선형가속기의 시간에 따르는 전자조사문영상기구의 잡음전력스펙트럼 비교 평가
Evaluation of Comparison of Noise Power Spectrum according to the Time of Using Electronic Portal Imaging Device (EPID) for LINAC System
0
조회
0
다운로드
0
인용
0
참고문헌
키워드
초록
This study was to assessment of quality assurance (QA) and noise characteristics of Noise Power Spectrum (NPS) according to the time of by using electronic portal imaging device (EPID) for LINAC (Linear Accelerator). LINAC device was (Varian ClinacR iX LINAC, USA) used and the were 40 × 30 cm2 of detector size were 1024 × 768 photo- electric diode array size. Signal could be obtained the K-space image of white noise images for NPS and we used to Overlap, Non-Overlap, Out of Penumbra, Flatness, Symmetry, Symmetry Rt, Lt methods. The 2013s NPS image Out of Penumbra quantitatively value more than 2013s NPS image Symmetry Rt, Lt methods quantitatively NPS based on the frequency of 1.0 mm-1. Thus, the 2022s NPS image Out of Penumbra quantitatively value more than 2022s NPS image Symmetry Rt, Lt methods quantitatively NPS based on the frequency of 1.0 mm-1. The assessment of comparison of white noise for NPS image noise and intensity of this study were to that should be used efficiently of the LINAC EPID detector system for Overlap method for International Electro-technical Commission (IEC).
Abstract
This study was to assessment of quality assurance (QA) and noise characteristics of Noise Power Spectrum (NPS) according to the time of by using electronic portal imaging device (EPID) for LINAC (Linear Accelerator). LINAC device was (Varian ClinacR iX LINAC, USA) used and the were 40 × 30 cm2 of detector size were 1024 × 768 photo- electric diode array size. Signal could be obtained the K-space image of white noise images for NPS and we used to Overlap, Non-Overlap, Out of Penumbra, Flatness, Symmetry, Symmetry Rt, Lt methods. The 2013s NPS image Out of Penumbra quantitatively value more than 2013s NPS image Symmetry Rt, Lt methods quantitatively NPS based on the frequency of 1.0 mm-1. Thus, the 2022s NPS image Out of Penumbra quantitatively value more than 2022s NPS image Symmetry Rt, Lt methods quantitatively NPS based on the frequency of 1.0 mm-1. The assessment of comparison of white noise for NPS image noise and intensity of this study were to that should be used efficiently of the LINAC EPID detector system for Overlap method for International Electro-technical Commission (IEC).
참고문헌 (15)
- Son SY, "Evaluation of image quality for various electronic portal imaging devices in radiation therapy", Journal of Radiological Science and Technology, 38(4), pp.451~461 (2015)
- Miyati T, "Measurements of MTF and SNR(f)using a subtraction method im MRI", Physics in Medicine and Biology, 47, pp.2961~2972 (2002)[DOI]
- Min JW, "Comparison study on CNR and SNR of thoracic spine Lateral radiography", Journal of Radiological Science and Technology, 36(4), pp.280~273 (2013)
- Lee YJ, "Optimization of non-local means algorithm in low-dose computed tomographic image using noise level and similarity evaluation parameters : A phantom study", Journal of Radiological Science and Technology, 47(1), pp.39~48 (2024)[DOI]
- Min JW, "Study on the resolution characteristics by using magnetic resonance imaging 3. 0T", Journal of Radiological Science and Technology, 43(4), pp.251~257 (2020)
- Min JW, "Evaluation of noise power spectrum characteristics by using magnetic resonance imaging 3. 0T", Journal of Radiological Science and Technology, 44(1), pp.279~288 (2021)
- Min JW, "Evaluation of the resolution characteristics by using american college of radiology phantom for magnetic resonance imaging", Journal of Radiological Science and Technology, 45(1), pp.11~17 (2022)[DOI]
- Jeong HW, "Performance characteristic of a CsI(Tl)flat panel detector radiography system", Journal of Radiological Science and Technology, 35(2), pp.109~117 (2012)
- Min JW, "Evaluation of the resolution characteristics by using ATS 535H phantom for ultrasound medical imaging", Journal of Radiological Science and Technology, 46(1), pp.15~21 (2023)[DOI]
- Kim KW, "Measurement of image quality according to the time of computed radiography system", Journal of Radiological Science and Technology, 38(4), pp.365~374 (2015)
- Min JW, "Comparison of noise power spectrum in measurements by using international electrotechnical commission standard devices in indirect digital radiography", Journal of Radiological Science and Technology, 41(5), pp.457~462 (2018)
- "IEC (International Electrotechnical Commission)62220-1. Medical electrical equipment characteristics of digital X-ray imaging devices Part 1: determination of the detective quantum efficiency" (2003)[URL]
- Samei E, "A method for measuring the presampled MTF of digital radiographic systems using an edge test device", Medical Physics, 25(1), pp.102~113 (1998)[DOI]
- Greer PB, "Evaluation of an algorithm for the assessment of the MTF using an edge method", Medical Physics, 27(9), pp.2048~2059 (2000)[DOI]
- Steckner MC, "Computing the modulation transfer function of a magnetic resonance imager", Medical Physics, 21(3), pp.483~489 (1994)[DOI]