Abstract
A new type of transducer, piezoelectric floating mass transducer (PFMT) which has advantages of piezoelectric and electromagnetic transducer has been proposed and implemented for the implantable middle ear hearing devices. By the uneven bonding of piezoelectric material to the inner bottom of transducer case, the PFMT can vibrate back-and-forth along the longitudinal axis of the transducer even though the piezoelectric material within the cylindrical case produces only the bilateral expansion and contraction according to the applied electrical signal. To improve efficiency of the PFMT, the multi-layered piezoelectric material has been adapted. The small number of components in the PFMT enables the simple manufacturing and the easy implanting into the middle ear. In order to examine the characteristics of vibration, mechanical modeling and finite element analyses of the proposed transducer have been performed. From the result of theoretical analyses and the measured data from the experiment, it is verified that the implemented PFMT can be used in implantable middle ear hearing devices.
참고문헌 (13)
- R. Goode, "The history and development of the implantable hearing aid", 28(1), pp.1995~ (1995)
- A. Chasin, "Current trends in implantable hearing aids", 2(3), pp.84 1997~107 (1997)
- R. Goode, "Current status and future of implantable electromagnetic hearing aids", 28(1), pp.141 1995~145 (1995)
- T. Tsuiki, "補聽器と 人工中耳Ó Journal of Oto- laryngology", 11, pp.533~536 (1995)
- N. Yanagihara, "Partially implantable hearing aid using piezoelectric ceramic ossicular vibration", 28(1), pp.85~97 (1995)
- W. Ko, "Engineering principles of mechanical stimulation of the middle ear", 28(1), pp.29~41 (1995)
- M. Kompis, "Design consideration for a contactless electromagnetic transducer for implantable hearing aids", (6), pp.3137~3176 (1998)
- F. Sink, "The effect of the floating mass transducer in hearing sensitivity", 21(1), pp.42 2000~48 (2000)
- J.-H. Cho, "Middle ear hearing aid transducer", 6 735, pp.318~ (2004)
- 조진호, "전자 트랜스듀서 방식에 의한 청각보조용 이식형 인공중이 시스템 설계", 6(6), pp.466~475 (1997)
- 송병섭, "이식형 인공중이 시스템을 위한 고효율 차동 전자 트랜스듀서의 설계", 11(3), pp.171~182 (2002)
- 선주헌 등, "압전 단결정 재료를 이용한 이식형 인공중이용 적층 액츄에이터 개발", 24(3), pp.127~132 (2005)
- Z. G. Wang, "Assessment of multi-layer piezoelectric actuator technology for middle-ear implants", 12, pp.3~17 (2002)