Evaluation of Image Quality for Various Electronic Portal Imaging Devices in Radiation Therapy
0
조회
0
다운로드
13
인용
0
참고문헌
키워드
초록
In megavoltage (MV) radiotherapy, delivering the dose to the target volume is important while protecting the surrounding normal tissue. The purpose of this study was to evaluate the modulation transfer function (MTF), the noise power spectrum (NPS), and the detective quantum efficiency (DQE) using an edge block in mega-voltage X-ray imaging (MVI).We used an edge block, which consists of tungsten with dimensions of 19 (thickness) × 10 (length) × 1 (width) cm3 and measured the pre-sampling MTF at 6 MV energy. Various radiation therapy (RT) devices such as TrueBeamTM (Varian), BEAMVIEWPLUS (Siemens), iViewGT (Elekta) and Clinac®iX (Varian) were used. As for MTF results, TrueBeamTM(Varian) flattening filter free(FFF) showed the highest values of 0.46 mm-1and1.40mm-1for MTF 0.5 and 0.1. In NPS, iViewGT (Elekta) showed the lowest noise distribution. In DQE, iViewGT (Elekta) showed the best efficiency at a peak DQE and 1 mm-1DQE of 0.0026 and 0.00014, respectively.This study could be used not only for traditional QA imaging but also for quantitative MTF, NPS, and DQE measurement for development of an electronic portal imaging device (EPID).
Abstract
In megavoltage (MV) radiotherapy, delivering the dose to the target volume is important while protecting the surrounding normal tissue. The purpose of this study was to evaluate the modulation transfer function (MTF), the noise power spectrum (NPS), and the detective quantum efficiency (DQE) using an edge block in mega-voltage X-ray imaging (MVI).We used an edge block, which consists of tungsten with dimensions of 19 (thickness) × 10 (length) × 1 (width) cm3 and measured the pre-sampling MTF at 6 MV energy. Various radiation therapy (RT) devices such as TrueBeamTM (Varian), BEAMVIEWPLUS (Siemens), iViewGT (Elekta) and Clinac®iX (Varian) were used. As for MTF results, TrueBeamTM(Varian) flattening filter free(FFF) showed the highest values of 0.46 mm-1and1.40mm-1for MTF 0.5 and 0.1. In NPS, iViewGT (Elekta) showed the lowest noise distribution. In DQE, iViewGT (Elekta) showed the best efficiency at a peak DQE and 1 mm-1DQE of 0.0026 and 0.00014, respectively.This study could be used not only for traditional QA imaging but also for quantitative MTF, NPS, and DQE measurement for development of an electronic portal imaging device (EPID).
참고문헌 (24)
- Samei E, "An experimental comparison of detector performance for direct and indirect digital radiography systems", Med Phys, 30, pp.608~ (2003)
- Kim HK, "Development and evaluation of a digital radiographic system based on CMOS image sensor", IEEE Trans Nucl Sci, 3, pp.662~ (2001)
- 민정환, "텅스텐 엣지 블록을 이용하여 Megavoltage (MV) 영상의 질 평가", 의학물리, 23(3), pp.154~161 (2012), 한국의학물리학회
- 정회원, "CsI(TI) Indirect Flat Panel Detector의 선질에 따른 물리적 영상 평가", 방사선기술과학, 35(2), pp.109~117 (2012), 대한방사선과학회[DOI]
- White JE, "The influence of radiation therapy quality control on survival, response and sites of relapse in oat cell carcinoma of the lung:preliminary report of a Southwest Oncology Group study", Cancer, 50, pp.1084~ (1982)
- Goitein M, "Immobilization errors: Some theoretical considerations", Radiology, 117, pp.407~ (1975)
- Marks JE, "Localization error in the radiotherapy of Hodgkin's disease and malignant lymphoma with extended mantle fields", Cancer, 34, pp.83~ (1974)
- Marks JE, "The value of frequent treatment verification films in reducing localization error in the irradiation of complex fields", Cancer, 37, pp.2755~ (1976)
- Gopal A, "Validity of the line-pair bar-pattern method in the measurement of the modulation transfer function (MTF) in megavoltage imaging", Med Phys, 35, pp.270~ (2008)
- 민정환, "Comparison of Noise Power Spectrum Methodologies in Measurements by Using Megavoltage X-ray Energies", Journal of the Korean Physical Society, 60(1), pp.129~136 (2012), 한국물리학회
- Giger ML, "Investigation of basic imaging properties in digital radiography. 1. Modulation transfer function", Med Phys, 287, pp.11~ (1984)
- Fujita H, "A simple method for determining the modulation transfer function in digital radiography", IEEE Trans Med Imaging, 11, pp.34~ (1992)
- Samei E, "A method for measuring the presampled MTF of digital radiographic systems using an edge test device", Med Phys, 25, pp.102~ (1998)
- Greer PB, "Evaluation of an algorithm for the assessment of the MTF using an edge method", Med Phys, 27, pp.2048~ (2000)
- Cunningham A, "Signal and noise in modulation transfer function determinations using the slit, wire and edge techniques", Med Phys, 19, pp.1037~ (1992)
- Hillen W, "Imaging performance of a digital storage phosphor system", Med Phys, 14, pp.744~ (1987)
- Rogge F, "A practical method for detected quantum efficiency (DQE) assessment of digital mammography systems in the radiological environment", Proc SPIE, 4682, pp.645~ (2002)
- Cremers F, "Performance of electronic portal imaging devices (EPIDs) used in radio-therapy: image quality and dose measurements", Med Phys, 31, pp.985~ (2004)
- Siewerdsen H, "Empirical and theoretical investigation of the noise performance of indirect detection, active matrix flatpanel imagers (AMFPIs) for diagnostic radiology", Med Phys, 24, pp.71~ (1997)
- Jung-Whan Min, "Performance of an Edge Block Used in a Configuration Detector: Image Quality Measurements", Journal of the Korean Physical Society, 64(5), pp.732~739 (2014), 한국물리학회
- Jung-Whan Min, "Evaluation of the Modulation Transfer Function of Megavoltage X-rays", Journal of the Korean Physical Society, 65(11), pp.1969~1975 (2014), 한국물리학회
- Soon-Yong Son, "Evaluation of an Edge Method for Computed Radiography and an Electronic Portal Imaging Device in Radiotherapy: Image Quality Measurements", Journal of the Korean Physical Society, 65(11), pp.1976~1984 (2014), 한국물리학회
- Andrew G, "Accurate MTF measurement in digital radiography using noise response", Med Phys, 37, pp.724~ (2010)
- Gopal A, "Use of a line-pair resolution phantom for comprehensive quality assurance of electronic portal imaging devices based on fundamental imaging metrics", Med Phys, 36, pp.2006~ (2009)
