Objectives. The loss of signal during intraoperative neuromonitoring (IONM) using electromyography (EMG) in thyroidectomy is one of the biggest problems. We have developed a novel IONM system with an endotracheal tube (ETT) with an attached pressure sensor instead of EMG to detect laryngeal twitching. The aim of the present study was to investigate the feasibility and reliability of this novel IONM system using an ETT with pressure sensor during thyroidectomy in a porcine model. Methods. We developed an ETT-attached pressure sensor that uses the piezoelectric effect to measure laryngeal muscle twitching. Stimulus thresholds, amplitude, and latency of laryngeal twitching evaluated using the pressure sensor were compared to those measured using transcartilage needle EMG. The measured amplitude changes by EMG and the pressure sensor during recurrent laryngeal nerve (RLN) traction injury were compared. Results. No significant differences in stimulus threshold intensity between EMG and the pressure sensor were observed. The EMG amplitude detected at 0.3 mA, increased with increasing stimulus intensity. When the stimulus was more than 1.0 mA, the amplitude showed a plateau. In a RLN traction injury experiment, the EMG amplitude did not recover even 20 minutes after stopping RLN traction. However, the pressure sensor showed a mostly recovery. Conclusion. The change in amplitude due to stimulation of the pressure sensor showed a pattern similar to EMG. Pressure sensors can be feasibly and reliably used for RLN traction injury prediction, RLN identification, and preservation through the detection of laryngeal muscle twitching. Our novel IONM system that uses an ETT with an attached pressure sensor to measure the change of surface pressure can be an alternative to EMG in the future.

Abstract

Objectives. The loss of signal during intraoperative neuromonitoring (IONM) using electromyography (EMG) in thyroidectomy is one of the biggest problems. We have developed a novel IONM system with an endotracheal tube (ETT) with an attached pressure sensor instead of EMG to detect laryngeal twitching. The aim of the present study was to investigate the feasibility and reliability of this novel IONM system using an ETT with pressure sensor during thyroidectomy in a porcine model. Methods. We developed an ETT-attached pressure sensor that uses the piezoelectric effect to measure laryngeal muscle twitching. Stimulus thresholds, amplitude, and latency of laryngeal twitching evaluated using the pressure sensor were compared to those measured using transcartilage needle EMG. The measured amplitude changes by EMG and the pressure sensor during recurrent laryngeal nerve (RLN) traction injury were compared. Results. No significant differences in stimulus threshold intensity between EMG and the pressure sensor were observed. The EMG amplitude detected at 0.3 mA, increased with increasing stimulus intensity. When the stimulus was more than 1.0 mA, the amplitude showed a plateau. In a RLN traction injury experiment, the EMG amplitude did not recover even 20 minutes after stopping RLN traction. However, the pressure sensor showed a mostly recovery. Conclusion. The change in amplitude due to stimulation of the pressure sensor showed a pattern similar to EMG. Pressure sensors can be feasibly and reliably used for RLN traction injury prediction, RLN identification, and preservation through the detection of laryngeal muscle twitching. Our novel IONM system that uses an ETT with an attached pressure sensor to measure the change of surface pressure can be an alternative to EMG in the future.

참고문헌 (37)

