory droplets and aerosol coming out as their coughing. This study is conducted on the air flow pattern optimization in BL-3 mobile laboratory with various design specifications of position of air supply & exhaust port and particle source. Air flow patterns of ceiling supply-exhaust and ceiling supply-bottom side exhaust with particle source were determined to compare the impact of the infection prevention. CFD simulation was used to analyze for two air flow patterns and particle source position. Numerical results showed that air flow pattern of air conditioning system with ceiling supply-exhaust in a row is more effective than that of ceiling supply-bottom side exhaust air flow pattern in terms of infection prevention in biosafety mobile laboratory.
참고문헌 (8)
- Morawaska L, "Airborne Transmission of SARS-CoV-2 : The World should Face The Reality", Environment International, 139(105730) (2022)
- World Health Organization (WHO), "Infection Prevention and Control of Epidemic and Pandemic Prone Acute Respiratory Infections in Healthcare", ~ (2014)
- "Environmental Factors Affecting the Transmission of Respiratory Viruses", Current Opinion in Virology, 2(1), pp.90~95 (2012)
- Li, H., "Dispersion of Evaporating Cough Droplets in Tropical Outdoor Environment", Physics of Fluids, 32(113301) (2020)
- Prateek Bahl, "Airborne or Droplet Precautions for Health Workers Treating Coronavirus Disease 2019", The Journal of Infectious Diseases (2019)[DOI]
- Bourouiba, L, "Violent Expiratory Events: On Coughing and Sneezing", Journal of Fluid Mechanics, 745, pp.537~563 (2014)