선형가속기의 시간에 따르는 전자조사문영상기구의 잡음전력스펙트럼 비교 평가

Evaluation of Comparison of Noise Power Spectrum according to the Time of Using Electronic Portal Imaging Device (EPID) for LINAC System

0

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류방사선과학

초록

This study was to assessment of quality assurance (QA) and noise characteristics of Noise Power Spectrum (NPS) according to the time of by using electronic portal imaging device (EPID) for LINAC (Linear Accelerator). LINAC device was (Varian ClinacR iX LINAC, USA) used and the were 40 × 30 cm2 of detector size were 1024 × 768 photo- electric diode array size. Signal could be obtained the K-space image of white noise images for NPS and we used to Overlap, Non-Overlap, Out of Penumbra, Flatness, Symmetry, Symmetry Rt, Lt methods. The 2013s NPS image Out of Penumbra quantitatively value more than 2013s NPS image Symmetry Rt, Lt methods quantitatively NPS based on the frequency of 1.0 mm-1. Thus, the 2022s NPS image Out of Penumbra quantitatively value more than 2022s NPS image Symmetry Rt, Lt methods quantitatively NPS based on the frequency of 1.0 mm-1. The assessment of comparison of white noise for NPS image noise and intensity of this study were to that should be used efficiently of the LINAC EPID detector system for Overlap method for International Electro-technical Commission (IEC).

Abstract

This study was to assessment of quality assurance (QA) and noise characteristics of Noise Power Spectrum (NPS) according to the time of by using electronic portal imaging device (EPID) for LINAC (Linear Accelerator). LINAC device was (Varian ClinacR iX LINAC, USA) used and the were 40 × 30 cm2 of detector size were 1024 × 768 photo- electric diode array size. Signal could be obtained the K-space image of white noise images for NPS and we used to Overlap, Non-Overlap, Out of Penumbra, Flatness, Symmetry, Symmetry Rt, Lt methods. The 2013s NPS image Out of Penumbra quantitatively value more than 2013s NPS image Symmetry Rt, Lt methods quantitatively NPS based on the frequency of 1.0 mm-1. Thus, the 2022s NPS image Out of Penumbra quantitatively value more than 2022s NPS image Symmetry Rt, Lt methods quantitatively NPS based on the frequency of 1.0 mm-1. The assessment of comparison of white noise for NPS image noise and intensity of this study were to that should be used efficiently of the LINAC EPID detector system for Overlap method for International Electro-technical Commission (IEC).

참고문헌 (15)

  1. Son SY, "Evaluation of image quality for various electronic portal imaging devices in radiation therapy", Journal of Radiological Science and Technology, 38(4), pp.451~461 (2015)
  2. Miyati T, "Measurements of MTF and SNR(f)using a subtraction method im MRI", Physics in Medicine and Biology, 47, pp.2961~2972 (2002)[DOI]
  3. Min JW, "Comparison study on CNR and SNR of thoracic spine Lateral radiography", Journal of Radiological Science and Technology, 36(4), pp.280~273 (2013)
  4. Lee YJ, "Optimization of non-local means algorithm in low-dose computed tomographic image using noise level and similarity evaluation parameters : A phantom study", Journal of Radiological Science and Technology, 47(1), pp.39~48 (2024)[DOI]
  5. Min JW, "Study on the resolution characteristics by using magnetic resonance imaging 3. 0T", Journal of Radiological Science and Technology, 43(4), pp.251~257 (2020)
  6. Min JW, "Evaluation of noise power spectrum characteristics by using magnetic resonance imaging 3. 0T", Journal of Radiological Science and Technology, 44(1), pp.279~288 (2021)
  7. Min JW, "Evaluation of the resolution characteristics by using american college of radiology phantom for magnetic resonance imaging", Journal of Radiological Science and Technology, 45(1), pp.11~17 (2022)[DOI]
  8. Jeong HW, "Performance characteristic of a CsI(Tl)flat panel detector radiography system", Journal of Radiological Science and Technology, 35(2), pp.109~117 (2012)
  9. Min JW, "Evaluation of the resolution characteristics by using ATS 535H phantom for ultrasound medical imaging", Journal of Radiological Science and Technology, 46(1), pp.15~21 (2023)[DOI]
  10. Kim KW, "Measurement of image quality according to the time of computed radiography system", Journal of Radiological Science and Technology, 38(4), pp.365~374 (2015)
  11. Min JW, "Comparison of noise power spectrum in measurements by using international electrotechnical commission standard devices in indirect digital radiography", Journal of Radiological Science and Technology, 41(5), pp.457~462 (2018)
  12. "IEC (International Electrotechnical Commission)62220-1. Medical electrical equipment characteristics of digital X-ray imaging devices Part 1: determination of the detective quantum efficiency" (2003)[URL]
  13. Samei E, "A method for measuring the presampled MTF of digital radiographic systems using an edge test device", Medical Physics, 25(1), pp.102~113 (1998)[DOI]
  14. Greer PB, "Evaluation of an algorithm for the assessment of the MTF using an edge method", Medical Physics, 27(9), pp.2048~2059 (2000)[DOI]
  15. Steckner MC, "Computing the modulation transfer function of a magnetic resonance imager", Medical Physics, 21(3), pp.483~489 (1994)[DOI]

관련 논문AI추천

방사선치료 시 전자포털 영상장치를 이용한 잡음전력스펙트럼 수평 및 수직 측정비권경태, 김기원, 민정환, 손순룡, 손진현, 장서구, 정회원, 최관우대한방사선과학회 · Journal of Radiological Science and Technology2016 · 제39권 제2호방사선치료 시 전자포털영상장치를 이용한 잡음전력스펙트럼 방법론 측정비교권경태, 김기원, 민정환, 손순룡, 손진현, 이영아, 정회원, 최관우대한방사선과학회 · Journal of Radiological Science and Technology2016 · 제39권 제1호Evaluation of Image Quality for Various Electronic Portal Imaging Devices in Radiation Therapy권경태, 김기원, 김정민, 민정환, 손순룡, 손진현, 이영아, 이제희, 정재용, 정회원, 조정희, 최관우대한방사선과학회 · Journal of Radiological Science and Technology2015 · 제38권 제4호간접평판형 검출기에서 국제전자기술위원회 기준을 통한 잡음전력스펙트럼 비교 연구권경태, 김기원, 김현수, 민정환, 손진현, 정재용, 정회원대한방사선과학회 · Journal of Radiological Science and Technology2018 · 제41권 제5호전산화단층 모의치료장치의 정도관리 항목 제안김연래, 윤영우, 이정우, 정재용, 정진범대한방사선과학회 · Journal of Radiological Science and Technology2021 · 제44권 제4호Measurement of Image Quality According to the Time of Computed Radiography System권경태, 김기원, 김정민, 민정환, 손순룡, 손진현, 이영아, 이익표, 정재용, 정회원, 최관우, 황선광대한방사선과학회 · Journal of Radiological Science and Technology2015 · 제38권 제4호

이 간행물의 다른 논문 (5건)