  1. Haugen BR, "2015 American Thyroid Association Management Guidelines for Adult Patients with Thyroid Nodules and Differentiated Thyroid Cancer : The American Thyroid Association Guidelines Task Force on Thyroid Nodules and Differentiated Thyroid Cancer", Thyroid, 26(1), pp.1~133 (2016)
  2. Chandrasekhar SS, "Clinical practice guideline: improving voice outcomes after thyroid surgery", Otolaryngol Head Neck Surg, 148(6 Suppl), pp.S1~S37 (2013)
  3. Jeannon JP, "Diagnosis of recurrent laryngeal nerve palsy after thyroidectomy : a systematic review", Int J Clin Pract, 63(4), pp.624~629 (2009)
  4. Riddell V, "Thyroidectomy: prevention of bilateral recurrent nerve palsy. Results of identification of the nerve over 23 consecutive years (1946-69) with a description of an additional safety measure", Br J Surg, 57(1), pp.1~11 (1970)
  5. Hermann M, "Laryngeal recurrent nerve injury in surgery for benign thyroid diseases : effect of nerve dissection and impact of individual surgeon in more than 27, 000 nerves at risk", Ann Surg, 235(2), pp.261~268 (2002)
  6. Chiang FY, "Recurrent laryngeal nerve palsy after thyroidectomy with routine identification of the recurrent laryngeal nerve", Surgery, 137(3), pp.342~347 (2005)
  7. Randolph GW, "Electrophysiologic recurrent laryngeal nerve monitoring during thyroid and parathyroid surgery: international standards guideline statement", Laryngoscope, 121(Suppl 1), pp.S1~S16 (2011)
  8. Sung ES, "Development of a novel detachable magnetic nerve stimulator for intraoperative neuromonitoring", World J Surg, 42(1), pp.137~142 (2018)
  9. Sung ES, "Development of an attachable endoscopic nerve stimulator for intraoperative neuromonitoring during endoscopic or robotic thyroidectomy", Otolaryngol Head Neck Surg, 158(3), pp.465~468 (2018)
  10. Shin SC, "Feasibility and safety of nerve stimulator attachment to energy-based devices : a porcine model study", Int J Surg, 48, pp.155~159 (2017)
  11. Snyder SK, "Intraoperative neurophysiology testing of the recurrent laryngeal nerve : plaudits and pitfalls", Surgery, 138(6), pp.1183~1191 (2005)
  12. Beldi G, "Evaluation of intraoperative recurrent nerve monitoring in thyroid surgery", World J Surg, 28(6), pp.589~591 (2004)
  13. Lu IC, "Optimal depth of NIM EMG endotracheal tube for intraoperative neuromonitoring of the recurrent laryngeal nerve during thyroidectomy", World J Surg, 32(9), pp.1935~1939 (2008)
  14. Wu CW, "Loss of signal in recurrent nerve neuromonitoring : causes and management", Gland Surg, 4(1), pp.19~26 (2015)
  15. Dionigi G, "What is the learning curve for intraoperative neuromonitoring in thyroid surgery?", Int J Surg, 6(Suppl 1), pp.S7~S12 (2008)
  16. Dralle H, "Intraoperative monitoring of the recurrent laryngeal nerve in thyroid surgery", World J Surg, 32(7), pp.1358~1366 (2008)
  17. Khan A, "Experience with two types of electromyography monitoring electrodes during thyroid surgery", Am J Otolaryngol, 18(2), pp.99~102 (1997)
  18. Tschopp KP, "Comparison of various methods of electromyographic monitoring of the recurrent laryngeal nerve in thyroid surgery", Ann Otol Rhinol Laryngol, 111(9), pp.811~816 (2002)
  19. Rea JL, "Clinical evoked electromyography for recurrent laryngeal nerve preservation : use of an endotracheal tube electrode and a postcricoid surface electrode", Laryngoscope, 108(9), pp.1418~1420 (1998)
  20. Marcus B, "Recurrent laryngeal nerve monitoring in thyroid and parathyroid surgery : the University of Michigan experience", Laryngoscope, 113(2), pp.356~361 (2003)
  21. Liddy W, "Monitoring of the posterior cricoarytenoid muscle represents another option for neural monitoring during thyroid surgery : normative vagal and recurrent laryngeal nerve posterior cricoarytenoid muscle electromyographic data", Laryngoscope, 128(1), pp.283~289 (2018)
  22. Wu CW, "Feasibility of intraoperative neuromonitoring during thyroid surgery using transcartilage surface recording electrodes", Thyroid, 28(11), pp.1508~1516 (2018)
  23. Wu CW, "Transcutaneous recording during intraoperative neuromonitoring in thyroid surgery", Thyroid, 28(11), pp.1500~1507 (2018)
  24. Chiang FY, "The mechanism of recurrent laryngeal nerve injury during thyroid surgery : the application of intraoperative neuromonitoring", Surgery, 143(6), pp.743~749 (2008)
  25. Chiang FY, "Standardization of intraoperative neuromonitoring of recurrent laryngeal nerve in thyroid operation", World J Surg, 34(2), pp.223~229 (2010)
  26. Schneider R, "Continuous intraoperative vagus nerve stimulation for identification of imminent recurrent laryngeal nerve injury", Head Neck, 35(11), pp.1591~1598 (2013)
  27. 성의숙, "Development of a Novel Intraoperative Neuromonitoring System Using a Surface Pressure Sensor to Detect Muscle Movement: A Rabbit Model Study", Clinical and Experimental Otorhinolaryngology, 12(2), pp.217~223 (2019), 대한이비인후과학회[DOI]
  28. Sanabria A, "Neuromonitoring of the laryngeal nerves in thyroid surgery : a critical appraisal of the literature", Eur Arch Otorhinolaryngol, 270(9), pp.2383~2395 (2013)
  29. Lifante JC, "A prospective, randomized trial of nerve monitoring of the external branch of the superior laryngeal nerve during thyroidectomy under local", regional anesthesia and IV sedation, 2009, pp.146(6), 1167~1173
  30. Barczynski M, "Randomized clinical trial of visualization versus neuromonitoring of recurrent laryngeal nerves during thyroidectomy", Br J Surg, 96(3), pp.240~246 (2009)
  31. Higgins TS, "Recurrent laryngeal nerve monitoring versus identification alone on post-thyroidectomy true vocal fold palsy : a meta-analysis", Laryngoscope, 121(5), pp.1009~1017 (2011)
  32. Dionigi G, "Neuromonitoring and video-assisted thyroidectomy : a prospective, randomized case-control evaluation", Surg Endosc, 23(5), pp.996~1003 (2009)
  33. Chan WF, "Pitfalls of intraoperative neuromonitoring for predicting postoperative recurrent laryngeal nerve function during thyroidectomy", World J Surg, 30(5), pp.806~812 (2006)
  34. Bai B, "Protective effects of intraoperative nerve monitoring(IONM)for recurrent laryngeal nerve injury in thyroidectomy : meta-analysis", Sci Rep, 8(1), pp.7761~ (2018)
  35. Chiang FY, "Comparison of EMG signals recorded by surface electrodes on endotracheal tube and thyroid cartilage during monitored thyroidectomy", Kaohsiung J Med Sci, 33(10), pp.503~509 (2017)
  36. 성의숙, "Development of a Novel Intraoperative Neuromonitoring System Using an Accelerometer Sensor in Thyroid Surgery: A Porcine Model Study", Clinical and Experimental Otorhinolaryngology, 12(4), pp.420~426 (2019), 대한이비인후과학회[DOI]
  37. Randolph GW, "Recurrent laryngeal nerve identification and assessment during thyroid surgery : laryngeal palpation", World J Surg, 28(8), pp.755~760 (2004)

