본문 바로가기
KERD
논문KCI 등재

Efficacy of Transcartilaginous Electrodes for Intraoperative Neural Monitoring During Thyroid Surgery

2

조회

0

다운로드

3

인용

0

참고문헌

키워드

초록

Objectives. This study was conducted evaluate the efficacy of electromyography (EMG) using transcartilaginous (TC) electrodes through the thyroid cartilage and perichondrium. Methods. We prospectively collected EMG data from intraoperative neuromonitoring (IONM) of 54 nerves at risk in 38 patients during thyroidectomy. We followed standardized IONM procedures in all operations. EMG signals from both endotracheal tube (ET) electrodes and TC needle electrodes were recorded simultaneously. We compared the characteristics of the EMG signals and the efficacy of both methods. Results. Significantly higher mean EMG amplitudes were recorded by TC electrodes than by ET electrodes in all four-step procedures (V1-R1-R2-V2, P<0.001). Loss of signal (LOS) occurred in five patients in ET electrodes, but in only two patients in TC electrodes. Postoperative laryngoscopy revealed recurrent laryngeal nerve palsy in the two patients who showed LOS from both the ET and TC electrodes, and vocal cord movement was intact in the other three patients. Therefore, the positive predictive values of LOS in ET and TC electrodes were 40% and 100%, respectively. Conclusion. EMG recording using TC needle electrodes is feasible and effective, making it a good alternative technique for IONM.

Abstract

Objectives. This study was conducted evaluate the efficacy of electromyography (EMG) using transcartilaginous (TC) electrodes through the thyroid cartilage and perichondrium. Methods. We prospectively collected EMG data from intraoperative neuromonitoring (IONM) of 54 nerves at risk in 38 patients during thyroidectomy. We followed standardized IONM procedures in all operations. EMG signals from both endotracheal tube (ET) electrodes and TC needle electrodes were recorded simultaneously. We compared the characteristics of the EMG signals and the efficacy of both methods. Results. Significantly higher mean EMG amplitudes were recorded by TC electrodes than by ET electrodes in all four-step procedures (V1-R1-R2-V2, P<0.001). Loss of signal (LOS) occurred in five patients in ET electrodes, but in only two patients in TC electrodes. Postoperative laryngoscopy revealed recurrent laryngeal nerve palsy in the two patients who showed LOS from both the ET and TC electrodes, and vocal cord movement was intact in the other three patients. Therefore, the positive predictive values of LOS in ET and TC electrodes were 40% and 100%, respectively. Conclusion. EMG recording using TC needle electrodes is feasible and effective, making it a good alternative technique for IONM.

참고문헌 (44)

