Efficacy of 4-Haptic Bitoric Intraocular Lens Implantation in Asian Patients with Cataract and Astigmatism

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류안과학

초록

Purpose: To determine the efficacy of 4-haptic bitoric intraocular lens (IOL) implantation in Asian patients withcataract and astigmatism. Methods: A total of 19 eyes with ≤25.0 mm axial length and ≥0.75 diopters (D) corneal astigmatism wereincluded in this prospective non-comparative study. All subjects underwent phacoemulsification with implantationof an AT Torbi 709M IOL. Visual and refractive outcomes as well as toric IOL axis were evaluatedduring a 3-month follow-up. Errors in predicted residual spherical equivalent were calculated by subtractingpredicted residual spherical equivalent from postoperative refraction. Results: Uncorrected and corrected distance visual acuity improved significantly 3 months after surgery, from0.43 to 0.05 and from 0.24 to -0.05, respectively. Mean refractive cylinders also decreased significantly, from-1.91 preoperatively to -0.54 D 3 months after surgery. Mean J0 and J45 decreased 3 months postoperatively,from 0.26 to 0.03 D and from 0.24 to -0.06 D, respectively. After 3 months, mean absolute IOL rotationwas 1.81º. Errors in predicted residual spherical equivalent showed a hyperopic shift of 0.35 D. Conclusions: Implantation of 4-haptic bitoric IOL proved to be effective for correcting astigmatism in Asianeyes during cataract surgery.

Abstract

Purpose: To determine the efficacy of 4-haptic bitoric intraocular lens (IOL) implantation in Asian patients withcataract and astigmatism. Methods: A total of 19 eyes with ≤25.0 mm axial length and ≥0.75 diopters (D) corneal astigmatism wereincluded in this prospective non-comparative study. All subjects underwent phacoemulsification with implantationof an AT Torbi 709M IOL. Visual and refractive outcomes as well as toric IOL axis were evaluatedduring a 3-month follow-up. Errors in predicted residual spherical equivalent were calculated by subtractingpredicted residual spherical equivalent from postoperative refraction. Results: Uncorrected and corrected distance visual acuity improved significantly 3 months after surgery, from0.43 to 0.05 and from 0.24 to -0.05, respectively. Mean refractive cylinders also decreased significantly, from-1.91 preoperatively to -0.54 D 3 months after surgery. Mean J0 and J45 decreased 3 months postoperatively,from 0.26 to 0.03 D and from 0.24 to -0.06 D, respectively. After 3 months, mean absolute IOL rotationwas 1.81º. Errors in predicted residual spherical equivalent showed a hyperopic shift of 0.35 D. Conclusions: Implantation of 4-haptic bitoric IOL proved to be effective for correcting astigmatism in Asianeyes during cataract surgery.

참고문헌 (22)

  1. Hoffer KJ, "Biometry of 7, 500 cataractous eyes", Am J Ophthalmol, 90, pp.360~368 (1980)
  2. Kim MJ, "Evaluation of optical performance of 4 aspheric toric intraocular lenses using an optical bench system : Influence of pupil size, decentration, and rotation", J Cataract Refract Surg, 41, pp.2274~2282 (2015)
  3. Kessel L, "Toric intraocular lenses in the correction of astigmatism during cataract surgery : a systematic review and meta-analysis", Ophthalmology, 123, pp.275~286 (2016)
  4. Novis C, "Astigmatism and toric intraocular lenses", Curr Opin Ophthalmol, 11, pp.47~50 (2000)
  5. Kretz FT, "Clinical outcomes and capsular bag stability of a four-point haptic bitoric intraocular lens", J Refract Surg, 31, pp.431~436 (2015)
  6. Mencucci R, "Clinical outcomes and rotational stability of a 4-haptic toric intraocular lens in myopic eyes", J Cataract Refract Surg, 40, pp.1479~1487 (2014)
  7. Kim JH, "The analysis of predicted capsular bag diameter using modified model of capsule measuring ring in Asians", Clin Exp Ophthalmol, 36, pp.238~244 (2008)
  8. Tehrani M, "Capsule measuring ring to predict capsular bag diameter and follow its course after foldable intraocular lens implantation", J Cataract Refract Surg, 29, pp.2127~2134 (2003)
  9. Thibos LN, "Power vector analysis of the optical outcome of refractive surgery", J Cataract Refract Surg, 27, pp.80~85 (2001)
  10. Buckhurst PJ, "Surgical correction of astigmatism during cataract surgery", Clin Exp Optom, 93, pp.409~418 (2010)
  11. Bascaran L, "Efficacy and stability of AT TORBI 709 M toric IOL", J Refract Surg, 29, pp.194~199 (2013)
  12. Yoon JJ, "Demographics and ocular biometric characteristics of patients undergoing cataract surgery in Auckland, New Zealand", Clin Exp Ophthalmol, 44, pp.106~113 (2016)
  13. Pan CW, "Ethnic variation in central corneal refractive power and steep cornea in Asians", Ophthalmic Epidemiol, 21, pp.99~105 (2014)
  14. Zhu X, "Factors influencing 1-year rotational stability of AcrySof Toric intraocular lenses", Br J Ophthalmol, 100, pp.263~268 (2016)
  15. Sheppard AL, "Clinical outcomes after implantation of a new hydrophobic acrylic toric IOL during routine cataract surgery", J Cataract Refract Surg, 39, pp.41~47 (2013)
  16. Ouchi M, "High-cylinder toric intraocular lens implantation versus combined surgery of low-cylinder intraocular lens implantation and limbal relaxing incision for high-astigmatism eyes", Clin Ophthalmol, 8, pp.661~667 (2014)
  17. Toto L, "Astigmatism correction with toric IOL : analysis of visual performance, position, and wavefront error", J Refract Surg, 29, pp.476~483 (2013)
  18. Chang DF, "Early rotational stability of the longer Staar toric intraocular lens : fifty consecutive cases", J Cataract Refract Surg, 29, pp.935~940 (2003)
  19. Patel CK, "Postoperative intraocular lens rotation : a randomized comparison of plate and loop haptic implants", Ophthalmology, 106, pp.2190~2195 (1999)
  20. Mamalis N, "Comparison of two plate-haptic intraocular lenses in a rabbit model", J Cataract Refract Surg, 22(Suppl 2), pp.1291~1295 (1996)
  21. Jampaulo M, "Long-term Staar toric intraocular lens rotational stability", Am J Ophthalmol, 146, pp.550~553 (2008)
  22. Li Y, "Effect of hydrophobic acrylic versus hydrophilic acrylic intraocular lens on posterior capsule opacification : meta-analysis", PLoS One, 8, pp.e77864~ (2013)

관련 논문AI추천

이 간행물의 다른 논문 (14건)