뇌동맥류 코일 색전술 시술 시 동맥류 크기 및 위치에 따른 방사선량 평가
Evaluation of Radiation Dose according to Aneurysm Size and Location during Cerebral Aneurysm Coil Embolization
1
조회
0
다운로드
1
인용
0
참고문헌
키워드
초록
Cerebral aneurysm coil embolization has the advantages of accurate, low patient burden, and fast recovery time, but efforts are needed to reduce dose due to the burden of exposure radiation dose during interventional procedures. In this study, the area dose product(DAP/Gy·cm2) and fluoro time(min) according to the size of the aneurysm and the location of aneurysm were investigated according to insurance recognition regulations aneurysm classification cerebral aneurysm coil embolization. According to the research method, classification according to the size and location of the aneurysm is first, the size of the aneurysm is divided into less than 4mm, more than 4mm to less than 8mm, and more than 8mm, and second, the dose to the area based on the location site (DAP/Gy·㎠) and fluoro time(min) based on the location site were observed. As a result, the location of the cerebral aneurysm procedure was found to be the Paraclinoid site. During cerebral aneurysm coil embolization, the area dose was 107 Gy·㎠ and fluoro time was 47.41 minutes, showing lower results than domestic studies, and when comparing the area dose product with foreign studies, the area dose product results were similar to that of Turkey and Saudi Arabia. It is expected that it can be used as an objective analysis indicator to establish diagnostic reference levels (DRLs) and patient radiation defense guidelines according to the size of cerebral aneurysm and location of cerebral aneurysm procedures during interventional procedures.
Abstract
Cerebral aneurysm coil embolization has the advantages of accurate, low patient burden, and fast recovery time, but efforts are needed to reduce dose due to the burden of exposure radiation dose during interventional procedures. In this study, the area dose product(DAP/Gy·cm2) and fluoro time(min) according to the size of the aneurysm and the location of aneurysm were investigated according to insurance recognition regulations aneurysm classification cerebral aneurysm coil embolization. According to the research method, classification according to the size and location of the aneurysm is first, the size of the aneurysm is divided into less than 4mm, more than 4mm to less than 8mm, and more than 8mm, and second, the dose to the area based on the location site (DAP/Gy·㎠) and fluoro time(min) based on the location site were observed. As a result, the location of the cerebral aneurysm procedure was found to be the Paraclinoid site. During cerebral aneurysm coil embolization, the area dose was 107 Gy·㎠ and fluoro time was 47.41 minutes, showing lower results than domestic studies, and when comparing the area dose product with foreign studies, the area dose product results were similar to that of Turkey and Saudi Arabia. It is expected that it can be used as an objective analysis indicator to establish diagnostic reference levels (DRLs) and patient radiation defense guidelines according to the size of cerebral aneurysm and location of cerebral aneurysm procedures during interventional procedures.
참고문헌 (19)
- 강병삼, "인터벤션시술 진단참고수준 평가", 방사선기술과학, 44(5), pp.451~457 (2021), 대한방사선과학회
- Berentein A, "Brugge: Surgical Neuroan giography", pp.365~453 (2003), Springer
- Brisman JL, "Cerebral aneurysms", The New England Journal of Medicine, 355(9), pp.928~939 (2006)[DOI]
- 조원상, "무증상으로 우연히 발견된 비파열성 뇌동맥류의 관리", 대한의사협회지, 63(5), pp.267~274 (2020), 대한의사협회
- Kim BT, "Pathogenesis of Cerebral Aneurysm", Korea Journal of Cerebrovascular Surgery, 10(1), pp.301~306 (2008)
- Vinuela F, "Guglielmi detach able coil embolization of acute intracranial aneurysm : Perioperative anatomical and clinical outcome in 403 patients", Journal of Neurosurgery, 86(3), pp.475~482 (1997)[DOI]
- Sharp C, "Avoidance of radiation injuries from medical interventional procedures", ICRP Pub, 85, pp.7~22 (2000)
- Kim JS, "Usefulness of Lens and Thyroid Screening Device for Hepatic Artery Embolization", Journal of Korean Society of Cardio-Vascular Interventional Technology, 19(1), pp.157~163 (2016)
- Kang BS, "Relationship between DAP and patient exposure dose during TACE procedure", Journal of the Korean Society of Cardiovascular Intervention Technology, 6(1), pp.191~198 (2003)
- Korea Centers for Disease Control and Prevention, "Establishment of diagnostic reference procedures for interventional radiotherapy, Induce reduction of patient exposure dose", ~ (2020)
- "Ministry of Health and Welfare Notice Revision Guide", ~ (2014)
- IEC, "IEC 60601-1-3: Medical Electrical Equipment-Part 1: General Requirements for Safety-3. Collateral Standard: General Requirements for Radiation Protection in Diagnostic X-ray Equipment", ~ (2008)
- IEC, "IEC 60601-2-54: Medical Electrical Equipment-Part 2–54: Particular Requirements for the Basic Safety and Essential Performance of X-ray Equipment for Radiography and Radioscopy", ~ (2009)
- 황준호, "인터벤션에서 자동노출제어장치와 수동노출 사용 시 두께 변화에 따른 선량감소 방안 연구", 방사선기술과학, 41(2), pp.115~122 (2018), 대한방사선과학회[DOI]
- Lee SH, "Clinical Analysis of Intracranial Aneurysms", Journal of Korean Neurosurgical Society, 18(3), pp.389~399 (1988)
- Vano E, "Patient dose in interventional radiology : A European survey", Radiation Protect Dosimetry, 129(1-3), pp.39~45 (2008)[DOI]
- Baiter S, "Patient radiation dose audits for fluoroscopically guided interventional procedures", Medical Physics, 38(3), pp.1611~1618 (2011)[DOI]
- Evaluation of Patient Doses in Interventional Radiology, "Chung JW", KFDA, ~ (2007)
- Mawad ME, "Long-term histopathologic changes in canine aneurysms embolized with Guglielmi detachable coils", American Journal of Neuroradiology, 16(1), pp.7~13 (1995)