  1. Vasileiadis I, "Association of intraoperative neuromonitoring with reduced recurrent laryngeal nerve injury in patients undergoing total thyroidectomy", JAMA Otolaryngol Head Neck Surg, 142(10), pp.994~1001 (2016)
  2. Riddell VH, "Injury to recurrent laryngeal nerves during thyroidectomy; a comparison between the results of identification and non-identification in 1022 nerves exposed to risk", Lancet, 271(6944), pp.638~641 (1956)
  3. Jatzko GR, "Recurrent nerve palsy after thyroid operations--principal nerve identification and a literature review", Surgery, 115(2), pp.139~144 (1994)
  4. Witt RL, "Recurrent laryngeal nerve electrophysiologic monitoring in thyroid surgery: the standard of care?", J Voice, 19(3), pp.497~500 (2005)
  5. Chan WF, "The role of intraoperative neuromonitoring of recurrent laryngeal nerve during thyroidectomy : a comparative study on 1000 nerves at risk", Surgery, 140(6), pp.866~872 (2006)
  6. Barczynski M, "Randomized clinical trial of visualization versus neuromonitoring of recurrent laryngeal nerves during thyroidectomy", Br J Surg, 96(3), pp.240~246 (2009)
  7. Flisberg K, "Electrical stimulation of the human recurrent laryngeal nerve during thyroid operation", Acta Otolaryngol Suppl, 263, pp.63~67 (1969)
  8. Gardner IH, "Intraoperative nerve monitoring during thyroid surgery", Curr Opin Endocrinol Diabetes Obes, 23(5), pp.394~399 (2016)
  9. Rea JL, "Postcricoid surface laryngeal electrode", Ear Nose Throat J, 71(6), pp.267~269 (1992)
  10. Tschopp K, "New aspects in surgery of the thyroid gland with intraoperative monitoring of the recurrent laryngeal nerve", Laryngorhinootologie, 73(11), pp.568~572 (1994)
  11. Mermelstein M, "Intraoperative identification of laryngeal nerves with laryngeal electromyography", Laryngoscope, 106(6), pp.752~756 (1996)
  12. Khan A, "Experience with two types of electromyography monitoring electrodes during thyroid surgery", Am J Otolaryngol, 18(2), pp.99~102 (1997)
  13. Maloney RW, "A new method for intraoperative recurrent laryngeal nerve monitoring", Ear Nose Throat J, 73(1), pp.30~33 (1994)
  14. 곽희용, "갑상선 수술 중 신경 모니터링의 현재와 미래", Journal of Endocrine Surgery, 14(2), pp.67~75 (2014), 대한내분비외과학회
  15. Randolph GW, "Electrophysiologic recurrent laryngeal nerve monitoring during thyroid and parathyroid surgery: international standards guideline statement", Laryngoscope, 121(Suppl 1), pp.S1~S16 (2011)
  16. Tsai CJ, "Electromyographic endotracheal tube placement during thyroid surgery in neuromonitoring of recurrent laryngeal nerve", Kaohsiung J Med Sci, 27(3), pp.96~101 (2011)
  17. Zhao Y, "Experimental study of needle recording electrodes placed on the thyroid cartilage for neuromonitoring during thyroid surgery", Br J Surg, 106(3), pp.245~254 (2019)
  18. Lu IC, "Optimal depth of NIM EMG endotracheal tube for intraoperative neuromonitoring of the recurrent laryngeal nerve during thyroidectomy", World J Surg, 32(9), pp.1935~1939 (2008)
  19. Kim HY, "Impact of positional changes in neural monitoring endotracheal tube on amplitude and latency of electromyographic response in monitored thyroid surgery : results from the porcine experiment", Head Neck, 38(Suppl 1), pp.E1004~E1008 (2016)
  20. Chiang FY, "Comparison of EMG signals recorded by surface electrodes on endotracheal tube and thyroid cartilage during monitored thyroidectomy", Kaohsiung J Med Sci, 33(10), pp.503~509 (2017)
  21. Chiang FY, "Anatomical variations of recurrent laryngeal nerve during thyroid surgery : how to identify and handle the variations with intraoperative neuromonitoring", Kaohsiung J Med Sci, 26(11), pp.575~583 (2010)
  22. Spahn JG, "Identification of the motor laryngeal nerves : a new electrical stimulation technique", Laryngoscope, 91(11), pp.1865~1868 (1981)
  23. Brauckhoff M, "Identification of the non-recurrent inferior laryngeal nerve using intraoperative neurostimulation", Langenbecks Arch Surg, 386(7), pp.482~487 (2002)
  24. Dralle H, "Intraoperative monitoring of the recurrent laryngeal nerve in thyroid surgery", World J Surg, 32(7), pp.1358~1366 (2008)
  25. Bergenfelz A, "Risk of recurrent laryngeal nerve palsy in patients undergoing thyroidectomy with and without intraoperative nerve monitoring", Br J Surg, 103(13), pp.1828~1838 (2016)
  26. Pisanu A, "Systematic review with meta-analysis of studies comparing intraoperative neuromonitoring of recurrent laryngeal nerves versus visualization alone during thyroidectomy", J Surg Res, 188(1), pp.152~161 (2014)
  27. Shedd DP, "Identification of the recurrent laryngeal nerve", Arch Surg, 92(6), pp.861~864 (1966)
  28. Chiang FY, "Intraoperative neuromonitoring for early localization and identification of recurrent laryngeal nerve during thyroid surgery", Kaohsiung J Med Sci, 26(12), pp.633~639 (2010)
  29. Eltzschig HK, "The use of readily available equipment in a simple method for intraoperative monitoring of recurrent laryngeal nerve function during thyroid surgery : initial experience with more than 300 cases", Arch Surg, 137(4), pp.452~456 (2002)
  30. Alesina PF, "Intraoperative neuromonitoring does not reduce the incidence of recurrent laryngeal nerve palsy in thyroid reoperations : results of a retrospective comparative analysis", World J Surg, 36(6), pp.1348~1353 (2012)
  31. Calo PG, "Interpretation of intraoperative recurrent laryngeal nerve monitoring signals : the importance of a correct standardization", Int J Surg, 28(Suppl 1), pp.S54~S58 (2016)
  32. Rulli F, "Meta-analysis of recurrent laryngeal nerve injury in thyroid surgery with or without intraoperative nerve monitoring", Acta Otorhinolaryngol Ital, 34(4), pp.223~229 (2014)
  33. Yang S, "Systematic review with meta-analysis of intraoperative neuromonitoring during thyroidectomy", Int J Surg, 39, pp.104~113 (2017)
  34. Snyder SK, "Intraoperative neurophysiology testing of the recurrent laryngeal nerve : plaudits and pitfalls", Surgery, 138(6), pp.1183~1191 (2005)
  35. Wu CW, "Transcutaneous recording during intraoperative neuromonitoring in thyroid surgery", Thyroid, 28(11), pp.1500~1507 (2018)
  36. Chavez KV, "Continuous intraoperative neural monitoring in thyroid surgery : a Mexican experience", Updates Surg, 69(4), pp.505~508 (2017)
  37. Sarkis LM, "Bilateral recurrent laryngeal nerve injury in a specialized thyroid surgery unit: would routine intraoperative neuromonitoring alter outcomes?", ANZ J Surg, 87(5), pp.364~367 (2017)
  38. Sitges-Serra A, "Contralateral surgery in patients scheduled for total thyroidectomy with initial loss or absence of signal during neural monitoring", Br J Surg, 106(4), pp.404~411 (2019)
  39. Sadowski SM, "Systematic use of recurrent laryngeal nerve neuromonitoring changes the operative strategy in planned bilateral thyroidectomy", Thyroid, 23(3), pp.329~333 (2013)
  40. Schneider R, "International neural monitoring study group guideline 2018 part I : staging bilateral thyroid surgery with monitoring loss of signal", Laryngoscope, 128(Suppl 3), pp.S1~S17 (2018)
  41. Genther DJ, "Correlation of final evoked potential amplitudes on intraoperative electromyography of the recurrent laryngeal nerve with immediate postoperative vocal fold function after thyroid and parathyroid surgery", JAMA Otolaryngol Head Neck Surg, 140(2), pp.124~128 (2014)
  42. Barczynski M, "International survey on the identification and neural monitoring of the EBSLN during thyroidectomy", Laryngoscope, 126(1), pp.285~291 (2016)
  43. Caragacianu D, "Intraoperative monitoring : normative range associated with normal postoperative glottic function", Laryngoscope, 123(12), pp.3026~3031 (2013)
  44. Pavier Y, "Acute prediction of laryngeal outcome during thyroid surgery by electromyographic laryngeal monitoring", Head Neck, 37(6), pp.835~839 (2015)