관련 논문AI추천

Development of a Novel Intraoperative Neuromonitoring System Using an Accelerometer Sensor in Thyroid Surgery: A Porcine Model Study권현근, 나한슬, 노정훈, 박다희, 성의숙, 신성찬, 이병주, 이진춘, 최성욱대한이비인후과학회 · Clinical and Experimental Otorhinolaryngology2019 · 제12권 제4호Development of a Novel Intraoperative Neuromonitoring System Using a Surface Pressure Sensor to Detect Muscle Movement: A Rabbit Model Study권현근, 김동주, 김민식, 노정훈, 성의숙, 신성찬, 이병주, 이진춘대한이비인후과학회 · Clinical and Experimental Otorhinolaryngology2019 · 제12권 제2호Efficacy of Transcartilaginous Electrodes for Intraoperative Neural Monitoring During Thyroid Surgery태경, 송창면, 이승환, 정선민, 정진혁, 지용배대한이비인후과학회 · Clinical and Experimental Otorhinolaryngology2020 · 제13권 제4호갑상선 수술을 위한 새로운 수술 중 신경감시시스템의개발성의숙, 이병주대한갑상선학회 · International Journal of Thyroidology2018 · 제11권 제2호갑상선 수술 중 바늘과 피부 전극을 이용한 새로운 신경감시술서명구, 신성찬, 이병주, 천용일대한갑상선학회 · International Journal of Thyroidology2022 · 제15권 제1호피부 부착형 전극을 이용한 갑상선 수술 중 신경 모니터링의 효용성김성원, 서상규, 이강대, 이형신, 조기헌대한이비인후과학회 · 대한이비인후과학회지 두경부외과학2021 · 제64권 제6호

이 간행물의 다른 논문 (15건)